Dosimetric evaluation of cone-beam CT-based synthetic CTs in pediatric patients undergoing intensity-modulated proton therapy

被引:13
作者
Sheikh, Khadija [1 ]
Liu, Dezhi [1 ]
Li, Heng [1 ]
Acharya, Sahaja [1 ]
Ladra, Matthew M. [1 ]
Hrinivich, William T. [1 ]
机构
[1] Johns Hopkins Univ, Sch Med, Dept Radiat Oncol & Mol Sci, Baltimore, MD USA
关键词
adaptive radiotherapy; cone-beam CT; image-guided radiotherapy; intensity-modulated proton therapy; pediatric cancer; DEFORMABLE IMAGE REGISTRATION; ONLY PHOTON;
D O I
10.1002/acm2.13604
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To evaluate dosimetric changes detected using synthetic computed tomography (sCT) derived from online cone-beam CTs (CBCT) in pediatric patients treated using intensity-modulated proton therapy (IMPT). Methods Ten pediatric patients undergoing IMPT and aligned daily using proton gantry-mounted CBCT were identified for retrospective analysis with treated anatomical sites fully encompassed in the CBCT field of view. Dates were identified when the patient received both a CBCT and a quality assurance CT (qCT) for routine dosimetric evaluation. sCTs were generated based on a deformable registration between the initial plan CT (pCT) and CBCT. The clinical IMPT plans were re-computed on the same day qCT and sCT, and dosimetric changes due to tissue change or response from the initial plan were computed using each image. Linear regression analysis was performed to determine the correlation between dosimetric changes detected using the qCT and the sCT. Gamma analysis was also used to compare the dose distributions computed on the qCT and sCT. Results The correlation coefficients (p-values) between qCTs and sCTs for changes detected in target coverage, overall maximum dose, and organ at risk dose were 0.97 (< .001), 0.84 (.002) and 0.91 (< .001), respectively. Mean +/- SD gamma pass rates of the sCT-based dose compared to the qCT-based dose at 3%/3 mm, 3%/2 mm, and 2%/2 mm criteria were 96.5%+/- 4.5%, 93.2%+/- 6.3%, and 91.3%+/- 7.8%, respectively. Pass rates tended to be lower for targets near lung. Conclusion While insufficient for re-planning, sCTs provide approximate dosimetry without administering additional imaging dose in pediatric patients undergoing IMPT. Dosimetric changes detected using sCTs are correlated with changes detected using clinically-standard qCTs; however, residual differences in dosimetry remain a limitation. Further improvements in sCT image quality may both improve online dosimetric evaluation and reduce imaging dose for pediatric patients by reducing the need for routine qCTs.
引用
收藏
页数:8
相关论文
共 20 条
[1]   EVALUATION OF PATIENT RESIDUAL DEVIATION AND ITS IMPACT ON DOSE DISTRIBUTION FOR PROTON RADIOTHERAPY [J].
Arjomandy, Bijan .
MEDICAL DOSIMETRY, 2011, 36 (03) :321-329
[2]   3D Slicer as an image computing platform for the Quantitative Imaging Network [J].
Fedorov, Andriy ;
Beichel, Reinhard ;
Kalpathy-Cramer, Jayashree ;
Finet, Julien ;
Fillion-Robin, Jean-Christophe ;
Pujol, Sonia ;
Bauer, Christian ;
Jennings, Dominique ;
Fennessy, Fiona ;
Sonka, Milan ;
Buatti, John ;
Aylward, Stephen ;
Miller, James V. ;
Pieper, Steve ;
Kikinis, Ron .
MAGNETIC RESONANCE IMAGING, 2012, 30 (09) :1323-1341
[3]   Deep learning-enabled MRI-only photon and proton therapy treatment planning for paediatric abdominal tumours [J].
Florkow, Mateusz C. ;
Guerreiro, Filipa ;
Zijlstra, Frank ;
Seravalli, Enrica ;
Janssens, Geert O. ;
Maduro, John H. ;
Knopf, Antje C. ;
Castelein, Rene M. ;
van Stralen, Marijn ;
Raaymakers, Bas W. ;
Seevinck, Peter R. .
RADIOTHERAPY AND ONCOLOGY, 2020, 153 :220-227
[4]   Revisiting the single-energy CT calibration for proton therapy treatment planning: a critical look at the stoichiometric method [J].
Goma, Carles ;
Almeida, Isabel P. ;
Verhaegen, Frank .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (23)
[5]   Cone-beam CT-derived relative stopping power map generation via deep learning for proton radiotherapy [J].
Harms, Joseph ;
Lei, Yang ;
Wang, Tonghe ;
McDonald, Mark ;
Ghavidel, Beth ;
Stokes, William ;
Curran, Walter J. ;
Zhou, Jun ;
Liu, Tian ;
Yang, Xiaofeng .
MEDICAL PHYSICS, 2020, 47 (09) :4416-4427
[6]   Multi-institutional Validation Study of Commercially Available Deformable Image Registration Software for Thoracic Images [J].
Kadoya, Noriyuki ;
Nakajima, Yujiro ;
Saito, Masahide ;
Miyabe, Yuki ;
Kurooka, Masahiko ;
Kito, Satoshi ;
Fujita, Yukio ;
Sasaki, Motoharu ;
Arai, Kazuhiro ;
Tani, Kensuke ;
Yagi, Masashi ;
Wakita, Akihisa ;
Tohyama, Naoki ;
Jingu, Keiichi .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02) :422-431
[7]   Investigating deformable image registration and scatter correction for CBCT-based dose calculation in adaptive IMPT [J].
Kurz, Christopher ;
Kamp, Florian ;
Park, Yang-Kyun ;
Zoellner, Christoph ;
Rit, Simon ;
Hansen, David ;
Podesta, Mark ;
Sharp, Gregory C. ;
Li, Minglun ;
Reiner, Michael ;
Hofmaier, Jan ;
Neppl, Sebastian ;
Thieke, Christian ;
Nijhuis, Reinoud ;
Ganswindt, Ute ;
Belka, Claus ;
Winey, Brian A. ;
Parodi, Katia ;
Landry, Guillaume .
MEDICAL PHYSICS, 2016, 43 (10) :5635-5646
[8]   Comparing cone- beam CT intensity correction methods for dose recalculation in adaptive intensity- modulated photon and proton therapy for head and neck cancer [J].
Kurz, Christopher ;
Dedes, George ;
Resch, Andreas ;
Reiner, Michael ;
Ganswindt, Ute ;
Nijhuis, Reinoud ;
Thieke, Christian ;
Belka, Claus ;
Parodi, Katia ;
Landry, Guillaume .
ACTA ONCOLOGICA, 2015, 54 (09) :1651-1657
[9]   A dosimetric comparison of proton and intensity modulated radiation therapy in pediatric rhabdorhyosarcoma patients enrolled on a prospective phase II proton study [J].
Ladra, Matthew M. ;
Edgington, Samantha K. ;
Mahajan, Anita ;
Grosshans, David ;
Szymonifka, Jackie ;
Khan, Fazal ;
Moteabbed, Maryam ;
Friedmann, Alison M. ;
MacDonald, Shannon M. ;
Tarbell, Nancy J. ;
Yock, Torunn I. .
RADIOTHERAPY AND ONCOLOGY, 2014, 113 (01) :77-83
[10]   Deep learning-based synthetic CT generation for paediatric brain MR-only photon and proton radiotherapy [J].
Maspero, Matteo ;
Bentvelzen, Laura G. ;
Savenije, Mark H. F. ;
Guerreiro, Filipa ;
Seravalli, Enrica ;
Janssens, Geert O. ;
van den Berg, Cornelis A. T. ;
Philippens, Marielle E. P. .
RADIOTHERAPY AND ONCOLOGY, 2020, 153 :197-204