Proximal humeral fractures: how stiff should an implant be? A comparative mechanical study with new implants in human specimens

被引:196
作者
Lill, H
Hepp, P
Korner, J
Kassi, JP
Verheyden, AP
Josten, C
Duda, GN
机构
[1] Univ Leipzig, Dept Trauma & Reconstruct Surg, D-04103 Leipzig, Germany
[2] Humboldt Univ, Dept Trauma & Reconstruct Surg, Res Lab, D-13353 Berlin, Germany
关键词
biomechanical testing; fractures; internal fixation; osteoporostis; proximal humerus;
D O I
10.1007/s00402-002-0465-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The objective of this study was to determine the in vitro characteristics of the clinically used and newly developed implants for the stabilization of proximal humeral fractures under static and cyclic loading. The goal was to optimize implant stiffness for fracture stabilization even in weak bone stock. Methods: In a laboratory study using 35 fresh human humeri, the specimens were randomized into 5 groups, which included the clinically used humerus T-plate (HTP), the cross-screw osteosynthesis (CSO), the unreamed proximal humerus nail with spiral blade (UHN), the recently developed Synclaw Proximal Humerus Nail (Synclaw PHN) and the angle-stable Locking Compression Plate Proximal Humerus (LCP-PH). The implant stiffness was determined for three clinically relevant load cases: axial compression, torsion and varus bending. In addition, a cyclic varus-bending test was performed to determine the implant properties under cyclic loading. Results: In contrast to a rather elastic and minimally invasive implant(LCP-PH), the conventionally designed ones (Synclaw PHN, CSO, HTP, UHN) showed rather high stiffness values under static loading. In cyclic loading, a strong decrease in stiffness (p<0.05) was found for the rigid implants HTP and UHN. In comparison with the other implants, only the elastic implant (LCP-PH) showed a significantly lower load reduction in a weak bone stock (17+/-6.2%). Conclusion: The high initial stiffness of rigid implants led to an early loosening and failure of the implant-bone interface under cyclic loading. Implants with low stiffness and elastic characteristics, however, appear to minimize the peak stresses at the bone-implant interface, making them particularly suitable for fracture fixation in osteoporotic bone.
引用
收藏
页码:74 / 81
页数:8
相关论文
共 21 条
[1]   Retrograde nailing of humeral shaft fractures with the UHN. An international multicenter survey [J].
Blum, J ;
Rommens, PM ;
Janzing, H ;
Langendorff, HS .
UNFALLCHIRURG, 1998, 101 (05) :342-352
[2]  
Blum J, 2000, UNFALLCHIRURG, V103, P183, DOI 10.1007/s001130050520
[3]   A BIOMECHANICAL COMPARISON OF INTRAMEDULLARY NAILING SYSTEMS FOR THE HUMERUS [J].
DALTON, JE ;
SALKELD, SL ;
SATTERWHITE, YE ;
COOK, SD .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1993, 7 (04) :367-374
[4]  
Duda GN, 1998, J BIOMECH, V31, P247, DOI 10.1016/S0021-9290(97)00115-2
[5]  
Henley M B, 1991, J Orthop Trauma, V5, P14, DOI 10.1097/00005131-199103000-00003
[6]   Semitubular blade plate fixation in proximal humeral fractures: A biomechanical study in a cadaveric model [J].
Instrum, K ;
Fennell, C ;
Shrive, N ;
Damson, E ;
Sonnabend, D ;
Hollinshead, R .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 1998, 7 (05) :462-466
[7]   Functional outcome after minimally displaced fractures of the proximal part of the humerus [J].
Koval, KJ ;
Gallagher, MA ;
Marsicano, JG ;
Cuomo, F ;
McShinawy, A ;
Zuckerman, JD .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (02) :203-207
[8]   Surgical neck fractures of the proximal humerus: A laboratory evaluation of ten fixation techniques [J].
Koval, KJ ;
Blair, B ;
Takei, R ;
Kummer, FJ ;
Zuckerman, JD .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1996, 40 (05) :778-783
[9]   MRT-arthrography - Improved diagnostic procedure in shoulder instability [J].
Lill, H ;
Lange, K ;
Reinbold, WD ;
Echtermeyer, V .
UNFALLCHIRURG, 1997, 100 (03) :186-192
[10]  
LILL H, 2001, IN PRESS UNFALLCHIRU