An in vitro study to optimise treatment of varicose veins with radiofrequency-induced thermo therapy

被引:19
作者
Badham, George E. [1 ]
Strong, Sophie M. [1 ]
Whiteley, Mark S. [1 ,2 ]
机构
[1] Whiteley Clin, Guildford GU2 7RF, Surrey, England
[2] Univ Surrey, Fac Hlth & Biomed Sci, Guildford GU2 5XH, Surrey, England
关键词
Radiofrequency ablation; thermoablation; radiofrequency-induced thermotherapy; varicose veins; INDUCED THERMOTHERAPY; SAPHENOUS-VEIN; BIPOLAR; RFITT;
D O I
10.1177/0268355514552005
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective To develop a reproducible method of using radiofrequency-induced thermotherapy with adequate thermal spread to ablate the whole vein wall in a truncal vein but avoiding carbonisation, device sticking and high impedance cut outs reducing interruptions during endovenous treatments. Methods Porcine liver was treated with radiofrequency-induced thermotherapy under glass to allow measurements, observation and video recording. Powers of 6-20W were used at varying pullback speeds to achieve linear endovenous energy densities of 18-100J/cm. Thermal spread, carbonisation of treated tissue and high-impedance cut outs were recorded. Results The currently recommended power settings of 18 and 20W produced sub-optimal results. If pulled back at 3s/cm to achieve linear endovenous energy densities around 60J/cm, tissue carbonisation and high impedance cut outs occurred. When high powers and fast pullbacks of 1s/cm were used, no carbonisation or cut outs occurred but thermal ablation of the liver model was below target due to reduced time for thermal conduction. Low powers (6-12W) with slow pullbacks (6-12s/cm) achieved target thermal ablation with minimal carbonisation and high impedance cut outs. Conclusions Using low power (6W) and a slow discontinuous pullback (6s every 0.5cm, in steps), we were able to achieve our target thermal ablation in the porcine liver model without carbonisation and high impedance cut outs. This suggests the currently recommended power levels could be reduced, reducing the need to remove the device to clean the electrodes during treatment while achieving target thermoablation of the treated tissue.
引用
收藏
页码:17 / 23
页数:7
相关论文
共 16 条
  • [1] Percutaneus treatment of varicose veins with bipolar radiofrequency ablation
    Boon, R.
    Akkersdijk, G. J. M.
    Nio, D.
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2010, 75 (01) : 43 - 47
  • [2] Radiofrequency-induced thermal therapy: results of a European multicentre study of resistive ablation of incompetent truncal varicose veins
    Braithwaite, B.
    Hnatek, L.
    Zierau, U.
    Camci, M.
    Akkersdijk, G. J. M.
    Nio, D.
    Sarlija, M.
    Ajduk, M.
    Santoro, P.
    Roche, E.
    [J]. PHLEBOLOGY, 2013, 28 (01) : 38 - 46
  • [3] Camci M, 2009, PHLEBOLOGIE, V38, P5
  • [4] Contrasting structure of the saphenous vein and internal mammary artery used as coronary bypass vessels
    Canham, PB
    Finlay, HM
    Boughner, DR
    [J]. CARDIOVASCULAR RESEARCH, 1997, 34 (03) : 557 - 567
  • [5] Technical characterization of a new bipolar and multipolar radiofrequency device for minimally invasive treatment of renal tumours
    Hacker, A
    Vallo, S
    Weiss, C
    Stein, T
    Alken, P
    Knoll, T
    Michel, MS
    [J]. BJU INTERNATIONAL, 2006, 97 (04) : 822 - 828
  • [6] Metcalfe M, 2007, SURGERY, V26, P4
  • [7] Strip-track revascularization after stripping of the great saphenous vein
    Munasinghe, A.
    Smith, C.
    Kianifard, B.
    Price, B. A.
    Holdstock, J. M.
    Whiteley, M. S.
    [J]. BRITISH JOURNAL OF SURGERY, 2007, 94 (07) : 840 - 843
  • [8] Optimising Treatment Parameters for Radiofrequency Induced Thermal Therapy (RFiTT): A Comparison of the Manufacturer's Treatment Guidance with a Locally Developed Treatment Protocol
    Newman, J. E.
    Meecham, L.
    Walker, R. J.
    Nyamekye, I. K.
    [J]. EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2014, 47 (06) : 664 - 669
  • [9] NICE National Institute for Health Care Excellence, 2013, Clinical guideline CG168
  • [10] Endovenous treatment of the great saphenous vein using a 1,320 nm Nd:YAG laser causes fewer side effects than using a 940 nm diode laser
    Proebstle, TM
    Moehler, T
    Gül, D
    Herdemann, S