3-DOF Force-Sensing Motorized Micro-Forceps for Robot-Assisted Vitreoretinal Surgery

被引:59
作者
Gonenc, Berk [1 ]
Chamani, Alireza [1 ]
Handa, James [2 ]
Gehlbach, Peter [2 ]
Taylor, Russell H. [1 ]
Iordachita, Iulian [1 ]
机构
[1] Johns Hopkins Univ, CISST ERC, Baltimore, MD 21218 USA
[2] Johns Hopkins Sch Med, Wilmer Eye Inst, Baltimore, MD 21287 USA
基金
美国国家卫生研究院;
关键词
Fiber Bragg grating; force sensing; micro-forceps; IATROGENIC RETINAL BREAKS; DIABETIC-RETINOPATHY; UNITED-STATES; MICROSURGICAL INSTRUMENT; EPIRETINAL MEMBRANES; STRAIN SENSOR; PREVALENCE; FEEDBACK; TEMPERATURE; TISSUE;
D O I
10.1109/JSEN.2017.2694965
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In vitreoretinal surgery, membrane peeling is a prototypical task where a layer of fibrous tissue is delaminated off the retina with a micro-forceps by applying very fine forces that are mostly imperceptible to the surgeon. Previously, we developed sensitized ophthalmic surgery tools based on fiber Bragg grating strain sensors, which were shown to precisely detect forces at the instrument's tip in two degrees of freedom perpendicular to the tool axis. This paper presents a new design that employs an additional sensor to capture also the tensile force along the tool axis. The grasping functionality is provided via a compact motorized unit. To compute forces, we investigate two distinct fitting methods: a linear regression and a nonlinear fitting based on second-order Bernstein polynomials. We carry out experiments to test the repeatability of sensor outputs, calibrate the sensor, and validate its performance. Results demonstrate sensor wavelength repeatability within 2 pm. Although the linear method provides sufficient accuracy in measuring transverse forces, in the axial direction, it produces a root mean square (rms) error over 3 mN even for a confined magnitude and direction of forces. On the other hand, the nonlinear method provides a more consistent and accurate measurement of both the transverse and axial forces for the entire force range (0-25 mN). Validation, including random samples, shows that our tool with the nonlinear force computation method can predict 3-D forces with an rms error under 0.15 mN in the transverse plane and within 2 mN accuracy in the axial direction.
引用
收藏
页码:3526 / 3541
页数:16
相关论文
共 68 条
  • [1] A Novel Technique Employing Tapered Fiber Bragg Grating to Solve the Axial/Transverse Forces Crosstalk in Microsurgical Instruments
    Abushagur, Abdulfatah A. Ghaith
    Bakar, Ahmad Ashrif A.
    Bin Zan, Mohd Saiful Dzulkefly
    Arsad, Norhana
    [J]. IEEE SENSORS JOURNAL, 2016, 16 (21) : 7671 - 7680
  • [2] Spinal needle force monitoring during lumbar puncture using fiber Bragg grating force device
    Ambastha, Shikha
    Umesh, Sharath
    Dabir, Sundaresh
    Asokan, Sundarrajan
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2016, 21 (11)
  • [3] Becker BC, 2012, P IEEE RAS-EMBS INT, P44, DOI 10.1109/BioRob.2012.6290813
  • [4] Integrating Optical Fiber Force Sensors into Microforceps for ORL Microsurgery
    Bell, Brett
    Stankowski, Stefan
    Moser, Benjamin
    Oliva, Vidina
    Stieger, Christof
    Nolte, Lutz-Peter
    Caversaccio, Marco
    Weber, Stefan
    [J]. 2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2010, : 1848 - 1851
  • [5] A miniature microsurgical instrument tip force sensor for enhanced force feedback during robot-assisted manipulation
    Berkelman, PJ
    Whitcomb, LL
    Taylor, RH
    Jensen, P
    [J]. IEEE TRANSACTIONS ON ROBOTICS AND AUTOMATION, 2003, 19 (05): : 917 - 922
  • [6] Chang D, 2013, IEEE INT C INT ROBOT, P413, DOI 10.1109/IROS.2013.6696384
  • [7] Double Peeling During Vitrectomy for Macular Pucker The Charles L-Schepens Lecture
    Chang, Stanley
    Gregory-Roberts, M.
    Park, Sungpyo
    Laud, Ketan
    Smith, Scott D.
    Hoang, Quan V.
    [J]. JAMA OPHTHALMOLOGY, 2013, 131 (04) : 525 - 530
  • [8] Robotic Vitreoretinal Surgery
    Channa, Roomasa
    Iordachita, Iulian
    Handa, James T.
    [J]. RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2017, 37 (07): : 1220 - 1228
  • [9] Techniques and tools for dissection of epiretinal membranes
    Charles, S
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2003, 241 (05) : 347 - 352
  • [10] Active depth-guiding handheld micro-forceps for Membranectomy based on CP-SSOCT
    Cheon, Gyeong Woo
    Lee, Phillip
    Gonenc, Berk
    Gehlbach, Peter L.
    Kang, Jin U.
    [J]. OPTICAL FIBERS AND SENSORS FOR MEDICAL DIAGNOSTICS AND TREATMENT APPLICATIONS XVI, 2016, 9702