Neurological and neuropsychological examination and outcome after use of an intra-aortic filter device during cardiac surgery

被引:14
作者
Eifert, S
Reichenspurner, H
Pfefferkorn, T
Baur, B
von Schlippenbach, C
Mayer, TE
Hamann, G
Reichart, B
机构
[1] Univ Munich, Klinikum Grosshadern, Herzchirurg Klin, Dept Cardiac Surg, D-81377 Munich, Germany
[2] Univ Hamburg, Dept Cardiothorac Surg, Hamburg, Germany
[3] Univ Munich, Klinikum Grosshadern, Dept Neurol, D-81377 Munich, Germany
[4] Univ Munich, Klinikum Grosshadern, Dept Neuroradiol, D-81377 Munich, Germany
来源
PERFUSION-UK | 2003年 / 18卷
关键词
D O I
10.1191/0267659103pf628oa
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Cerebral embolization of particles after cardiac surgery is frequently associated with neurological deficits. Aortic crossclamp manipulation seems to be the most significant cause of emboli release during cardiac surgery. The goal of this study was to demonstrate whether the use of an intra-aortic filter device has an effect on the magnet resonance imaging (MRI) and functional neurological outcome. Twenty-four patients undergoing cardiosurgical procedures using cardiopulmonary bypass (CPB) were selected: coronary artery bypass graft (CABG) surgery (n = 17), aortic valve replacement (AVR) surgery (n = 4) or combined procedures (n = 3). Patients were evaluated by diffusion weighted MRI of the brain, neurological examination and neuropsychological assessment regarding alertness as well as divided and selective attention before and five to seven days after surgery. The patients were divided into two groups. In group I, 12 patients received a filter through a modified 24 F arterial cannula immediately before the aortic crossclamp was released. Filters remained in the aorta until CPB was discontinued. Intraoperatively, bilateral middle cerebral artery transcranial Doppler (TCD) was monitored at baseline, at the beginning of CPB, at a timepoint when the aorta was crossclamped, when the filter was inserted and while the crossclamp was switched to partial clamping until the CPB was discontinued. TCD was used for detection of microembolic signals (MES). The captured material in the filter was examined histologically. Twelve patients served as controls without aortic filtration (group II). The MRI of the brain did not show any diffusion alterations in either group before or after surgery. No patient developed a focal neurological deficit or stroke. Intraoperative quantitative MES detection revealed a four to tenfold increase in patients of group I compared with group II (5-6 versus 0.5-1 MES/min) during the filter dwell time. There was no consistent pattern regarding the neurobehavioural sequelae. Filters showed arteriosclerotic debris in 75% of the patients. The use of the intra-aortic filter device did not show a positive effect on neurological, neuroradiographical and neuropsychological outcomes. The increase of the MES rate in group I patients may be due to microbubbles generated as microcavitations by the filter or the aortic filter cannula. The intra-aortic filter was able to capture atheromatous material in 75% of the patients.
引用
收藏
页码:55 / 60
页数:6
相关论文
共 20 条
[1]   Aortic atheroma is related to outcome but not numbers of emboli during coronary bypass [J].
Barbut, D ;
Lo, YW ;
Hartman, GS ;
Yao, FSF ;
Trifiletti, RR ;
Hager, DN ;
Hinton, RB ;
Gold, JP ;
Isom, OW .
ANNALS OF THORACIC SURGERY, 1997, 64 (02) :454-459
[2]   Aortic atheromatosis and risks of cerebral embolization [J].
Barbut, D ;
Gold, JP .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 1996, 10 (01) :24-30
[3]  
Barbut D, 1997, ANN THORAC SURG, V63, P1262
[4]   CEREBRAL EMBOLI DETECTED DURING BYPASS-SURGERY ARE ASSOCIATED WITH CLAMP REMOVAL [J].
BARBUT, D ;
HINTON, RB ;
SZATROWSKI, TP ;
HARTMAN, GS ;
BRUEFACH, M ;
WILLIAMSRUSSO, P ;
CHARLSON, ME ;
GOLD, JP .
STROKE, 1994, 25 (12) :2398-2402
[5]   Comparison of transcranial Doppler ultrasonography and transesophageal echocardiography to monitor emboli during coronary artery bypass surgery [J].
Barbut, D ;
Yao, FS ;
Hager, DN ;
Kavanaugh, P ;
Trifiletti, RR ;
Gold, JP .
STROKE, 1996, 27 (01) :87-90
[6]   Effect of aortic cannula characteristics and blood velocity on transcranial Doppler-detected microemboli during cardiopulmonary bypass [J].
Benaroia, M ;
Baker, AJ ;
Mazer, D ;
Errett, L .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 1998, 12 (03) :266-269
[7]  
Braekken SK, 1998, J NEUROL NEUROSUR PS, V65, P573
[8]  
Braekken SK, 1997, STROKE, V28, P1988
[9]   Neuropsychologic dysfunction after cardiac surgery: What is the problem? [J].
Gill, R ;
Murkin, JM .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 1996, 10 (01) :91-98
[10]   Capture of particulate emboli during cardiac procedures in which aortic cross-clamp is used [J].
Harringer, W .