Comparison of IDEXX SediVue Dx(R) urine sediment analyzer to manual microscopy for detection of casts in canine urine

被引:6
作者
Vasilatis, Demitria M. [1 ]
Cowgill, Larry D. [2 ]
Farace, Giosi [3 ]
Peterson, Sarah [3 ]
Yerramilli, Murthy [3 ]
Owens, Sean D. [1 ]
机构
[1] Univ Calif Davis, Sch Vet Med, Dept Pathol Microbiol & Immunol, Davis, CA 95616 USA
[2] Univ Calif Davis, Sch Vet Med, Dept Med & Epidemiol, Davis, CA 95616 USA
[3] IDEXX Labs Inc, Westbrook, ME USA
关键词
automated analyzer; dog; prevalence; urinalysis; urine sediment;
D O I
10.1111/jvim.16090
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background Detection of urinary casts is difficult due to their intermittent presence and deterioration in urine samples. Objective To compare the performance of the IDEXX SediVue Dx(R) Urine Sediment Analyzer (SediVue) with manual microscopy for the detection of urinary casts. We hypothesized that the SediVue analyzer would perform similarly to manual microscopy in cast detection. Animals Four hundred forty-three samples from 420 dogs from a hospital population. Methods This is a prospective, cross-sectional study. For SediVue analysis (software version [SW] 1.0.1.3), uncentrifuged urine was pipetted into a disposable cartridge. Seventy images were captured and processed by an onboard algorithm. For manual microscopy, urine was centrifuged to obtain sediment. Any cast identified by either method was considered a positive result (>0/low-power field [LPF]). SediVue images were evaluated if casts were detected by either methodology. A revised sensitivity and specificity were calculated after image review and when using a threshold of >1 cast/LPF. Results The sensitivity of the SediVue analysis for the detection of urinary casts was 53.7% (43.85%-63.35%), and specificity was 86.0% (81.78%-89.51%). After image review, the revised sensitivity/specificity was 52.0% (42.89%-61.02%) and 90.6% (86.81%-93.54%), respectively. When using a more clinically relevant threshold of >1/LPF, the sensitivity was 52.6% (35.82%-69.02%) and specificity was 99.3% (97.85%-99.85%). Conclusions and Clinical Importance The SediVue provides moderate agreement to manual methodology for detection of casts in urine.
引用
收藏
页码:1439 / 1447
页数:9
相关论文
共 19 条
  • [1] [Anonymous], 2011, GRAFFS TXB ROUTINE U
  • [2] Campbell T.W., 2012, Veterinary hematology and clinical chemistry, VSecond
  • [3] Comparison of the performance of the IDEXX SediVue Dx® with manual microscopy for the detection of cells and 2 crystal types in canine and feline urine
    Hernandez, Annalisa M.
    Bilbrough, Graham E. A.
    DeNicola, Dennis B.
    Myrick, Celine
    Edwards, Suzanne
    Hammond, Jeremy M.
    Myers, Alex N.
    Heseltine, Johanna C.
    Russell, Karen
    Giraldi, Marco
    Nabity, Mary B.
    [J]. JOURNAL OF VETERINARY INTERNAL MEDICINE, 2019, 33 (01): : 167 - 177
  • [4] IDEXX Laboratories Inc, 2016, SEDIVUE DX UR SED AN
  • [5] Ince Fatma Demet, 2016, Pract Lab Med, V5, P14, DOI 10.1016/j.plabm.2016.03.002
  • [6] Urine sediment examination: A comparison between the manual method and the iQ200 automated urine microscopy analyzer
    Lamchiagdhase, P
    Preechaborisutkul, K
    Lomsomboon, P
    Srisuchart, P
    Tantiniti, P
    Khan-u-ra, N
    Preechaborisutkul, B
    [J]. CLINICA CHIMICA ACTA, 2005, 358 (1-2) : 167 - 174
  • [7] MEASUREMENT OF OBSERVER AGREEMENT FOR CATEGORICAL DATA
    LANDIS, JR
    KOCH, GG
    [J]. BIOMETRICS, 1977, 33 (01) : 159 - 174
  • [8] Lerma EV., 2019, NEPHROLOGY SECRETS
  • [9] Analytical performance of the Iris iQ200 automated urine microscopy analyzer
    Linko, Solveig
    Kouri, Timo T.
    Toivonen, Ella
    Ranta, Paivi H.
    Chapoulaud, Eric
    Lalla, Martti
    [J]. CLINICA CHIMICA ACTA, 2006, 372 (1-2) : 54 - 64
  • [10] Assessment of Interobserver Reliability of Nephrologist Examination of Urine Sediment
    Palsson, Ragnar
    Colona, Mia R.
    Hoenig, Melanie P.
    Lundquist, Andrew L.
    Novak, James E.
    Perazella, Mark A.
    Waikar, Sushrut S.
    [J]. JAMA NETWORK OPEN, 2020, 3 (08)