Development and implementation of an automatic air delineation technique for MRI-guided adaptive radiation therapy

被引:5
作者
Ahunbay, Ergun [1 ]
Parchur, Abdul K. [1 ]
Paulson, Eric [1 ]
Chen, Xinfeng [1 ]
Omari, Eenas [1 ]
Li, X. Allen [1 ]
机构
[1] Med Coll Wisconsin, Dept Radiat Oncol, Milwaukee, WI 53226 USA
基金
美国国家卫生研究院;
关键词
MR-Linac; online adaptation; synthetic CT; MRgOART; RADIOTHERAPY; FIELD;
D O I
10.1088/1361-6560/ac7b65
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. Auto-delineation of air regions on daily MRI for MR-guided online adaptive radiotherapy (MRgOART) of abdominal tumors is challenging since the air packets occur randomly and their MR intensities can be similar to some other tissue types. This work reports a new method to auto-delineate air regions on MRI. Approach. The proposed method (named DIFF method) consists of (1) generating a combined volume V ( comb ), which is a union of the air-containing organs on a reference MR image offline, (2) transferring V ( comb ) from the reference MR to a daily MR via DIR, (3) combining the transferred V ( comb ) with a region of high DIR inaccuracy, and (4) applying a threshold to the obtained final combined volume to generate the air volumes. The high DIR inaccuracy region was calculated from the absolute difference between the deformed daily and the reference images. This method was tested on 36 abdominal daily MRI sets acquired from 7 patients on a 1.5 T MR-Linac. The performance of DIFF was compared with alternative auto-air generation methods that (1) does not account for DIR inaccuracies, and (2) uses rigid registration instead of DIR. Main results. The results show that the proposed DIFF method can be fully automated and can be executed within 25 s. The Dice similarity coefficient of manual and DIFF auto-generated air contours was >92% for all cases, while it was 90% for the alternative auto-delineation methods. Dosimetrically, the auto-generated air regions using DIFF resulted in practically identical DVHs as those generated by using manual air contours. Significance. The DIFF method is robust and accurate and can be implemented to automatically consider the inter- and intra- fractional air volume variations during MRgOART for abdominal tumors. The use of DIFF method improves dosimetric accuracy as compared to other methods, especially beneficial for the patients with large daily abdominal air volume variations.
引用
收藏
页数:9
相关论文
共 12 条
[1]   A Technique to Rapidly Generate Synthetic Computed Tomography for Magnetic Resonance Imaging-Guided Online Adaptive Replanning: An Exploratory Study [J].
Ahunbay, Ergun E. ;
Thapa, Ranjeeta ;
Chen, Xinfeng ;
Paulson, Eric ;
Li, X. Allen .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 103 (05) :1261-1270
[2]  
Chen X, 2021, 63 AAPM ANN M
[3]   A review of substitute CT generation for MRI-only radiation therapy [J].
Edmund, Jens M. ;
Nyholm, Tufve .
RADIATION ONCOLOGY, 2017, 12
[4]   The Magnetic Resonance Imaging-Linac System [J].
Lagendijk, Jan J. W. ;
Raaymakers, Bas W. ;
van Vulpen, Marco .
SEMINARS IN RADIATION ONCOLOGY, 2014, 24 (03) :207-209
[5]   Variations of MRI-assessed peristaltic motions during radiation therapy [J].
Mostafaei, Farshad ;
Tai, An ;
Omari, Eenas ;
Song, Yingqiu ;
Christian, James ;
Paulson, Eric ;
Hall, William ;
Erickson, Beth ;
Li, X. Allen .
PLOS ONE, 2018, 13 (10)
[6]   4D-MRI driven MR-guided online adaptive radiotherapy for abdominal stereotactic body radiation therapy on a high field MR-Linac: Implementation and initial clinical experience [J].
Paulson, Eric S. ;
Ahunbay, Ergun ;
Chen, Xinfeng ;
Mickevicius, Nikolai J. ;
Chen, Guang-Pei ;
Schultz, Christopher ;
Erickson, Beth ;
Straza, Michael ;
Hall, William A. ;
Li, X. Allen .
CLINICAL AND TRANSLATIONAL RADIATION ONCOLOGY, 2020, 23 :72-79
[7]  
Piper J., 2013, Deformable Image Registration in MIM Maestro Evaluation and Description, White Paper, MIM Software
[8]   Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue-air interfaces in a lateral magnetic field due to returning electrons [J].
Raaijmakers, AJE ;
Raaymakers, BW ;
Lagendijk, JJW .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (07) :1363-1376
[9]   First patients treated with a 1.5 T MRI-Linac: clinical proof of concept of a high-precision, high-field MRI guided radiotherapy treatment [J].
Raaymakers, B. W. ;
Jurgenliemk-Schulz, I. M. ;
Bol, G. H. ;
Glitzner, M. ;
Kotte, A. N. T. J. ;
van Asselen, B. ;
de Boer, J. C. J. ;
Bluemink, J. J. ;
Hackett, S. L. ;
Moerland, M. A. ;
Woodings, S. J. ;
Wolthaus, J. W. H. ;
van Zijp, H. M. ;
Philippens, M. E. P. ;
Tijssen, R. ;
Kok, J. G. M. ;
de Groot-van Breugel, E. N. ;
Kiekebosch, I. ;
Meijers, L. T. C. ;
Nomden, C. N. ;
Sikkes, G. G. ;
Doornaert, P. A. H. ;
Eppinga, W. S. C. ;
Kasperts, N. ;
Kerkmeijer, L. G. W. ;
Tersteeg, J. H. A. ;
Brown, K. J. ;
Pais, B. ;
Woodhead, P. ;
Lagendijk, J. J. W. .
PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (23) :L41-L50
[10]   Automated air region delineation on MRI for synthetic CT creation [J].
Thapa, Ranjeeta ;
Ahunbay, Ergun ;
Nasief, Haidy ;
Chen, Xinfeng ;
Li, X. Allen .
PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (02)