Comparison between Deff and Dw approaches for estimation of size-specific dose in paediatric CT imaging

被引:6
|
作者
Abdulkadir, Muhammad Kabir [1 ,2 ]
Shuaib, Ibrahim Lutfi [1 ]
Nasirudin, R. A. [1 ]
Samsudin, Ahmad Hadif Zaidin [3 ]
Osman, Noor Diyana [1 ,4 ]
机构
[1] Univ Sains Malaysia, Adv Med & Dent Inst, Bertam 13200, Penang, Malaysia
[2] Univ Ilorin, Fac Basic Clin Sci, Dept Med Radiog, Ilorin 240213, Nigeria
[3] Univ Sains Malaysia, Sch Med Sci, Dept Radiol, Kubang Kerian 16150, Kelantan, Malaysia
[4] Univ Sains Malaysia, Adv Med & Dent Inst, Oncol & Radiol Sci Cluster, Bertam 13200, Penang, Malaysia
关键词
Computed tomography; Size-specific dose estimate; Paediatric; Water-equivalent diameter; Effective diameter; ABDOMEN; CHEST; HEAD;
D O I
10.1016/j.radphyschem.2022.110205
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Individual patient size is an important determinant for size-specific dose estimate (SSDE) since the size correction factors are highly dependent on this factor. This study aims to compare two approaches for patient size measurement, the effective diameter (Deff) and water-equivalent diameter (Dw) for estimation of paediatric SSDE. The individual size of 318 paediatric patients (160 head and 158 abdomen CT) aged 0-12 years were measured retrospectively. The patient size was estimated using attenuation-based; Dw approach and geometry-based; Deff approach using anteroposterior (AP) and lateral dimensions on CT axial images. A comparative study was performed to determine the difference and association between these two approaches and their effects on SSDE. The measured patient size using Dw was higher than Deff by an average difference of about 12.8% and 9.3% for head and abdomen, respectively. There were high positive correlations between both approaches, head (R2 = 0.9176) and abdomen (R2 = 0.9491). The difference between measured Dw and Deff were observed by the Bland-Altman's analysis that showed good agreement between Dw and Deff. The overall disparity in SSDE calculated using Deff and Dw was less than 10%. The disparity was higher in head CT SSDE as compared to abdomen CT SSDE, due to lesser correlation between Dw and Deff in head CT scan. Our findings show Deff and Dw to be comparable size metrics for size-specific dose estimate in paediatric CT. The discrepancy between both dose metrics led to slight difference in SSDE calculated using Deff and Dw.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Advancing size-specific dose estimates in CT examinations: Dose estimates at longitudinal positions of scans
    Li, Xinhua
    Marschall, Theodore A.
    Yang, Kai
    Liu, Bob
    MEDICAL PHYSICS, 2022, 49 (02) : 1303 - 1311
  • [22] Optimizing CT radiation dose based on patient size and image quality: the size-specific dose estimate method
    Larson, David B.
    PEDIATRIC RADIOLOGY, 2014, 44 : 501 - 505
  • [23] SIZE-specific dose estimate for lower-limb CT
    Kobayashi, Masanao
    Nishihara, Yusei
    Haba, Tomonobu
    Matsunaga, Yuta
    Asada, Yasuki
    Kobayashi, Shigeki
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2022, 45 (04) : 1183 - 1191
  • [24] Evaluation of Size-Specific Dose Estimates for Optimizing Pediatric Chest CT Protocol
    Zarei F.
    Nasiri M.
    Etemadi Z.
    Haghighi R.R.
    Chatterjee S.
    Khaneghah P.A.
    Chatterjee V.V.
    Iranian Journal of Medical Physics, 2022, 19 (05): : 315 - 321
  • [25] Relationships between Size-specific Dose Estimate and Signal to Noise Ratio under Chest CT Examinations with Tube Current Modulation
    Qin, Tian
    Wang, Jing
    Wang, Mengting
    Gu, Ye
    Xie, Zongyu
    Liang, Baohui
    CURRENT MEDICAL IMAGING, 2024, 20
  • [26] Size-specific Dose Estimates for Adult Patients at CT of the Torso
    Christner, Jodie A.
    Braun, Natalie N.
    Jacobsen, Megan C.
    Carter, Rickey E.
    Kofler, James M.
    McCollough, Cynthia H.
    RADIOLOGY, 2012, 265 (03) : 841 - 847
  • [27] SIZE-specific dose estimate for lower-limb CT
    Masanao Kobayashi
    Yusei Nishihara
    Tomonobu Haba
    Yuta Matsunaga
    Yasuki Asada
    Shigeki Kobayashi
    Physical and Engineering Sciences in Medicine, 2022, 45 : 1183 - 1191
  • [28] Optimizing CT radiation dose based on patient size and image quality: the size-specific dose estimate method
    David B. Larson
    Pediatric Radiology, 2014, 44 : 501 - 505
  • [29] An improved method for automated calculation of the water-equivalent diameter for estimating size-specific dose in CT
    Anam, Choirul
    Mahdani, Fahmi Rosydiansyah
    Dewi, Winda Kusuma
    Sutanto, Heri
    Triadyaksa, Pandji
    Haryanto, Freddy
    Dougherty, Geoff
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (09): : 313 - 323
  • [30] Determination of water-equivalent diameter (Dw) and size-specific dose estimates (SSDE) for adult patients in abdominal CT examinations
    El Mansouri, M'hamed
    Tahiri, Zakaria
    Talbi, Mohammed
    Nhila, Oussama
    Sekkat, Hamza
    Choukri, Abdelmajid
    RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2024, 179 (5-6): : 634 - 641