Novel MEMS stiffness sensor for force and elasticity measurements

被引:34
作者
Peng, P. [1 ]
Sezen, A. S. [1 ]
Rajamani, R. [1 ]
Erdman, A. G. [1 ]
机构
[1] Univ Minnesota, Dept Mech Engn, Minneapolis, MN 55455 USA
基金
美国国家科学基金会;
关键词
Stiffness sensor; MEMS capacitive sensor; Surface micromachining; Silicon nitride membranes; ARTICULAR-CARTILAGE; TACTILE SENSOR; POISSONS RATIO; TISSUE; SURGERY;
D O I
10.1016/j.sna.2009.12.002
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper presents the design, mathematical model, fabrication and testing of a novel type of MEMS stiffness sensor. The proposed sensor can measure both contact force and the stiffness of the object under contact. The sensing concept utilizes multiple membranes with varying stiffnesses. The relative deflection of the sensing membranes can be measured to calculate the stiffness or elasticity of the targeted object. In order to validate the new sensing concept. MEMS capacitive sensors are fabricated using surface micromachining with each fabricated sensor having a 1 mm x 1 mm square active sensor area. Finally, the sensors are characterized by using them to touch polymers of different elastic stiffness values. The test results show that the fabricated sensor can discriminate elasticity variation of polymer specimens with a resolution of 0.1 MPa over a range of 0.7-1.2 MPa. The force sensing resolution is excellent and is as small as 0.2 mN over a force range of 0-0.5 N. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:10 / 17
页数:8
相关论文
共 25 条
[1]   Estimation of Young's modulus and Poisson's ratio of soft tissue from indentation using two different-sized indentors: finite element analysis of the finite deformation effect [J].
Choi, APC ;
Zheng, YP .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2005, 43 (02) :258-264
[2]   Modelling and testing of a sensor capable of determining the stiffness of biological tissues [J].
Dargahi, Javad ;
Najarian, Siamak ;
Mirjalili, Vahid ;
Liu, Bin .
CANADIAN JOURNAL OF ELECTRICAL AND COMPUTER ENGINEERING-REVUE CANADIENNE DE GENIE ELECTRIQUE ET INFORMATIQUE, 2007, 32 (01) :45-51
[3]   Tactile sensing technology for minimal access surgery - a review [J].
Eltaib, MEH ;
Hewit, JR .
MECHATRONICS, 2003, 13 (10) :1163-1177
[4]  
Giovanni M.D., 1982, Flat and corrugated diaphgram design handbook
[5]  
Hasler EM, 1999, CRIT REV BIOMED ENG, V27, P415
[6]   Determination of Poisson's ratio of articular cartilage by indentation using different-sized indenters [J].
Jin, H ;
Lewis, JL .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2004, 126 (02) :138-145
[7]  
Johnson K L, 1987, CONTACT MECH
[8]  
Lee HK, 2006, PROC IEEE MICR ELECT, P606
[9]  
Lyyra T, 1999, J BIOMED MATER RES, V48, P482, DOI 10.1002/(SICI)1097-4636(1999)48:4<482::AID-JBM13>3.0.CO
[10]  
2-M