Effect of Silver Annealing Conditions on the Performance of Electrolytic Silver/Silver Chloride Electrodes used in Harned Cell Measurements of pH

被引:14
作者
Brewer, Paul. J. [1 ]
Brown, Richard J. C. [1 ]
机构
[1] Natl Phys Lab, Analyt Sci Div, Teddington TW11 0LW, Middx, England
关键词
pH; Harned cell; Ag/AgCl electrodes; annealing; stability;
D O I
10.3390/s100302202
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We have studied the long and short term stability of electrolytic Ag/AgCl electrodes fabricated from Ag wire that has been subjected to a range of different annealing conditions. At elevated temperatures, the presence of oxygen during the annealing process has been shown to be detrimental to the performance of electrodes produced. This phenomenon has been attributed to the dissolution of oxygen in the Ag lattice leading to structural changes in the Ag/AgCl electrode material. Electrodes prepared from Ag wire annealed in the absence of oxygen have shown no appreciable change in performance throughout the temperature range employed. This work has resulted in an improved understanding of the optimum annealing conditions required for Ag used in the preparation of electrolytic Ag/AgCl reference electrodes. This work has positive implications for the accuracy of Harned cell measurements of pH.
引用
收藏
页码:2202 / 2216
页数:15
相关论文
共 23 条
[1]  
Bates R.G., 1973, DETERMINATION PH THE, V2nd
[2]   Effect of Structural Design of Silver/Silver Chloride Electrodes on Stability and Response Time and the Implications for Improved Accuracy in pH Measurement [J].
Brewer, Paul J. ;
Brown, Richard J. C. .
SENSORS, 2009, 9 (01) :118-130
[3]  
British Standards Institution, 1979, 25861979 BSI
[4]  
BROWN HC, 1971, GMELIN HDB ANORGANIS
[5]   Long-term equilibrium potential and electrochemical impedance study of Ag/AgCl electrodes used in Harned Cell measurements of pH [J].
Brown, Richard J. C. ;
Brewer, Paul J. ;
Brett, Dan J. L. .
ACCREDITATION AND QUALITY ASSURANCE, 2009, 14 (03) :139-145
[6]   The microporous structure of silver/silver chloride electrodes and the implications for Harned cell operation [J].
Brown, RJC ;
Milton, MJT .
ACCREDITATION AND QUALITY ASSURANCE, 2005, 10 (07) :352-355
[7]   Observation of a combined dilution and salting effect in buffers under conditions of high dilution and high ionic strength [J].
Brown, RJC ;
Milton, MJT .
ACCREDITATION AND QUALITY ASSURANCE, 2003, 8 (11) :505-510
[8]   Measurement of pH. Definition, standards, and procedures [J].
Buck, RP ;
Rondinini, S ;
Covington, AK ;
Baucke, FGK ;
Brett, CMA ;
Camoes, MF ;
Milton, MJT ;
Mussini, T ;
Naumann, R ;
Pratt, KW ;
Spitzer, P ;
Wilson, GS .
PURE AND APPLIED CHEMISTRY, 2002, 74 (11) :2169-2200
[9]  
*CCQM, 1998, CCQM REP, V4
[10]   pH electrode performance under automated management conditions [J].
Comer, JEA ;
Hibbert, C .