A Compact, Bone-Attached Robot for Mastoidectomy

被引:26
作者
Dillon, Neal P. [1 ]
Balachandran, Ramya [2 ]
Fitzpatrick, J. Michael [3 ]
Siebold, Michael A. [3 ]
Labadie, Robert F. [2 ]
Wanna, George B. [2 ]
Withrow, Thomas J. [1 ]
Webster, Robert J., III [1 ]
机构
[1] Vanderbilt Univ, Dept Mech Engn, 2301 Vanderbilt Pl,PMB 351592, Nashville, TN 37235 USA
[2] Vanderbilt Univ, Med Ctr, Dept Otolaryngol, Nashville, TN 37232 USA
[3] Vanderbilt Univ, Dept Elect Engn & Comp Sci, Nashville, TN 37235 USA
来源
JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME | 2015年 / 9卷 / 03期
关键词
SYSTEM;
D O I
10.1115/1.4030083
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Otologic surgery often involves a mastoidectomy, which is the removal of a portion of the mastoid region of the temporal bone, to safely access the middle and inner ear. The surgery is challenging because many critical structures are embedded within the bone, making them difficult to see and requiring a high level of accuracy with the surgical dissection instrument, a high-speed drill. We propose to automate the mastoidectomy portion of the surgery using a compact, bone-attached robot. The system described in this paper is a milling robot with four degrees-of-freedom (DOF) that is fixed to the patient during surgery using a rigid positioning frame screwed into the surface of the bone. The target volume to be removed is manually identified by the surgeon pre-operatively in a computed tomography (CT) scan and converted to a milling path for the robot. The surgeon attaches the robot to the patient in the operating room and monitors the procedure. Several design considerations are discussed in the paper as well as the proposed surgical workflow. The mean targeting error of the system in free space was measured to be 0.5 mm or less at vital structures. Four mastoidectomies were then performed in cadaveric temporal bones, and the error at the edges of the target volume was measured by registering a postoperative computed tomography (CT) to the pre-operative CT. The mean error along the border of the milled cavity was 0.38 mm, and all critical anatomical structures were preserved.
引用
收藏
页数:7
相关论文
共 25 条
[1]  
[Anonymous], 2014, CENGAGE LEARNING
[2]   Accuracy Evaluation of microTargeting Platforms for Deep-Brain Stimulation Using Virtual Targets [J].
Balachandran, Ramya ;
Mitchell, Jason E. ;
Dawant, Benoit M. ;
Fitzpatrick, J. Michael .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2009, 56 (01) :37-44
[3]  
Bell B, 2013, OTOL NEUROTOL, V34, P1284, DOI 10.1097/MAO.0b013e31829561b6
[4]   A self-developed and constructed robot for minimally invasive cochlear implantation [J].
Bell, Brett ;
Stieger, Christof ;
Gerber, Nicolas ;
Arnold, Andreas ;
Nauer, Claude ;
Hamacher, Volkmar ;
Kompis, Martin ;
Nolte, Lutz ;
Caversaccio, Marco ;
Weber, Stefan .
ACTA OTO-LARYNGOLOGICA, 2012, 132 (04) :355-360
[5]  
Danilchenko A., 2011, THESIS VANDERBILT U
[6]   Robotic Mastoidectomy [J].
Danilchenko, Andrei ;
Balachandran, Ramya ;
Toennies, Jenna L. ;
Baron, Stephan ;
Munske, Benjamin ;
Fitzpatrick, J. Michael ;
Withrow, Thomas J. ;
Webster, Robert J., III ;
Labadie, Robert F. .
OTOLOGY & NEUROTOLOGY, 2011, 32 (01) :11-16
[7]   Preliminary Testing of a Compact, Bone-Attached Robot for Otologic Surgery [J].
Dillon, Neal P. ;
Balachandran, Ramya ;
Falisse, Antoine Motte Dit ;
Wanna, George B. ;
Labadie, Robert F. ;
Withrow, Thomas J. ;
Fitzpatrick, J. Michael ;
Webster, Robert J., III .
MEDICAL IMAGING 2014: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2014, 9036
[8]  
Dillon NP, 2013, OTOL NEUROTOL, V34, pE93, DOI 10.1097/MAO.0b013e318291c76b
[9]   Development of the first force-controlled robot for otoneurosurgery [J].
Federspil, PA ;
Geisthoff, UW ;
Henrich, D ;
Plinkert, PK .
LARYNGOSCOPE, 2003, 113 (03) :465-471
[10]  
Fitzpatrick J. M., 2000, HDB MED IMAGING MED