Analytical expressions for the reconstructed image of a homogeneous cylindrical sample exhibiting a beam hardening artifact in X-ray computed tomography

被引:5
|
作者
Nakano, Tsukasa [1 ]
Nakashima, Yoshito [1 ]
机构
[1] Natl Inst Adv Ind Sci & Technol, Cent 7,Higashi 1-1-1, Tsukuba, Ibaraki 3058567, Japan
基金
日本学术振兴会;
关键词
Beam hardening; cupping artifact; exact solution; phantom; polychromatic X-ray spectrum; X-ray CT; SPECTRUM ESTIMATION; CT;
D O I
10.3233/XST-18378
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
BACKGROUND: Cylindrical phantoms are often imaged by X-ray computed tomography (CT) to evaluate the extent of beam hardening (or cupping artifact) resulting from a polychromatic X-ray source. OBJECTIVE: Our goal was to derive analytical expressions for the reconstructed image of a homogeneous cylindrical phantom exhibiting a cupping artifact, to permit a quantitative comparison with experimental cupping data. METHODS: A filtered backprojection method was employed to obtain the analytical cupping profile for the phantom, assuming that the projection data could be approximated as a power series with respect to the sample penetration thickness. RESULTS: The cupping profile was obtained analytically as a series of functions by employing Ramachandran filtering with an infinite Nyquist wavenumber. The quantitative relationship between the power series of the projection and the nth moment of the linear attenuation coefficient spectrum of the phantom was also determined. Application of the obtained cupping profile to the evaluation of the practical reconstruction filters with a finite Nyquist wavenumber and to the best choice of the contrast agent was demonstrated. CONCLUSIONS: The set of exact solutions derived in this work should be applicable to the analysis of cylindrical phantom experiments intended to evaluate CT systems.
引用
收藏
页码:691 / 705
页数:15
相关论文
共 50 条
  • [1] Statistical X-ray computed tomography image reconstruction with beam hardening correction
    Elbakri, IA
    Fessler, JA
    MEDICAL IMAGING: 2001: IMAGE PROCESSING, PTS 1-3, 2001, 4322 : 1 - 12
  • [2] Metal Artifact Reduction Based on Beam Hardening Correction and Statistical Iterative Reconstruction for X-ray Computed Tomography
    Zhang, Yanbo
    Mou, Xuanqin
    MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [3] Beam Hardening Correction Method for X-Ray Computed Tomography Based on Subsection Beam Hardening Curves
    Huang Kui-dong
    Zhang Ding-hua
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2009, 29 (09) : 2581 - 2584
  • [4] Beam hardening correction for X-ray computed tomography of heterogeneous natural materials
    Ketcham, Richard A.
    Hanna, Romy D.
    COMPUTERS & GEOSCIENCES, 2014, 67 : 49 - 61
  • [5] Noise Suppression Methods in Beam Hardening Correction for X-Ray Computed Tomography
    Huang, Kuidong
    Zhang, Dinghua
    Li, Mingjun
    PROCEEDINGS OF THE 2009 2ND INTERNATIONAL CONGRESS ON IMAGE AND SIGNAL PROCESSING, VOLS 1-9, 2009, : 564 - 568
  • [6] Reduce beam hardening artifacts of polychromatic X-ray computed tomography by an iterative approximation approach
    Shi, Hongli
    Yang, Zhi
    Luo, Shuqian
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2017, 25 (03) : 417 - 428
  • [7] A new approach for reducing beam hardening artifacts in polychromatic X-ray computed tomography using more accurate prior image
    Wang, Hui
    Xu, Yanan
    Shi, Hongli
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2018, 26 (04) : 593 - 602
  • [8] Theoretical Beam Hardening Correction for Industrial X-ray Computed Tomography
    Ahmed, Osama M. H.
    Song, YuShou
    Xie, Zhaoyang
    JOURNAL OF ENGINEERING AND TECHNOLOGICAL SCIENCES, 2019, 51 (06): : 869 - 880
  • [9] A Review of Common Beam Hardening Correction Methods for Industrial X-ray Computed Tomography
    Ahmed, O. M. H.
    Song, Yushou
    SAINS MALAYSIANA, 2018, 47 (08): : 1883 - 1890
  • [10] Beam hardening: Analytical considerations of the effective attenuation coefficient of x-ray tomography
    Alles, J.
    Mudde, R. F.
    MEDICAL PHYSICS, 2007, 34 (07) : 2882 - 2889