Recent advances in parallel imaging for MRI

被引:165
作者
Hamilton, Jesse [1 ]
Franson, Dominique [1 ]
Seiberlich, Nicole [1 ,2 ]
机构
[1] Case Western Reserve Univ, Biomed Engn, Cleveland, OH 44106 USA
[2] Univ Hosp Cleveland, Med Ctr, Radiol, Cleveland, OH 44106 USA
基金
美国国家科学基金会;
关键词
Parallel imaging; Non-Cartesian; Simultaneous multi-slice; Spectroscopic imaging; Phase-constrained; SIMULTANEOUS MULTISLICE EXCITATION; SENSITIVITY ENCODING SENSE; K-SPACE NEIGHBORHOODS; TIME SPIRAL GRAPPA; RADIAL GRAPPA; PHASED-ARRAY; TESLA; TEMPORAL RESOLUTION; SPATIAL-RESOLUTION; VOLUMETRIC MRI;
D O I
10.1016/j.pnmrs.2017.04.002
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Magnetic Resonance Imaging (MRI) is an essential technology in modern medicine. However, one of its main drawbacks is the long scan time needed to localize the MR signal in space to generate an image. This review article summarizes some basic principles and recent developments in parallel imaging, a class of image reconstruction techniques for shortening scan time. First, the fundamentals of MRI data acquisition are covered, including the concepts of k-space, undersampling, and aliasing. It is demonstrated that scan time can be reduced by sampling a smaller number of phase encoding lines in k space; however, without further processing, the resulting images will be degraded by aliasing artifacts. Nearly all modern clinical scanners acquire data from multiple independent receiver coil arrays. Parallel imaging methods exploit properties of these coil arrays to separate aliased pixels in the image domain or to estimate missing k-space data using knowledge of nearby acquired k-space points. Three parallel imaging methods SENSE, GRAPPA, and SPIRiT are described in detail, since they are employed clinically and form the foundation for more advanced methods. These techniques can be extended to non Cartesian sampling patterns, where the collected k-space points do not fall on a rectangular grid. Non Cartesian acquisitions have several beneficial properties, the most important being the appearance of incoherent aliasing artifacts. Recent advances in simultaneous multi-slice imaging are presented next, which use parallel imaging to disentangle images of several slices that have been acquired at once. Parallel imaging can also be employed to accelerate 3D MRI, in which a contiguous volume is scanned rather than sequential slices. Another class of phase-constrained parallel imaging methods takes advantage of both image magnitude and phase to achieve better reconstruction performance. Finally, some applications are presented of parallel imaging being used to accelerate MR Spectroscopic Imaging. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:71 / 95
页数:25
相关论文
共 169 条
[1]   Evaluation of left ventricular ejection fraction using through-time radial GRAPPA [J].
Aandal, Gunhild ;
Nadig, Vidya ;
Yeh, Victoria ;
Rajiah, Prabhakar ;
Jenkins, Trevor ;
Sattar, Abdus ;
Griswold, Mark ;
Gulani, Vikas ;
Gilkeson, Robert C. ;
Seiberlich, Nicole .
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2014, 16
[2]   Application of TWIST MR angiography to aid successful central venous access in challenging patients: initial single-centre experience [J].
Armstrong, L. ;
Rodrigues, J. C. L. ;
Lawton, C. S. ;
Tyrell-Price, J. ;
Hamilton, M. C. K. ;
Manghat, N. E. .
CLINICAL RADIOLOGY, 2016, 71 (11) :1104-1112
[3]   Accelerated spectroscopic imaging of hyperpolarized C-13 pyruvate using SENSE parallel imaging [J].
Arunachalam, Arjun ;
Whitt, David ;
Fish, Kenneth ;
Giaquinto, Randy ;
Piel, Joseph ;
Watkins, Ronald ;
Hancu, Ileana .
NMR IN BIOMEDICINE, 2009, 22 (08) :867-873
[4]   Simultaneous Multislice (SMS) Imaging Techniques [J].
Barth, Markus ;
Breuer, Felix ;
Koopmans, Peter J. ;
Norris, David G. ;
Poser, Benedikt A. .
MAGNETIC RESONANCE IN MEDICINE, 2016, 75 (01) :63-81
[5]   The Effect of Reconstruction and Acquisition Parameters for GRAPPA-Based Parallel Imaging on the Image Quality [J].
Bauer, S. ;
Markl, M. ;
Honal, M. ;
Jung, B. A. .
MAGNETIC RESONANCE IN MEDICINE, 2011, 66 (02) :402-409
[6]  
Bell JB, 1978, Mathematics of Computation, V32, P1320, DOI DOI 10.2307/2006360
[7]   BOLD sensitivity and SNR characteristics of parallel imaging-accelerated single-shot multi-echo EPI for fMRI [J].
Bhavsar, Saurabh ;
Zvyagintsev, Mikhail ;
Mathiak, Klaus .
NEUROIMAGE, 2014, 84 :65-75
[8]   Wave-CAIPI for highly accelerated 3D imaging [J].
Bilgic, Berkin ;
Gagoski, Borjan A. ;
Cauley, Stephen F. ;
Fan, Audrey P. ;
Polimeni, Jonathan R. ;
Grant, P. Ellen ;
Wald, Lawrence L. ;
Setsompop, Kawin .
MAGNETIC RESONANCE IN MEDICINE, 2015, 73 (06) :2152-2162
[9]   2D-GRAPPA-operator for faster 3D parallel MRI [J].
Blaimer, Martin ;
Breuer, Felix A. ;
Mueller, Matthias ;
Seiberlich, Nicole ;
Ebel, Dmitry ;
Heidemann, Robin M. ;
Griswold, Mark A. ;
Jakob, Peter M. .
MAGNETIC RESONANCE IN MEDICINE, 2006, 56 (06) :1359-1364
[10]   Accelerated volumetric MRI with a SENSE/GRAPPA combination [J].
Blaimer, Martin ;
Breuer, Felix A. ;
Seiberlich, Nicole ;
Mueller, Matthias F. ;
Heidemann, Robin M. ;
Jellus, Vladimir ;
Wiggins, Graham ;
Wald, Lawrence L. ;
Griswold, Mark A. ;
Jakob, Peter M. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2006, 24 (02) :444-450