High-Energy X-Ray Imaging Applied to Non Destructive Characterization of Large Nuclear Waste Drums

被引:0
作者
Estre, Nicolas [1 ]
Eck, Daniel [1 ]
Pettier, Jean-Luc [1 ]
Payan, Emmanuel [1 ]
Roure, Christophe [1 ]
Simon, Eric [1 ]
机构
[1] CEN Cadarache, DEN Nucl Measurement Lab, F-13108 St Paul Les Durance, France
来源
2013 3RD INTERNATIONAL CONFERENCE ON ADVANCEMENTS IN NUCLEAR INSTRUMENTATION, MEASUREMENT METHODS AND THEIR APPLICATIONS (ANIMMA) | 2013年
关键词
High-Energy Imaging; nuclear waste drums; linear accelerators; radiography; tomography; scintillators; CdTe; COMPUTED-TOMOGRAPHY; CT;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
As part of its R&D programs on non-destructive testing of nuclear waste drums, CEA is commissioning an irradiation cell named CINPHONIE, at Cadarache. This cell allows high-energy imaging (radiography and tomography) on large volumes (up to 5 m 3) and heavy weights (up to 5 tons). A demonstrator has been finalized, based on existing components. The X-ray source is a 9 MeV LINAC which produces Bremsstrahlung X-rays (up to 23 Gy/min at 1 meter in the beam axis). The mechanical bench is digitally controlled on three axes (translation, rotation, elevation) and can handle objects up to 2 t. This bench performs trajectories necessary for acquisition of projections (sinograms) according to different geometries: Translation-Rotation, Fan-Beam and Cone-Beam. Two detection systems both developed by CEA-Leti are available. The first one is a large GADOX scintillating screen (800x600 mm(2)) coupled to a low-noise pixelated camera. The second one is a multi-CdTe semiconductor detector, offering measurements up to 5 decades of attenuation (equivalent to 25 cm of lead or 180 cm of standard concrete). At the end of the acquisition, a Filtered Back Projection-based algorithm is performed. Then, a density slice (fan-beam tomography) or a density volume (cone-beam tomography or helical tomography) is produced and used to examine the waste. Characterization of LINAC, associated detectors as well as the full acquisition chain, are presented. Experimental performances on phantoms and real drum are discussed and expected limits on defect detectability are evaluated by simulation. The final system, designed to handle objects up to 5 tons is then presented.
引用
收藏
页数:6
相关论文
共 50 条
[31]   Electronic imaging of high energy nanosecond x-ray pulse accelerators [J].
Smith, GW ;
Bell, JD ;
Seymour, CLG .
25TH INTERNATIONAL CONGRESS ON HIGH-SPEED PHOTOGRAPHY AND PHOTONICS, 2003, 4948 :580-591
[32]   Development of soft X-ray stereo imaging system for time-evolution measurement of high-energy electron distribution [J].
Takeda S. ;
Xiang J. ;
Cai Y. ;
Tanabe H. ;
Ono Y. .
IEEJ Transactions on Fundamentals and Materials, 2021, 141 (10) :604-605
[33]   Digital CZT Detector System for High Flux Energy-Resolved X-ray Imaging [J].
Abbene, L. ;
Principato, F. ;
Gerardi, G. ;
Benassi, G. ;
Zambelli, N. ;
Zappettini, A. ;
Bettelli, M. ;
Seller, P. ;
Veale, M. C. .
2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,
[34]   High-energy cryo X-ray nanoimaging at the ID16A beamline of ESRF [J].
da Silva, Julio C. ;
Pacureanu, Alexandra ;
Yang, Yang ;
Fus, Florin ;
Hubert, Maxime ;
Bloch, Leonid ;
Salome, Murielle ;
Bohic, Sylvain ;
Cloetens, Peter .
X-RAY NANOIMAGING: INSTRUMENTS AND METHODS III, 2017, 10389
[36]   A Talbot-Lau X-Ray Deflectometer as a High-Energy Density Plasma Diagnostic [J].
Valdivia, Maria Pia ;
Stutman, Dan ;
Stoeckl, Christian ;
Mileham, Chad S. ;
Begishev, Ildar A. ;
Bromage, Jake ;
Regan, Sean P. .
IEEE TRANSACTIONS ON PLASMA SCIENCE, 2016, 44 (09) :1592-1598
[37]   Quantitative non-destructive evaluation of drilling defects in SiCf/SiC composites using low-energy X-ray imaging technique [J].
Liu, Feifei ;
Liu, Songping ;
Zhang, Qingle ;
Li, Zhiying ;
Qiu, Haipeng .
NDT & E INTERNATIONAL, 2020, 116
[38]   High Energy X-ray Phase-Contrast Imaging Using Glancing Angle Grating Interferometers [J].
Stutman, D. ;
Stayman, J. W. ;
Finkenthal, M. ;
Siewerdsen, J. H. .
MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
[39]   Non-Destructive Imaging Materials Investigation by Microfocus 3D X-Ray Computed Tomography [J].
Lima, I. ;
Assis, J. T. ;
Apoloni, C. R. ;
Mendonca de Souza, S. M. F. ;
Duarte, M. E. L. ;
Lopes, R. T. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (03) :1448-1453
[40]   Large field of view, fast and low dose multimodal phase-contrast imaging at high x-ray energy [J].
Astolfo, Alberto ;
Endrizzi, Marco ;
Vittoria, Fabio A. ;
Diemoz, Paul C. ;
Price, Benjamin ;
Haig, Ian ;
Olivo, Alessandro .
SCIENTIFIC REPORTS, 2017, 7