Reliability of the Respiratory Movement Measuring Instrument, RMMI

被引:14
作者
Olsen, Monika Fagevik [1 ,2 ,3 ]
Romberg, Karin [1 ]
机构
[1] Sahlgrens Univ Hosp, Dept Phys Therapy, SE-41345 Gothenburg, Sweden
[2] Gothenburg Univ, Dept Occupat, Sahlgrenska Acad, Gothenburg, Sweden
[3] Gothenburg Univ, Dept Phys Therapy, Sahlgrenska Acad, Gothenburg, Sweden
关键词
breathing movements; inter-rater; intrarater; reliability; Respiratory Movement Measuring Instrument; PULMONARY-FUNCTION; BREATHING PATTERNS; CARDIAC-SURGERY; CHEST-WALL; CAGE; MOTION;
D O I
10.1111/j.1475-097X.2010.00951.x
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
P>Introduction: A new instrument to measure breathing movements, Respiratory Movement Measuring Instrument (RMMI), has been developed. There is still a lack of knowledge about the reliability of the instrument. Objectives: The aim was to investigate reliability of the RMMI. Methods: In this trial, RMMI (ReMo Inc. Keldnaholt, Reykjavik, Iceland) was used. Inter- and intrarater reliability was performed on 30 volunteers who were tested three times in the supine position. Two of the tests were performed by one of the authors and the third test by the other. Intrasubject reliability was tested on 10 volunteers 12 times randomly during 1 h. Intra-instrument reliability was performed by measuring 12 solid surfaces where the instrument was placed horizontally against a couch or vertically against a chair. Results: Intra-rater reliability: The correlations were moderate to strong (r = 0 center dot 54-0 center dot 94) except for the left lower thoracic position r = 0 center dot 35. Inter-rater reliability: The correlations were strong (r = 0 center dot 71-0 center dot 99) except for the left lower thoracic position r = 0 center dot 35. The strongest correlations were seen on the tests on abdominal breathing movements. Intra subject reliability: The over-all difference within and between the participants and test was non-significant (P = 0 center dot 98). The average difference between the tests was 1 center dot 4 mm. Intra instrument: The differences between the measurements were small (mean 0 center dot 15 mm). The over-all difference was non-significant (P = 0 center dot 79). Conclusion: The RMMI is a reliable instrument and usable in both clinical practice and research.
引用
收藏
页码:349 / 353
页数:5
相关论文
共 23 条
  • [1] Avnon LS, 2009, ISR MED ASSOC J, V11, P83
  • [2] Cahalin LP., 2002, Chest, V122, p207S
  • [3] Domholdt E., 2005, Rehabilitation research: principles and applications, V3rd
  • [4] Elkins MR, 2006, COCHRANE DB SYST REV, DOI [DOI 10.1002/14651858., 10.1002/14651858.CD003147.pub3]
  • [5] CLINICAL APPLICATION OF INDUCTIVE PLETHYSMOGRAPHY
    HANNING, CD
    SMITH, HC
    LEDINGHAM, IM
    [J]. SCOTTISH MEDICAL JOURNAL, 1978, 23 (04) : 310 - 311
  • [6] FACTORS DETERMINING PULMONARY-FUNCTION IN ADOLESCENT IDIOPATHIC THORACIC SCOLIOSIS
    KEARON, C
    VIVIANI, GR
    KIRKLEY, A
    KILLIAN, KJ
    [J]. AMERICAN REVIEW OF RESPIRATORY DISEASE, 1993, 148 (02): : 288 - 294
  • [7] KONDO T, 1989, RESPIRATION, V56, P116
  • [8] KOTANI T, 2002, RES SPINAL DEFORMITI, V4, P135
  • [9] Chest wall motion and pulmonary function are more diminished following cardiac surgery when the internal mammary artery retractor is used
    Kristjánsdóttir, A
    Ragnarsdóttir, M
    Hannesson, P
    Beck, HJ
    Torfason, B
    [J]. SCANDINAVIAN CARDIOVASCULAR JOURNAL, 2004, 38 (06) : 369 - 374
  • [10] Respiratory movements are altered three months and one year following cardiac surgery
    Kristjánsdóttir, A
    Ragnarsdóttir, M
    Hannesson, P
    Beck, HJ
    Torfason, B
    [J]. SCANDINAVIAN CARDIOVASCULAR JOURNAL, 2004, 38 (02) : 98 - 103