High-throughput flow cytometry: Validation in microvolume bioassays

被引:44
作者
Ramirez, S [1 ]
Aiken, CT [1 ]
Andrzejewski, B [1 ]
Sklar, LA [1 ]
Edwards, BS [1 ]
机构
[1] Univ New Mexico, Hlth Sci Ctr, Dept Pathol, Canc Res & Treatment Ctr, Albuquerque, NM 87131 USA
关键词
flow cytometry; automation; drug discovery; biomolecular screening; SELECTIN GLYCOPROTEIN LIGAND-1; P-SELECTIN; DRUG DISCOVERY;
D O I
10.1002/cyto.a.10035
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Background: We recently reported an automated sample handling system, designated HyperCyt, by which samples are aspirated from microplate wells and delivered to the flow cytometer for analysis at rates approaching 100 samples per minute. In this approach, an autosampler and peristaltic pump introduce samples into a tubing line that directly connects to the flow cytometer. Air bubbles are inserted between samples to prevent sample dispersion. In the present work, we compare results of HyperCyt with those of conventional manual flow cytometric analysis in representative flow cytometric bioassays and describe a cell suspension method in which HyperCyt exploits the use of microvolume wells. Methods: Human eosinophils and neutrophils were treated with trypsin to generate a wide (>25-fold) range of membrane P-selectin glycoprotein ligand-1 (PSGL-1) expression and then stained with fluorescent anti-PSGL-1 antibodies. Human peripheral blood mononuclear cells were stained with fluorescein isothiocyanate- and phycoerythrin-conjugated monoclonal antibodies for multiparameter immunophenotype analysis. U937 cells labeled with PKH62GL were used to assess cell settling in microplate wells. Results: Differences in PSGL-1 expression levels were detected by HyperCyt autosampling of leukocytes from 96-well plates at an analysis rate of approximately 1.5 s/well. Hyper-Cyt measurements linearly correlated with parallel manual measurements (r(2) = 0.98). Lymphocyte subpopulations were accurately distinguished and reproducibly quantified in multiparameter immunophenotyping assays performed over a range of HyperCyt analysis rates (1.4-5.5 s/sample). When assay volumes were reduced to 10 mul/well in 60-well Terasaki plates, cells could be maintained in uniform suspension for up to 30 min by periodically inverting plates on a rotating carousel before HyperCyt analysis. HyperCyt analysis of five fluorescence-level Cyto-Plex beads sampled from Temsaki plate microwells at 2.5 s/well produced highly reproducible results over a wide range of input bead concentrations (from 7 X 10(5) to 20 X 10(6) beads/ml) that linearly correlated with manual analysis results. Conclusions: The HyperCyt autosampling system enabled a 10-fold or greater increase in sample thoughput compared with conventional manual flow cytometric sample analysis, with comparable analysis results. Assays were performed efficiently in 10-mul volumes to enable significant reagent cost savings, use of quantity-limited reagents at otherwise prohibitive concentrations, and maintenance of uniform suspensions of cells for prolonged periods. Cytometry Part A 53A:55-65, 2003. (C) 2003 Wiley-Liss, Inc.
引用
收藏
页码:55 / 65
页数:11
相关论文
共 50 条
  • [41] Development of a Human Whole Blood Screening Platform to Monitor JAK/STAT Signaling Using High-Throughput Flow Cytometry
    Fereshteh, Mark P.
    Li, Xin
    Li, Sha
    Fan, Yi
    Zhang, Rosemary
    Farr, Glen A.
    Kolodin, Garrett
    Lippy, Jonathan
    Naglich, Joseph G.
    Schieven, Gary
    Schweizer, Liang
    Zhang, Litao
    JOURNAL OF BIOMOLECULAR SCREENING, 2016, 21 (08) : 866 - 874
  • [42] A high-throughput flow cytometry system for early screening of in vitro made polyploids in Dendrobium hybrids
    Grosso, V.
    Farina, A.
    Giorgi, D.
    Nardi, L.
    Diretto, G.
    Lucretti, S.
    PLANT CELL TISSUE AND ORGAN CULTURE, 2018, 132 (01) : 57 - 70
  • [43] High-throughput vibrational cytometry based on nonlinear Raman microspectroscopy
    Arora, R.
    Petrov, G. I.
    Yakovlev, V. V.
    IMAGING, MANIPULATION, AND ANALYSIS OF BIOMOLECULES, CELLS, AND TISSUES VIII, 2010, 7568
  • [44] Miniaturized High-Throughput Multiparameter Flow Cytometry Assays Measuring In Vitro Human Dendritic Cell Maturation and T-Cell Activation in Mixed Lymphocyte Reactions
    Fan, Yi
    Naglich, Joseph G.
    Koenitzer, Jennifer D.
    Ribeiro, Humberto
    Lippy, Jonathan
    Blum, Jordan
    Li, Xin
    Milburn, Christina
    Barnhart, Bryan
    Zhang, Litao
    Fereshteh, Mark P.
    SLAS DISCOVERY, 2018, 23 (07) : 742 - 750
  • [45] CE-LIF coupled with flow cytometry for high-throughput quantitation of fluorophores in single intact cells
    Xiao, Hua
    Li, Xin
    Zou, Hanfa
    Yang, Ling
    Wang, Yulin
    Wang, Hailin
    Le, X. Chris
    ELECTROPHORESIS, 2006, 27 (17) : 3452 - 3459
  • [46] Identification of neutrophil surface marker changes in health and inflammation using high-throughput screening flow cytometry
    Lakschevitz, Flavia S.
    Hassanpour, Siavash
    Rubin, Ayala
    Fine, Noah
    Sun, Chunxiang
    Glogauer, Michael
    EXPERIMENTAL CELL RESEARCH, 2016, 342 (02) : 200 - 209
  • [47] High-Throughput Isolation of Giant Viruses in Liquid Medium Using Automated Flow Cytometry and Fluorescence Staining
    Khalil, Jacques Y. B.
    Robert, Stephane
    Reteno, Dorine G.
    Andreani, Julien
    Raoult, Didier
    La Scola, Bernard
    FRONTIERS IN MICROBIOLOGY, 2016, 7
  • [48] Assessment of Microbial Community Dynamics in River Bank Filtrate Using High-Throughput Sequencing and Flow Cytometry
    Fiedler, Christina J.
    Schoenher, Christoph
    Proksch, Philipp
    Kerschbaumer, David Johannes
    Mayr, Ernest
    Zunabovic-Pichler, Marija
    Domig, Konrad J.
    Perfler, Reinhard
    FRONTIERS IN MICROBIOLOGY, 2018, 9
  • [49] Duplex High-Throughput Flow Cytometry Screen Identifies Two Novel Formylpeptide Receptor Family Probes
    Young, Susan M.
    Bologa, Cristian M.
    Fara, Dan
    Bryant, Bj K.
    Strouse, Juan Jacob
    Arterburn, Jeffrey B.
    Ye, Richard D.
    Oprea, Tudor I.
    Prossnitz, Eric R.
    Sklar, Larry A.
    Edwards, Bruce S.
    CYTOMETRY PART A, 2009, 75A (03) : 253 - 263
  • [50] Development and Validation of an Automated High-Throughput System for Zebrafish In Vivo Screenings
    Letamendia, Ainhoa
    Quevedo, Celia
    Ibarbia, Izaskun
    Virto, Juan M.
    Holgado, Olaia
    Diez, Maria
    Izpisua Belmonte, Juan Carlos
    Callol-Massot, Carles
    PLOS ONE, 2012, 7 (05): : e36690