Prediction of aortoiliac stent graft length: Comparison of a semiautomated computed tomography angiography method and calibrated aortography

被引:16
作者
Coenegrachts, K
Rigauts, H
De Letter, J
机构
[1] AZ St Jan AV, Dept Radiol, B-8000 Brugge, Belgium
[2] AZ St Jan AV, Dept Vasc Surg, Brugge, Belgium
关键词
stent graft; length measurement; computerized tomographic angiography; calibrated aortography;
D O I
10.1097/00004728-200303000-00031
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The aim of this study is to compare multislice computed tomography (MSCT) in combination with a newly developed semiautomated software program with calibrated aortography in patients who are scheduled for endovascular aortic stent graft placement. Methods: From November 2000 until December 2001, seven patients with an abdominal aortic aneurysm (AAA) underwent both calibrated aortography and MSCT for preoperative endovascular stent graft planning. Both studies were performed within 14 days. Further, length measurements were performed with a semiautomated computerized tomographic angiography (CTA) calibration method and a conventional calibrated aortography technique using three differently configured tubes with variable tortuosity. The AAA length measurements of the semiautomated CTA calibration method and the calibrated aortography were compared. Results: Statistical analysis included linear regression analysis and revealed a probability value of 0.000381 and an r(2) value of 0.93. Using phantoms, it is proven by the authors that the accuracy of the semiautomated CTA calibration method increases with increasing tortuosity when compared with the conventional calibrated aortography technique. Conclusions: Our preliminary results show that the semiautomated CTA calibration method has a potentially advantageous role in preoperative stent graft planning regarding the aortic length measurements and seems to be more accurate than calibrated aortography, especially in extremely tortuous vessels. Further studies have to be performed, however.
引用
收藏
页码:284 / 288
页数:5
相关论文
共 15 条
[1]   Abdominal aortic aneurysm measurements for endovascular repair: Intra- and interobserver variability of CT measurements [J].
Aarts, NJM ;
Schurink, GWH ;
Kool, LJS ;
Bode, PJ ;
van Baalen, JM ;
Hermans, J ;
van Bockel, JH .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 1999, 18 (06) :475-480
[2]  
Broeders IAMJ, 1997, J ENDOVASC SURG, V4, P252, DOI 10.1583/1074-6218(1997)004<0252:PSOGFT>2.0.CO
[3]  
2
[4]   Abdominal aortic aneurysm in high-risk patients: Short- to intermediate-term results of endovascular repair [J].
Chuter, TAM ;
Gordon, RL ;
Reilly, LM ;
Kerlan, RK ;
Sawhney, R ;
Jean-Claude, J ;
Canto, CJ ;
LaBerge, JM ;
Ring, EJ ;
Wall, SD ;
Messina, LM .
RADIOLOGY, 1999, 210 (02) :361-365
[5]  
Ichikawa Kazuo, 2001, Journal of Nippon Medical School, V68, P498, DOI 10.1272/jnms.68.498
[6]  
Kritpracha B, 2002, J ENDOVASC THER, V9, P103, DOI 10.1583/1545-1550(2002)009<0103:CAOTPA>2.0.CO
[7]  
2
[8]   The value of intraoperative intravascular ultrasound for determining stent graft size (excluding abdominal aortic aneurysm) with a modular system [J].
Nolthenius, RPT ;
van den Berg, JC ;
Moll, FL .
ANNALS OF VASCULAR SURGERY, 2000, 14 (04) :311-317
[9]  
Resch T, 1999, J ENDOVASC SURG, V6, P227, DOI 10.1583/1074-6218(1999)006<0227:AAAMIC>2.0.CO
[10]  
2