A dynamic collimation system for penumbra reduction in spot-scanning proton therapy: Proof of concept

被引:68
|
作者
Hyer, Daniel E. [1 ]
Hill, Patrick M. [1 ]
Wang, Dongxu [1 ]
Smith, Blake R. [1 ]
Flynn, Ryan T. [1 ]
机构
[1] Univ Iowa, Dept Radiat Oncol, Iowa City, IA 52242 USA
关键词
intensity modulated proton therapy (IMPT); spot scanning; protons; penumbra; collimation; RADIOTHERAPY; BEAM;
D O I
10.1118/1.4837155
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In the absence of a collimation system the lateral penumbra of spot scanning (SS) dose distributions delivered by low energy proton beams is highly dependent on the spot size. For current commercial equipment, spot size increases with decreasing proton energy thereby reducing the benefit of the SS technique. This paper presents a dynamic collimation system (DCS) for sharpening the lateral penumbra of proton therapy dose distributions delivered by SS. Methods: The collimation system presented here exploits the property that a proton pencil beam used for SS requires collimation only when it is near the target edge, enabling the use of trimmers that are in motion at times when the pencil beam is away from the target edge. The device consists of two pairs of parallel nickel trimmer blades of 2 cm thickness and dimensions of 2 cm x 18 cm in the beam's eye view. The two pairs of trimmer blades are rotated 90 degrees relative to each other to form a rectangular shape. Each trimmer blade is capable of rapid motion in the direction perpendicular to the central beam axis by means of a linear motor, with maximum velocity and acceleration of 2.5 m/s and 19.6m/s(2), respectively. The blades travel on curved tracks to match the divergence of the proton source. An algorithm for selecting blade positions is developed to minimize the dose delivered outside of the target, and treatment plans are created both with and without the DCS. Results: The snout of the DCS has outer dimensions of 22.6 x 22.6 cm(2) and is capable of delivering a minimum treatment field size of 15 x 15 cm(2). Using currently available components, the constructed system would weigh less than 20 kg. For irregularly shaped fields, the use of the DCS reduces the mean dose outside of a 2D target of 46.6 cm(2) by approximately 40% as compared to an identical plan without collimation. The use of the DCS increased treatment time by 1-3 s per energy layer. Conclusions: The spread of the lateral penumbra of low-energy SS proton treatments may be greatly reduced with the use of this system at the cost of only a small penalty in delivery time. (C) 2014 American Association of Physicists in Medicine.
引用
收藏
页数:9
相关论文
共 50 条
  • [31] Lung Stereotactic Body Radiotherapy (SBRT) Using Spot-Scanning Proton Arc (SPArc) Therapy: A Feasibility Study
    Liu, Gang
    Zhao, Lewei
    Qin, An
    Grills, Inga
    Deraniyagala, Rohan
    Stevens, Craig
    Zhang, Sheng
    Yan, Di
    Li, Xiaoqiang
    Ding, Xuanfeng
    FRONTIERS IN ONCOLOGY, 2021, 11
  • [32] Spot-scanning hadron arc (SHArc) therapy: A proof of concept using single- and multi-ion strategies with helium, carbon, oxygen, and neon ions
    Mein, Stewart
    Kopp, Benedikt
    Tessonnier, Thomas
    Liermann, Jakob
    Abdollahi, Amir
    Debus, Juergen
    Haberer, Thomas
    Mairani, Andrea
    MEDICAL PHYSICS, 2022, 49 (09) : 6082 - 6097
  • [33] Clinical implementation of respiratory-gated spot-scanning proton therapy: An efficiency analysis of active motion management
    Gelover, Edgar
    Deisher, Amanda J.
    Herman, Michael G.
    Johnson, Jedediah E.
    Kruse, Jon J.
    Tryggestad, Erik J.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (05): : 99 - 108
  • [34] Investigating aperture-based approximations to model a focused dynamic collimation system for pencil beam scanning proton therapy
    Nelson, Nicholas P.
    Culberson, Wesley S.
    Hyer, Daniel E.
    Smith, Blake R.
    Flynn, Ryan T.
    Hill, Patrick M.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2022, 8 (02)
  • [35] Implementation of planar proton minibeam radiation therapy using a pencil beam scanning system: A proof of concept study
    De Marzi, Ludovic
    Patriarca, Annalisa
    Nauraye, Catherine
    Hierso, Eric
    Dendale, Remi
    Guardiola, Consuelo
    Prezado, Yolanda
    MEDICAL PHYSICS, 2018, 45 (11) : 5305 - 5316
  • [36] Postoperative spot-scanning proton radiation therapy for chordoma and chondrosarcoma in children and adolescents: Initial experience at Paul Scherrer Institute
    Rutz, Hans Peter
    Weber, Damien C.
    Goitein, Gudrun
    Ares, Carmen
    Alessandra, Bolsi
    Antony, J. Lomax
    Pedroni, Eros
    Adolf, Coray
    Eugen, B. Hug
    Timmermann, Beate
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (01): : 220 - 225
  • [37] Spot-scanning proton therapy for early breast cancer in free breathing versus deep inspiration breath-hold
    Stick, Line Bjerregaard
    Nielsen, Louise Laerke
    Trinh, Cecilia Bui
    Bahij, Ihsan
    Jensen, Maria Fuglsang
    Kronborg, Camilla Jen- senius Skovhus
    Petersen, Stine Elleberg
    Thai, Linh My Hoang
    Martinsen, May-Lin
    Precht, Helle
    Offersen, Birgitte Vrou
    ACTA ONCOLOGICA, 2024, 63 : 56 - 61
  • [38] Biological effect of dose distortion by fiducial markers in spot-scanning proton therapy with a limited number of fields: A simulation study
    Matsuura, Taeko
    Maeda, Kenichiro
    Sutherland, Kenneth
    Takayanagi, Taisuke
    Shimizu, Shinichi
    Takao, Seishin
    Miyamoto, Naoki
    Nihongi, Hideaki
    Toramatsu, Chie
    Nagamine, Yoshihiko
    Fujimoto, Rintaro
    Suzuki, Ryusuke
    Ishikawa, Masayori
    Umegaki, Kikuo
    Shirato, Hiroki
    MEDICAL PHYSICS, 2012, 39 (09) : 5584 - 5591
  • [39] Comparison of organ-at-risk sparing and plan robustness for spot-scanning proton therapy and volumetric modulated arc photon therapy in head-and-neck cancer
    Barten, Danique L. J.
    Tol, Jim P.
    Dahele, Max
    Slotman, Ben J.
    Verbakel, Wilko F. A. R.
    MEDICAL PHYSICS, 2015, 42 (11) : 6589 - 6598
  • [40] Dynamic gating window technique for the reduction of dosimetric error in respiratory-gated spot-scanning particle therapy: An initial phantom study using patient tumor trajectory data
    Miyamoto, Naoki
    Yokokawa, Kouhei
    Takao, Seishin
    Matsuura, Taeko
    Tanaka, Sodai
    Shimizu, Shinichi
    Shirato, Hiroki
    Umegaki, Kikuo
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (04): : 13 - 21