A dynamic collimation system for penumbra reduction in spot-scanning proton therapy: Proof of concept

被引:68
|
作者
Hyer, Daniel E. [1 ]
Hill, Patrick M. [1 ]
Wang, Dongxu [1 ]
Smith, Blake R. [1 ]
Flynn, Ryan T. [1 ]
机构
[1] Univ Iowa, Dept Radiat Oncol, Iowa City, IA 52242 USA
关键词
intensity modulated proton therapy (IMPT); spot scanning; protons; penumbra; collimation; RADIOTHERAPY; BEAM;
D O I
10.1118/1.4837155
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In the absence of a collimation system the lateral penumbra of spot scanning (SS) dose distributions delivered by low energy proton beams is highly dependent on the spot size. For current commercial equipment, spot size increases with decreasing proton energy thereby reducing the benefit of the SS technique. This paper presents a dynamic collimation system (DCS) for sharpening the lateral penumbra of proton therapy dose distributions delivered by SS. Methods: The collimation system presented here exploits the property that a proton pencil beam used for SS requires collimation only when it is near the target edge, enabling the use of trimmers that are in motion at times when the pencil beam is away from the target edge. The device consists of two pairs of parallel nickel trimmer blades of 2 cm thickness and dimensions of 2 cm x 18 cm in the beam's eye view. The two pairs of trimmer blades are rotated 90 degrees relative to each other to form a rectangular shape. Each trimmer blade is capable of rapid motion in the direction perpendicular to the central beam axis by means of a linear motor, with maximum velocity and acceleration of 2.5 m/s and 19.6m/s(2), respectively. The blades travel on curved tracks to match the divergence of the proton source. An algorithm for selecting blade positions is developed to minimize the dose delivered outside of the target, and treatment plans are created both with and without the DCS. Results: The snout of the DCS has outer dimensions of 22.6 x 22.6 cm(2) and is capable of delivering a minimum treatment field size of 15 x 15 cm(2). Using currently available components, the constructed system would weigh less than 20 kg. For irregularly shaped fields, the use of the DCS reduces the mean dose outside of a 2D target of 46.6 cm(2) by approximately 40% as compared to an identical plan without collimation. The use of the DCS increased treatment time by 1-3 s per energy layer. Conclusions: The spread of the lateral penumbra of low-energy SS proton treatments may be greatly reduced with the use of this system at the cost of only a small penalty in delivery time. (C) 2014 American Association of Physicists in Medicine.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Effects of spot size and spot spacing on lateral penumbra reduction when using a dynamic collimation system for spot scanning proton therapy
    Hyer, Daniel E.
    Hill, Patrick M.
    Wang, Dongxu
    Smith, Blake R.
    Flynn, Ryan T.
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (22) : N187 - N196
  • [2] Evaluation of the relative biological effectiveness of spot-scanning proton irradiation in vitro
    Maeda, Kenichiro
    Yasui, Hironobu
    Matsuura, Taeko
    Yamamori, Tohru
    Suzuki, Motofumi
    Nagane, Masaki
    Nam, Jin-Min
    Inanami, Osamu
    Shirato, Hiroki
    JOURNAL OF RADIATION RESEARCH, 2016, 57 (03) : 307 - 311
  • [3] Physical and biological impacts of collimator-scattered protons in spot-scanning proton therapy
    Ueno, Koki
    Matsuura, Taeko
    Hirayama, Shusuke
    Takao, Seishin
    Ueda, Hideaki
    Matsuo, Yuto
    Yoshimura, Takaaki
    Umegaki, Kikuo
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (07): : 48 - 57
  • [4] The Effects of a Dynamic Collimation System On Proton Pencil Beams to Improve Lateral Tissue Sparing in Spot Scanned Proton Therapy
    Hill, P.
    Wang, D.
    Flynn, R.
    Hyer, D.
    MEDICAL PHYSICS, 2014, 41 (06) : 298 - 298
  • [5] Interplay Effect in Spot-Scanning Proton Therapy with Rescanning, Breath Holding, and Gating: A Phantom Study
    Belikhin, Mikhail
    Shemyakov, Alexander
    Ivanov, Dmitry
    Zavestovskaya, Irina
    APPLIED SCIENCES-BASEL, 2024, 14 (18):
  • [6] Explore the feasibility of using spot-scanning proton arc therapy for a synchrotron accelerator-based proton therapy system - A simulation study
    Cong, Xiaoda
    Liu, Gang
    Liu, Peilin
    Zhao, Lewei
    Chen, Shupeng
    Li, Xiaoqiang
    Shen, Jiajian
    Ding, Xuanfeng
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025, 26 (01):
  • [7] Integration and dosimetric validation of a dynamic collimation system for pencil beam scanning proton therapy
    Nelson, Nicholas P.
    Culberson, Wesley S.
    Hyer, Daniel E.
    Geoghegan, Theodore J.
    Patwardhan, Kaustubh A.
    Smith, Blake R.
    Flynn, Ryan T.
    Gutierrez, Alonso N.
    Boland, Thibault
    Hill, Patrick M.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2023, 9 (06)
  • [8] Toward improved target conformity for two spot scanning proton therapy delivery systems using dynamic collimation
    Moignier, Alexandra
    Gelover, Edgar
    Smith, Blake R.
    Wang, Dongxu
    Flynn, Ryan T.
    Kirk, Maura L.
    Lin, Liyong
    Solberg, Timothy D.
    Lin, Alexander
    Hyer, Daniel E.
    MEDICAL PHYSICS, 2016, 43 (03) : 1421 - 1427
  • [9] Rotationally Intensified Proton Lattice: A Novel Lattice Technique Using Spot-Scanning Proton Arc Therapy
    Lee, Joseph S.
    Mumaw, Derek A.
    Liu, Peilin
    Loving, Bailey A.
    Sebastian, Ebin
    Cong, Xiaoda
    Stefani, Mark S.
    Loughery, Brian F.
    Li, Xiaoqiang
    Deraniyagala, Rohan
    Almahariq, Muayad F.
    Ding, Xuanfeng
    Quinn, Thomas J.
    ADVANCES IN RADIATION ONCOLOGY, 2024, 9 (11)
  • [10] Spot-scanning proton therapy for rhabdomyosarcomas of early childhood. First experiences at PSI
    Timmermann, Beate
    Schuck, Andreas
    Niggli, Felix
    Weiss, Markus
    Lomax, Anthony
    Goitein, Gudrun
    STRAHLENTHERAPIE UND ONKOLOGIE, 2006, 182 (11) : 653 - 659