Zero-Heat-Flux Thermometry for Non-Invasive Measurement of Core Body Temperature in Pigs

被引:15
作者
Guschlbauer, Maria [1 ]
Maul, Alexandra C. [1 ]
Yan, Xiaowei [2 ]
Herff, Holger [2 ]
Annecke, Thorsten [2 ]
Sterner-Kock, Anja [1 ]
Boettiger, Bernd W. [2 ]
Schroeder, Daniel C.
机构
[1] Univ Hosp Cologne, Ctr Med Expt, Cologne, Germany
[2] Univ Hosp Cologne, Dept Anaesthesiol & Intens Care Med, Cologne, Germany
来源
PLOS ONE | 2016年 / 11卷 / 03期
关键词
MILD PERIOPERATIVE HYPOTHERMIA; THERAPEUTIC HYPOTHERMIA; HEMORRHAGIC-SHOCK; WOUND-INFECTION; CARDIAC-ARREST; BLOOD-LOSS; ANESTHESIA; SWINE; MODEL; NORMOTHERMIA;
D O I
10.1371/journal.pone.0150759
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Hypothermia is a severe, unpleasant side effect during general anesthesia. Thus, temperature surveillance is a prerequisite in general anesthesia settings during experimental surgeries. The gold standard to measure the core body temperature (T-core) is placement of a Swan-Ganz catheter in the pulmonary artery, which is a highly invasive procedure. Therefore, T-core is commonly examined in the urine bladder and rectum. However, these procedures are known for their inaccuracy and delayed record of temperatures. Zero-heat-flux (ZHF) thermometry is an alternative, non-invasive method quantifying T-core in human patients by applying a thermosensoric patch to the lateral forehead. Since the porcine cranial anatomy is different to the human's, the optimal location of the patch remains unclear to date. The aim was to compare three different patch locations of ZHF thermometry in a porcine hypothermia model. Hypothermia (33.0 degrees C T-core) was conducted in 11 anesthetized female pigs (26-30kg). T-core was measured continuously by an invasive Swan-Ganz catheter in the pulmonary artery (T-pulm). A ZHF thermometry device was mounted on three different defined locations. The smallest average difference between T-pulm and T-ZHF during stable temperatures was 0.21 +/- 0.16 degrees C at location A, where the patch was placed directly behind the eye. Also during rapidly changing temperatures location A showed the smallest bias with 0.48 +/- 0.29 degrees C. Location A provided the most reliable data for T-core. Therefore, the ZHF thermometry patch should be placed directly behind the left temporal corner of the eye to provide a non-invasive method for accurate measurement of T-core in pigs.
引用
收藏
页数:11
相关论文
共 42 条
  • [1] MEASUREMENT IN MEDICINE - THE ANALYSIS OF METHOD COMPARISON STUDIES
    ALTMAN, DG
    BLAND, JM
    [J]. JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES D-THE STATISTICIAN, 1983, 32 (03) : 307 - 317
  • [2] Efficacy of intraoperative heat administration by ventilation with warm humidified gases and an oesophageal warming system
    Beck, E
    Langer, M
    DiMauro, P
    Prato, P
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1996, 77 (04) : 530 - 533
  • [3] Confounding factors in the use of the zero-heat-flow method for non-invasive muscle temperature measurement
    Brajkovic, D
    Ducharme, M
    [J]. EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2005, 94 (04) : 386 - 391
  • [4] An Evaluation of a Zero-Heat-Flux Cutaneous Thermometer in Cardiac Surgical Patients
    Eshraghi, Yashar
    Nasr, Vivian
    Parra-Sanchez, Ivan
    Van Duren, Albert
    Botham, Mark
    Santoscoy, Thomas
    Sessler, Daniel I.
    [J]. ANESTHESIA AND ANALGESIA, 2014, 119 (03) : 543 - 549
  • [5] NEW METHOD FOR MONITORING DEEP BODY-TEMPERATURE FROM SKIN SURFACE
    FOX, RH
    SOLMAN, AJ
    ISAACS, R
    FRY, AJ
    MACDONALD, IC
    [J]. CLINICAL SCIENCE, 1973, 44 (01) : 81 - 86
  • [6] Understanding Bland Altman analysis
    Giavarina, Davide
    [J]. BIOCHEMIA MEDICA, 2015, 25 (02) : 141 - 151
  • [7] Effect of Mild Hypothermia on the Coagulation-Fibrinolysis System and Physiological Anticoagulants after Cardiopulmonary Resuscitation in a Porcine Model
    Gong, Ping
    Zhang, Ming-Yue
    Zhao, Hong
    Tang, Zi-Ren
    Hua, Rong
    Mei, Xue
    Cui, Juan
    Li, Chun-Sheng
    [J]. PLOS ONE, 2013, 8 (06):
  • [8] Comparison of methods of temperature measurement in swine
    Hanneman, SK
    Jesurum-Urbaitis, JT
    Bickel, DR
    [J]. LABORATORY ANIMALS, 2004, 38 (03) : 297 - 306
  • [9] Intraoperative temperature monitoring with zero heat flux technology (3M SpotOn sensor) in comparison with sublingual and nasopharyngeal temperature An observational study
    Iden, Timo
    Horn, Ernst-Peter
    Bein, Berthold
    Boehm, Ruwen
    Beese, Janne
    Hoecker, Jan
    [J]. EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2015, 32 (06) : 387 - 391
  • [10] BODY-TEMPERATURE AND ANESTHESIA
    IMRIE, MM
    HALL, GM
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1990, 64 (03) : 346 - 354