Development of the chemical exposure monitor with indoor positioning (CEMWIP) for workplace VOC surveys

被引:9
作者
Brown, K. K. [1 ]
Shaw, P. B. [1 ]
Mead, K. R. [1 ]
Kovein, R. J. [1 ]
Voorhees, R. T. [2 ]
Brandes, A. R. [3 ]
机构
[1] NIOSH, Div Appl Res & Technol, 4676 Columbia Pkwy MS-R7, Cincinnati, OH 45226 USA
[2] Univ Cincinnati, Cincinnati, OH USA
[3] MeasureNet Technol Ltd, Cincinnati, OH USA
关键词
3-D hazard map; direct-reading method (DRM); exposure assessment; photo-ionization detector (PID); real-time location system (RTLS); volatile organic compounds (VOCs);
D O I
10.1080/15459624.2015.1125488
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The purpose of this article was to research and develop a direct-reading exposure assessment method that combined a real-time location system with a wireless direct-reading personal chemical sensor. The personal chemical sensor was a photoionization device for detecting volatile organic compounds. The combined system was calibrated and tested against the same four standard gas concentrations and calibrated at one standard location and tested at four locations that included the standard locations. Data were wirelessly collected from the chemical sensor every 1.4 sec, for volatile organic compounds concentration, location, temperature, humidity, and time. Regression analysis of the photo-ionization device voltage response against calibration gases showed the chemical sensor had a limit of detection of 0.2 ppm. The real-time location system was accurate to 13 cm +/- 6 cm( standard deviation) in an open area and to 57 cm +/- 31 cm in a closed room where the radio frequency has to penetrate drywall-finished walls. The streaming data were collected and graphically displayed as a three-dimensional hazard map for assessment of peak exposure with location. A real-time personal exposure assessment device with indoor positioning was practical and provided new knowledge on direct reading exposure assessment methods.
引用
收藏
页码:401 / 412
页数:12
相关论文
共 17 条
[1]   Definition and assessment of sampling and analytical accuracy [J].
Bartley, DL .
ANNALS OF OCCUPATIONAL HYGIENE, 2001, 45 (05) :357-364
[2]  
Boulos M.N.K., REAL TIME LOCATING S
[3]   Environmental and biological monitoring of volatile organic compounds in the workplace [J].
Caro, J. ;
Gallego, M. .
CHEMOSPHERE, 2009, 77 (03) :426-433
[4]   A new sensor for the assessment of personal exposure to volatile organic compounds [J].
Chen, Cheng ;
Campbell, Katherine Driggs ;
Negi, Indira ;
Iglesias, Rodrigo A. ;
Owens, Patrick ;
Tao, Nongjian ;
Tsow, Francis ;
Forzani, Erica S. .
ATMOSPHERIC ENVIRONMENT, 2012, 54 :679-687
[5]   Measurement Capability of Field Portable Organic Vapor Monitoring Instruments Under Different Experimental Conditions [J].
Coffey, Christopher C. ;
Pearce, Terri A. ;
Lawrence, Robert B. ;
Hudnall, Judith B. ;
Slaven, James E. ;
Martin, Stephen B., Jr. .
JOURNAL OF OCCUPATIONAL AND ENVIRONMENTAL HYGIENE, 2009, 6 (01) :1-8
[6]  
Colwell S., 2008, GPS WORLD, P34
[7]  
Driscoll J.N., 2013, IMPORTANT INSTRUMENT, P39
[8]   ELECTROCHEMICAL SENSORS FOR MONITORING ENVIRONMENTAL-POLLUTANTS [J].
FLEET, B ;
GUNASINGHAM, H .
TALANTA, 1992, 39 (11) :1449-1457
[9]  
Howard J., DIRECT READING SENSO
[10]   Time location analysis for exposure assessment studies of indoor workers based on active RFID technology [J].
Huang, Fu-Chuan ;
Shih, Tung-Sheng ;
Lee, Jiunn-Fwu ;
Chao, Huan-Ping ;
Wang, Peng-Yau .
JOURNAL OF ENVIRONMENTAL MONITORING, 2010, 12 (02) :514-523