Cone beam computed tomography based image guidance and quality assessment of prostate cancer for magnetic resonance imaging-only radiotherapy in the pelvis

被引:4
|
作者
Edmund, Jens M. [1 ,2 ]
Andreasen, Daniel [3 ]
Van Leemput, Koen [3 ,4 ]
机构
[1] Univ Copenhagen, Gentofte & Herlev Hosp, Dept Oncol, Radiotherapy Res Unit, DK-2730 Herlev, Denmark
[2] Univ Copenhagen, Niels Bohr Inst, DK-2100 Copenhagen, Denmark
[3] Tech Univ Denmark, Dept Hlth Technol, DK-2800 Lyngby, Denmark
[4] Harvard Med Sch, Massachusetts Gen Hosp, Dept Radiol, Martinos Ctr Biomed Imaging, Boston, MA 02114 USA
来源
PHYSICS & IMAGING IN RADIATION ONCOLOGY | 2021年 / 18卷
关键词
MRI-only RT; IGRT; Synthetic CT; Pseudo CT; Magnetic resonance imaging; Radiotherapy; Cone beam CT; SYNTHETIC CT GENERATION; RADIATION-THERAPY; NEURAL-NETWORK; MRI SEQUENCES; PSEUDO-CT;
D O I
10.1016/j.phro.2021.05.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Radiotherapy (RT) based on magentic resonance imaging (MRI) only is currently used clinically in the pelvis. A synthetic computed tomography (sCT) is needed for dose planning. Here, we investigate the accuracy of cone beam CT (CBCT) based MRI-only image guided RT (IGRT) and sCT image quality. Materials and methods: CT, MRI and CBCT scans of ten prostate cancer patients were included. The MRI was converted to a sCT using a multi-atlas approach. The sCT, CT and MR images were auto-matched with the CBCT on the bony anatomy. Paired sCT-CT and sCT-CBCT data were created. CT numbers were converted to relative electron (RED) and mass densities (DES) using a standard calibration curve for the CT and sCT. For the CBCT RED/DES conversion, a phantom and paired CT-CBCT population based calibration curve was used. For the latter, the CBCT numbers were averaged in 100 HU bins and the known RED/DES of the CT were assigned. The paired sCT-CT and sCT-CBCT data were averaged in bins of 10 HU or 0.01 RED/DES. The median absolute error (MeAE) between the sCT-CT and sCT-CBCT bins was calculated. Wilcoxon rank-sum tests were carried out for the IGRT and MeAE study. Results: The mean sCT or MR IGRT difference from CT was <= 2 mm but significant differences were observed. A CBCT HU or phantom-based RED/DES MeAE did not estimate the sCT quality similar to a CT based MeAE but the CBCT population-based RED/DES MeAE did. Conclusions: MRI-only CBCT-based IGRT seems feasible but caution is advised. A MeAE around 0.1 DES could call for sCT quality inspection.
引用
收藏
页码:55 / 60
页数:6
相关论文
共 50 条
  • [1] Cone beam computed tomography image guidance within a magnetic resonance imaging-only planning workflow
    O'Connor, Laura M.
    Quinn, Alesha
    Denley, Samuel
    Leigh, Lucy
    Martin, Jarad
    Dowling, Jason A.
    Skehan, Kate
    Warren-Forward, Helen
    Greer, Peter B.
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2023, 27
  • [2] Cone beam computed tomography for dose calculation quality assurance for magnetic resonance-only radiotherapy
    Wyatt, Jonathan J.
    Pearson, Rachel A.
    Walker, Christopher P.
    Brooks, Rachel L.
    Pilling, Karen
    McCallum, Hazel M.
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2021, 17 : 71 - 76
  • [3] Cone beam computed tomography guided treatment delivery and planning verification for magnetic resonance imaging only radiotherapy of the brain
    Edmund, Jens M.
    Andreasen, Daniel
    Mahmood, Faisal
    Van Leemput, Koen
    ACTA ONCOLOGICA, 2015, 54 (09) : 1496 - 1500
  • [4] Feasibility of magnetic resonance imaging-only rectum radiotherapy with a commercial synthetic computed tomography generation solution
    Maspero, Matteo
    Tyyger, Marcus D.
    Tijssen, Rob H. N.
    Seevinck, Peter R.
    Intven, Martijn P. W.
    van den Berg, Cornelis A. T.
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2018, 7 : 58 - 64
  • [5] The accuracy of Magnetic Resonance - Cone Beam Computed Tomography soft-tissue matching for prostate radiotherapy
    Wyatt, Jonathan J.
    Brooks, Rachel L.
    Ainslie, Dean
    Wilkins, Emily
    Raven, Elizabeth
    Pilling, Karen
    Pearson, Rachel A.
    McCallum, Hazel M.
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2019, 12 : 49 - 55
  • [6] Assessment of dosimetric and positioning accuracy of a magnetic resonance imaging-only solution for external beam radiotherapy of pelvic anatomy
    Kemppainen, Reko
    Suilamo, Sami
    Ranta, Iiro
    Pesola, Marko
    Halkola, Aleksi
    Eufemio, Alvin
    Minn, Heikki
    Keyrilainen, Jani
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2019, 11 : 1 - 8
  • [7] LIPIODOL: A POTENTIAL DIRECT SURROGATE FOR CONE-BEAM COMPUTED TOMOGRAPHY IMAGE GUIDANCE IN RADIOTHERAPY OF LIVER TUMOR
    Yue, Jinbo
    Sun, Xindong
    Cai, Jing
    Yin, Fang-Fang
    Yin, Yong
    Zhu, Jian
    Lu, Jie
    Liu, Tonghai
    Yu, Jinming
    Shi, Xuetao
    Song, Jinlong
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 82 (02): : 834 - 841
  • [8] IMAGE-GUIDED RADIOTHERAPY (IGRT) FOR PROSTATE CANCER COMPARING kV IMAGING OF FIDUCIAL MARKERS WITH CONE BEAM COMPUTED TOMOGRAPHY (CBCT)
    Barney, Brandon M.
    Lee, R. Jeffrey
    Handrahan, Diana
    Welsh, Keith T.
    Cook, J. Taylor
    Sause, William T.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 80 (01): : 301 - 305
  • [9] Results of 2023 survey on the use of synthetic computed tomography for magnetic resonance Imaging-only radiotherapy: Current status and future steps
    Fusella, M.
    Andres, E. Alvarez
    Villegas, F.
    Milan, L.
    Janssen, T. M.
    Dal Bello, R.
    Garibaldi, C.
    Placidi, L.
    Cusumano, D.
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2024, 32
  • [10] Evaluation of the Geometric and Dosimetric Accuracy of Synthetic Computed Tomography Images for Magnetic Resonance Imaging-only Stereotactic Radiosurgery
    Fatemi, Ali
    Kanakamedala, Madhava R.
    Yang, Claus Chunli
    Morris, Bart
    Duggar, William N.
    Vijayakumar, Srinivasan
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2019, 11 (04)