Assessing the Accuracy of Eyelid Measurements Utilizing the Volk Eye Check System and Clinical Measurements

被引:18
|
作者
Sinha, Kunal R. [1 ,2 ]
Yeganeh, Amir [2 ,3 ]
Goldberg, Robert A. [2 ]
Rootman, Daniel B. [2 ,4 ]
机构
[1] Univ Calif Los Angeles, David Geffen Sch Med, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, Stein Eye Inst, Div Orbital & Ophthalm Plast Surg, Los Angeles, CA USA
[3] Northeast Ohio Med Univ, Rootstown, OH USA
[4] Univ Calif Los Angeles, Doheny Eye Ctr, Div Orbital & Ophthalm Plast Surg, Los Angeles, CA USA
来源
OPHTHALMIC PLASTIC AND RECONSTRUCTIVE SURGERY | 2018年 / 34卷 / 04期
关键词
RESECTION; POSITION;
D O I
10.1097/IOP.0000000000000991
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: The purpose of this study was to validate the accuracy of marginal reflex distance 1 (MRD1) measurements obtained by the Volk Eye Check system, a modified smartphone that measures MRD1 automatically, relative to clinical and digital measurements. Methods: In this prospective observational study of adults with normal eyelids and ptosis, MRD1 was measured clinically, digitally, and automatically with the Volk device. Eyes were divided into successful versus unsuccessful Volk trial groups; successful eyes were then subdivided into control and ptosis subgroups. The primary outcome measures were mean MRD1 obtained by the 3 modalities. Secondary outcome measures included the success rate of the device and the prevalence of ptosis within the successful and unsuccessful groups. Results: In the overall sample of 88 eyes, clinical and digital MRD1 were not significantly different. Among eyes with successful Volk trials, significant differences in MRD1 measured by the 3 modalities were as follows: in the successful group, Volk MRD1 (3.05mm) was significantly (p < 0.01) higher than digital MRD1 (2.68mm); in the ptosis subgroup, Volk MRD1 (2.47mm) was significantly higher than clinical (2.05mm; p < 0.05) and digital MRD1 (1.91mm; p < 0.01). Eyes in the successful group (66% of attempted collections) demonstrated significantly higher MRD1 (mean difference, 1.21mm; p < 0.01) and were significantly less likely to demonstrate ptosis (2.7x; p < 0.05) than those in the unsuccessful group. Conclusions: The Volk device measures MRD1 well in normal patients but overestimates MRD1 in patients with ptosis. It may be most appropriate in assessing patients with normal or elevated eyelid position. Clinical and digital MRD1 measurements were not different than each other.
引用
收藏
页码:346 / 350
页数:5
相关论文
共 50 条
  • [41] CENTRAL METERING SYSTEM INCREASES ACCURACY OF PIPELINE MEASUREMENTS
    THEODULE, P
    CULOT, P
    OIL & GAS JOURNAL, 1978, 76 (17) : 91 - 93
  • [42] Development of a Neutron Spectroscopic System Utilizing Compressed Sensing Measurements
    Vargas, Danilo
    Kurwitz, R. Cable
    Carron, Igor
    DePriest, K. Russell
    ISRD 15 - INTERNATIONAL SYMPOSIUM ON REACTOR DOSIMETRY, 2016, 106
  • [43] Clinical Accuracy of Time-Delayed Laboratory Glucose Measurements
    Cheng, Royce
    Dunn, Timothy C.
    Hayter, Gary
    DIABETES, 2011, 60 : A582 - A583
  • [44] IS THE ACCURACY OF SERUM ALBUMIN MEASUREMENTS SUITABLE FOR CLINICAL APPLICATION OF THE TEST?
    Infusino, I.
    Braga, F.
    Mozzi, R.
    Valente, C.
    Panteghini, M.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S805 - S805
  • [45] Is the accuracy of serum albumin measurements suitable for clinical application of the test?
    Infusino, Ilenia
    Braga, Federica
    Mozzi, Roberta
    Valente, Cristina
    Panteghini, Mauro
    CLINICA CHIMICA ACTA, 2011, 412 (9-10) : 791 - 792
  • [46] Accuracy of volume measurements by a clinical spirometer in multiple veterinary hospitals
    Raillard, Mathieu
    Mosing, Martina
    Raisis, Anthea
    Auckburally, Adam
    Borland, Karla
    Canfran, Susana
    Downing, Frances
    Mateo, Alejandra Garcia de Carellan
    Macfarlane, Paul
    Mcfadzean, William
    Merlin, Tristan
    Portier, Karine
    Robertson, Josephine
    Soares, Joao Henrique Neves
    Steblaj, Barbara
    Zoff, Aurora
    Levionnois, Olivier L.
    FRONTIERS IN VETERINARY SCIENCE, 2024, 11
  • [47] Continuous ambulatory diaphragm emg measurements: Assessing respiratory control and function utilizing the Diaphragm Pacing Stimulation (DPS) system
    Onders, Raymond P.
    Katirji, Bashar
    Schilz, Robert
    Nemunaitis, Greg
    Elmo, MaryJo
    Ignagni, Anthony
    CHEST, 2007, 132 (04) : 649S - 650S
  • [48] ASSESSING AND REPORTING THE ACCURACY OF POSITION MEASUREMENTS MADE WITH OPTICAL TRACKING-SYSTEMS
    HAGGARD, P
    WING, AM
    JOURNAL OF MOTOR BEHAVIOR, 1990, 22 (02) : 315 - 321
  • [49] Measurement Methods and Procedures for Assessing Accuracy of Instrument Transformers for Power Quality Measurements
    Crotti, G.
    van den Brom, H. E.
    Mohns, E.
    Tinarelli, R.
    Luiso, M.
    Styblikova, R.
    Agazar, M.
    Cayci, H.
    Mazza, P.
    Meyer, J.
    Almutairi, M.
    2020 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM), 2020,
  • [50] Assessing the accuracy of Greenland ice sheet ice ablation measurements by pressure transducer
    Fausto, Robert S.
    Van As, Dirk
    Ahlstrom, Andreas P.
    Citterio, Michele
    JOURNAL OF GLACIOLOGY, 2012, 58 (212) : 1144 - 1150