Assessing the Accuracy of Eyelid Measurements Utilizing the Volk Eye Check System and Clinical Measurements

被引:18
|
作者
Sinha, Kunal R. [1 ,2 ]
Yeganeh, Amir [2 ,3 ]
Goldberg, Robert A. [2 ]
Rootman, Daniel B. [2 ,4 ]
机构
[1] Univ Calif Los Angeles, David Geffen Sch Med, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, Stein Eye Inst, Div Orbital & Ophthalm Plast Surg, Los Angeles, CA USA
[3] Northeast Ohio Med Univ, Rootstown, OH USA
[4] Univ Calif Los Angeles, Doheny Eye Ctr, Div Orbital & Ophthalm Plast Surg, Los Angeles, CA USA
来源
OPHTHALMIC PLASTIC AND RECONSTRUCTIVE SURGERY | 2018年 / 34卷 / 04期
关键词
RESECTION; POSITION;
D O I
10.1097/IOP.0000000000000991
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: The purpose of this study was to validate the accuracy of marginal reflex distance 1 (MRD1) measurements obtained by the Volk Eye Check system, a modified smartphone that measures MRD1 automatically, relative to clinical and digital measurements. Methods: In this prospective observational study of adults with normal eyelids and ptosis, MRD1 was measured clinically, digitally, and automatically with the Volk device. Eyes were divided into successful versus unsuccessful Volk trial groups; successful eyes were then subdivided into control and ptosis subgroups. The primary outcome measures were mean MRD1 obtained by the 3 modalities. Secondary outcome measures included the success rate of the device and the prevalence of ptosis within the successful and unsuccessful groups. Results: In the overall sample of 88 eyes, clinical and digital MRD1 were not significantly different. Among eyes with successful Volk trials, significant differences in MRD1 measured by the 3 modalities were as follows: in the successful group, Volk MRD1 (3.05mm) was significantly (p < 0.01) higher than digital MRD1 (2.68mm); in the ptosis subgroup, Volk MRD1 (2.47mm) was significantly higher than clinical (2.05mm; p < 0.05) and digital MRD1 (1.91mm; p < 0.01). Eyes in the successful group (66% of attempted collections) demonstrated significantly higher MRD1 (mean difference, 1.21mm; p < 0.01) and were significantly less likely to demonstrate ptosis (2.7x; p < 0.05) than those in the unsuccessful group. Conclusions: The Volk device measures MRD1 well in normal patients but overestimates MRD1 in patients with ptosis. It may be most appropriate in assessing patients with normal or elevated eyelid position. Clinical and digital MRD1 measurements were not different than each other.
引用
收藏
页码:346 / 350
页数:5
相关论文
共 50 条
  • [1] Volk eye check ocular measurement device for objective periorbital measurements: A prospective study
    Naik, Milind N.
    Tiwari, Arundhati
    Gattu, Naresh
    Ali, Mohammad Hasnat
    JOURNAL OF COSMETIC DERMATOLOGY, 2022, 21 (04) : 1582 - 1587
  • [2] Accuracy of Marginal Reflex Distance Measurements in Eyelid Surgery
    Nemet, Arie Y.
    JOURNAL OF CRANIOFACIAL SURGERY, 2015, 26 (07) : E569 - E571
  • [3] LINAC Commissioning Measurements Utilizing the Cylindrical Arc Check Phantom
    Aristophanous, M.
    Balter, P.
    Martel, M.
    Murray, N.
    Court, L.
    MEDICAL PHYSICS, 2012, 39 (06) : 3745 - 3745
  • [4] Utilizing CMM measurements in the evaluation of machine tool accuracy
    Holsa, J
    LASER METROLOGY AND MACHINE PERFORMANCE III, 1997, : 487 - 496
  • [5] The repeatability of clinical measurements of dry eye
    Nichols, KK
    Mitchell, GL
    Zadnik, K
    CORNEA, 2004, 23 (03) : 272 - 285
  • [6] ACCURACY OF CLINICAL TOOTH MOBILITY MEASUREMENTS
    SELIPSKY, H
    ERICKSON, JD
    JOURNAL OF DENTAL RESEARCH, 1975, 54 : L65 - L65
  • [7] ACCURACY OF CLINICAL TOOTH MOBILITY MEASUREMENTS
    SELIPSKY, H
    ERICKSON, JD
    JOURNAL OF PERIODONTOLOGY, 1975, 46 (10) : 631 - 631
  • [8] Frequency of IOP measurements and eye check-ups in elderly Finns
    Vesti, Eija
    Nevalainen, Tanja
    Jaatinen, Jarkko
    Pitkanen, Leena
    ACTA OPHTHALMOLOGICA, 2017, 95 (04) : E346 - E347
  • [9] Assessing the Accuracy of Tree Diameter Measurements Collected at a Distance
    Weaver, Steven A.
    Ucar, Zennure
    Bettinger, Pete
    Merry, Krista
    Faw, Krisha
    Cieszewski, Chris J.
    CROATIAN JOURNAL OF FOREST ENGINEERING, 2015, 36 (01) : 73 - 83
  • [10] Assessing the Accuracy of SWOT Measurements of Water Bodies in Australia
    Maubant, Louise
    Dodd, Lachlan
    Tregoning, Paul
    GEOPHYSICAL RESEARCH LETTERS, 2025, 52 (06)