Metal artifact reduction in CT, a phantom study: subjective and objective evaluation of four commercial metal artifact reduction algorithms when used on three different orthopedic metal implants

被引:54
作者
Bolstad, Kirsten [1 ]
Flatabo, Silje [1 ]
Aadnevik, Daniel [1 ]
Dalehaug, Ingvild [1 ]
Vetti, Nils [2 ]
机构
[1] Haukeland Hosp, Dept Oncol & Med Phys, Bergen, Norway
[2] Haukeland Hosp, Dept Radiol, Bergen, Norway
关键词
Tomography; X-ray computed; metals; artifacts; algorithms; phantoms; diagnostic imaging; orthopedic fixation devices; COMPUTED-TOMOGRAPHY; RADIATION-THERAPY;
D O I
10.1177/0284185117751278
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Metal implants may introduce severe artifacts in computed tomography (CT) images. Over the last few years dedicated algorithms have been developed in order to reduce metal artifacts in CT images. Purpose: To investigate and compare metal artifact reduction algorithms (MARs) from four different CT vendors when imaging three different orthopedic metal implants. Material and Methods: Three clinical metal implants were attached to the leg of an anthropomorphic phantom: cobalt-chrome; stainless steel; and titanium. Four commercial MARs were investigated: SmartMAR (GE); O-MAR (Philips); iMAR (Siemens); and SEMAR (Toshiba). The images were evaluated subjectively by three observers and analyzed objectively by calculating the fraction of pixels with CT number above 500 HU in a region of interest around the metal. The average CT number and image noise were also measured. Results: Both subjective evaluation and objective analysis showed that MARs reduced metal artifacts and improved the image quality for CT images containing metal implants of steel and cobalt-chrome. When using MARs on titanium, all MARs introduced new visible artifacts. Conclusion: The effect of MARs varied between CT vendors and different metal implants used in orthopedic surgery. Both in subjective evaluation and objective analysis the effect of applying MARs was most obvious on steel and cobalt-chrome implants when using SEMAR from Toshiba followed by SmartMAR from GE. However, MARs may also introduce new image artifacts especially when used on titanium implants. Therefore, it is important to reconstruct all CT images containing metal with and without MARs.
引用
收藏
页码:1110 / 1118
页数:9
相关论文
共 17 条
  • [1] Metal artefact reduction in CT imaging of hip prostheses-an evaluation of commercial techniques provided by four vendors
    Andersson, K. M.
    Nowik, P.
    Persliden, J.
    Thunberg, P.
    Norrman, E.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1052)
  • [2] Visual grading evaluation of commercially available metal artefact reduction techniques in hip prosthesis computed tomography
    Andersson, Karin M.
    Norrman, Eva
    Geijer, Hakan
    Krauss, Wolfgang
    Cao, Yang
    Jendeberg, Johan
    Geijer, Mats
    Liden, Mats
    Thunberg, Per
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1063)
  • [3] Artifacts in CT: Recognition and avoidance
    Barrett, JF
    Keat, N
    [J]. RADIOGRAPHICS, 2004, 24 (06) : 1679 - 1691
  • [4] Boas FE., 2012, IMAGING MED, V4, P229, DOI DOI 10.2217/IIM.12.13
  • [5] Dual-Energy Computed Tomography-How Accurate Is Gemstone Spectrum Imaging Metal Artefact Reduction? Its Application to Orthopedic Metal Implants
    Dabirrahmani, Dane
    Magnussen, John
    Appleyard, Richard Charles
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2015, 39 (06) : 925 - 935
  • [6] GE Healthcare, 2013, SMART MET ART RED MA
  • [7] Metal artifacts in computed tomography for radiation therapy planning: dosimetric effects and impact of metal artifact reduction
    Giantsoudi, Drosoula
    De Man, Bruno
    Verburg, Joost
    Trofimov, Alexei
    Jin, Yannan
    Wang, Ge
    Gjesteby, Lars
    Paganetti, Harald
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (08) : R49 - R80
  • [8] Metal Artifact Reduction: Standard and Advanced Magnetic Resonance and Computed Tomography Techniques
    Gupta, Amit
    Subhas, Naveen
    Primak, Andrew N.
    Nittka, Mathias
    Liu, Kecheng
    [J]. RADIOLOGIC CLINICS OF NORTH AMERICA, 2015, 53 (03) : 531 - +
  • [9] An evaluation of three commercially available metal artifact reduction methods for CT imaging
    Huang, Jessie Y.
    Kerns, James R.
    Nute, Jessica L.
    Liu, Xinming
    Balter, Peter A.
    Stingo, Francesco C.
    Followill, David S.
    Mirkovic, Dragan
    Howell, Rebecca M.
    Kry, Stephen F.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (03) : 1047 - 1067
  • [10] Reduction of dental metallic artefacts in CT: Value of a newly developed algorithm for metal artefact reduction (O-MAR)
    Kidoh, M.
    Nakaura, T.
    Nakamura, S.
    Tokuyasu, S.
    Osakabe, H.
    Harada, K.
    Yamashita, Y.
    [J]. CLINICAL RADIOLOGY, 2014, 69 (01) : E11 - E16