A Programmable Aerosol Diluter for Generating Time-Varied Sub-Micrometer Particles

被引:2
|
作者
Liu, Qiaoling [1 ]
Chen, Da-Ren [1 ]
机构
[1] Virginia Commonwealth Univ, Dept Mech & Nucl Engn, Particle Lab, Richmond, VA 23294 USA
关键词
Diluter; Dilution ratio; Dynamic aerosol; Programmable control; SIZE DISTRIBUTIONS; NUCLEATION; PERFORMANCE; COAGULATION; GROWTH; SYSTEM; RATES;
D O I
10.4209/aaqr.2014.02.0032
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Real-world aerosols often vary over time. A time-varied aerosol generation system is thus needed to evaluate the performance of aerosol sensors for characterizing time-dependent aerosol size distribution. However, all laboratory aerosol generators are designed for the production of aerosols with stable concentrations and steady size distributions. A simple way to produce reliable time-dependent aerosols is to combine a stable aerosol generator with a programmable aerosol diluter. Aerosol diluters are also needed to measure aerosols in high concentrations using existing aerosol instruments. In this study we focused on the design and evaluation of a programmable aerosol diluter. The dilution flow rate of the aerosol diluter was controlled by a programmable mass flow controller with a flow rate of up to 200 L/min. Steady and dynamic dilution processes in the diluter were programmed using Visual Basic. Experiments were carried out to characterize the steady and dynamic dilution performance of the aerosol diluter for DMA-classified particles with sizes ranging from 10 nm to 1.0 mu m. The steady dilution result shows that the diluter has a non-size dependent dilution performance, and there is a good linear relationship between the aerosol dilution ratio and dilution flow rate ratio (with a calibrated line slope of 1.03, close to the ideal line slope of 1.0). Our experiments further indicate that efficient aerosol mixing in the diluter can be achieved when operated at a flow Reynolds number above 450. The evaluation of four dynamic dilution modes also provides evidence of the excellent performance of the diluter, which has the capability of continuously producing well-defined and concentration-varied aerosols. A simple empirical model was also proposed to describe the steady and dynamic dilution performance of the diluter.
引用
收藏
页码:1838 / 1850
页数:13
相关论文
共 50 条
  • [41] Highly monodisperse chemically reactive sub-micrometer particles: polymer colloidal photonic crystals
    Zhenyu Luo
    Chunxiao Zou
    Saad Syed
    Luthfia A. Syarbaini
    Guofang Chen
    Colloid and Polymer Science, 2012, 290 : 141 - 150
  • [42] Rapid measurement of sub-micrometer aerosol size distribution using a fast integrated mobility spectrometer
    Wang, Yang
    Pinterich, Tamara
    Wang, Jian
    JOURNAL OF AEROSOL SCIENCE, 2018, 121 : 12 - 20
  • [43] Polyol mediated synthesis of sub-micrometer Bi2O3 particles
    H.-O. Jungk
    C. Feldmann
    Journal of Materials Science, 2001, 36 : 297 - 299
  • [44] Control strategies for sub-micrometer particles indoors: model study of air filtration and ventilation
    Jamriska, M
    Morawska, L
    Ensor, DS
    INDOOR AIR, 2003, 13 (02) : 96 - 105
  • [45] Ag impregnated sub-micrometer crystalline jute cellulose particles: Catalytic and antibacterial properties
    Rabbi, Mohammad Ahasanur
    Rahman, Mohammad Mahbubor
    Minami, Hideto
    Habib, Mohammad Rowshanul
    Ahmad, Hasan
    CARBOHYDRATE POLYMERS, 2020, 233
  • [46] Enhanced Catalytic Combustion Using Sub-micrometer and Nano-size Platinum Particles
    Ma, Yi
    Ricciuti, Christopher
    Miller, Tiffany
    Kadlowec, Jennifer
    Pearlman, Howard
    ENERGY & FUELS, 2008, 22 (06) : 3695 - 3700
  • [47] DEPOSITION OF SUB-MICROMETER PARTICLES ONTO A HEATED SURFACE WITH WIDELY SPACED ROUGHNESS ELEMENTS
    OWEN, I
    ELKADY, AA
    CLEAVER, JW
    JOURNAL OF AEROSOL SCIENCE, 1989, 20 (06) : 671 - 681
  • [48] Air-stable silver-coated copper particles of sub-micrometer size
    Jung, D. S.
    Lee, H. M.
    Kang, Y. C.
    Park, S. B.
    JOURNAL OF COLLOID AND INTERFACE SCIENCE, 2011, 364 (02) : 574 - 581
  • [49] Size-resolved flux measurement of sub-micrometer particles over an urban area
    Deventer, Malte Julian
    Griessbaum, Frank
    Klemm, Otto
    METEOROLOGISCHE ZEITSCHRIFT, 2013, 22 (06) : 729 - 737
  • [50] Sub-Micrometer Particles Produced by a Low-Powered AC Electric Arc in Liquids
    Jaworski, Jacek A.
    Fleury, Eric
    JOURNAL OF NANOSCIENCE AND NANOTECHNOLOGY, 2012, 12 (01) : 604 - 609