Biomechanical Comparison of Titanium Locking Fragment-Specific and Volar Locking Plates for AO B1 and B2 Fractures of the Distal Radius

被引:1
|
作者
Martin, Dennis P. [1 ]
Park, Andrew G. [1 ]
Jamison, David [2 ]
Jones, Christopher M. [1 ]
Ilyas, Asif M. [1 ]
机构
[1] Thomas Jefferson Univ, Sidney Kimmel Med Coll, Dept Orthoped Surg, Rothman Inst, 125 S 9th St,Suite 1000, Philadelphia, PA 19107 USA
[2] Villanova Univ, Dept Mech Engn, Villanova, PA 19085 USA
来源
JOURNAL OF HAND SURGERY-AMERICAN VOLUME | 2019年 / 44卷 / 12期
关键词
Distal radius; dorsal shear; fragment-specific; radial styloid; volar locking; LOW-PROFILE DORSAL; DISPLACED FRACTURES; SURGICAL-TREATMENT; FIXATION; PART; COMPLICATIONS; STABILITY; FORCES; SYSTEM; SCREW;
D O I
10.1016/j.jhsa.2019.02.003
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose This biomechanical study compared the stability of volar locking plates (VLPs) and locking fragment-specific (LFS) dorsal and radial styloid plates for the fixation of dorsal (AO 23-B2) and radial styloid (AO 23-B1) shear fractures of the distal radius, respectively. Methods Two groups of 6 composite radii were fixed with a VLP or an LFS dorsal plate over a simulated dorsal shear fracture. Two additional groups of 6 radii received the same VLP or an LFS radial plate to fix a radial styloid fracture. Each plated radius was tested under cyclic axial compression by a servohydraulic testing machine that recorded axial displacement per cycle. Construct stiffness was calculated from the slope of the force-displacement curve. Results In the dorsal shear fracture model, the dorsal LFS plate exhibited less displacement than the VLP (0.32 +/- 0.04 vs 0.43 +/- 0.07 mm, respectively) and showed greater average stiffness (645 +/- 64 vs 433 +/- 88 N/mm, respectively). Plate type was responsible for 53.1% of the variation in displacement and 68.6% of the variation in stiffness. In the radial styloid fracture model, variations due to number of cycles elapsed and plate type were similar for displacement and stiffness in both groups. The average stiffness during cyclical nondestructive testing was 566 +/- 45 and 573 +/- 60 N/mm for VLP and LFS radial plating groups, respectively. Conclusions For AO 23-B2 (dorsal rim) fractures, the dorsal LFS plates exhibited significantly less displacement and greater stiffness in axial loading than VLPs. ForAO23-B1 (radial styloid) fractures, the VLP displayed similar displacement and stiffness to the radial LFS plates. Copyright (C) 2019 by the American Society for Surgery of the Hand. All rights reserved.
引用
收藏
页码:1093.e1 / 1093.e8
页数:8
相关论文
共 31 条
  • [1] Comparison between Carbon-Peek volar locking plates and titanium volar locking plates in the treatment of distal radius fractures
    Perugia, Dario
    Guzzini, Matteo
    Mazza, Daniele
    Iorio, Carlo
    Civitenga, Carolina
    Ferretti, Andrea
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2017, 48 : S24 - S29
  • [2] Biomechanical Comparison of Volar Fixed-Angle Locking Plates for AO C3 Distal Radius Fractures: Titanium Versus Stainless Steel With Compression
    Marshall, Tyler
    Momaya, Amit
    Eberhardt, Alan
    Chaudhari, Nilesh
    Hunt, Thomas R., III
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2015, 40 (10): : 2032 - 2038
  • [3] Biomechanical comparison between stainless steel, titanium and carbon-fiber reinforced polyetheretherketone volar locking plates for distal radius fractures
    Mugnai, Raffaele
    Tarallo, Luigi
    Capra, Francesco
    Catani, Fabio
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2018, 104 (06) : 877 - 882
  • [4] Randomized Comparison of Volar Locking Plates and Intramedullary Nails for Unstable Distal Radius Fractures
    Plate, Johannes F.
    Gaffney, Daniel L.
    Emory, Cynthia L.
    Mannava, Sandeep
    Smith, Beth P.
    Koman, L. Andrew
    Wiesler, Ethan R.
    Li, Zhongyu
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2015, 40 (06): : 1095 - 1101
  • [5] Letter Regarding "Biomechanical Comparison of Volar Fixed-Angle Locking Plates for AO C3 Distal Radius Fractures: Titanium Versus Stainless Steel With Compression"
    Spiegel, Adrian
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2016, 41 (07): : E225 - E225
  • [6] Functional Outcomes in Volar-Displaced Distal Radius Fractures Patients with Marginal Rim Fragment Treated by Volar Distal Locking Plates
    Naito, Kiyohito
    Sugiyama, Yoichi
    Kinoshita, Mayuko
    Obata, Hiroyuki
    Goto, Kenji
    Nagura, Nana
    Iwase, Yoshiyuki
    Obayashi, Osamu
    Kaneko, Kazuo
    JOURNAL OF HAND AND MICROSURGERY, 2019, 11 (02) : 100 - 105
  • [7] Biomechanical Comparison of Locking versus Nonlocking Volar and Dorsal T-Plates for Fixation of Dorsally Comminuted Distal Radius Fractures
    Gondusky, Joseph S.
    Carney, Joseph
    Erpenbach, Jonathan
    Robertson, Claire
    Mahar, Andrew
    Oka, Richard
    Thompson, Michael
    Mazurek, Michael
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2011, 25 (01) : 44 - 50
  • [8] Radiographic Alignment of Unstable Distal Radius Fractures Fixed With 1 or 2 Rows of Screws in Volar Locking Plates
    Neuhaus, Valentin
    Badri, Omar
    Ferree, Steven
    Bot, Arjan G. J.
    Ring, David C.
    Mudgal, Chaitanya S.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2013, 38A (02): : 297 - 301
  • [9] Can the Use of Variable-Angle Volar Locking Plates Compensate for Suboptimal Plate Positioning in Unstable Distal Radius Fractures? A Biomechanical Study
    Hart, Adam
    Collins, Melissa
    Chhatwal, Dane
    Steffen, Thomas
    Harvey, Edward J.
    Martineau, Paul A.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2015, 29 (01) : E1 - E6
  • [10] Stability of volar locking plate systems for AO type C3 fractures of the distal radius: biomechanical study in a cadaveric model
    Kamei, Shuzo
    Osada, Denju
    Tamai, Kazuya
    Kato, Nakayuki
    Takai, Morimitsu
    Kameda, Masahiro
    Nohara, Yutaka
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2010, 15 (03) : 357 - 364