Accuracy Assessment of the Dinamap ProCare 400 in Pregnancy and Preeclampsia

被引:17
作者
De Greeff, Annemarie
Ghosh, Donna
Anthony, John [2 ]
Shennan, Andrew [1 ]
机构
[1] Kings Coll London, Maternal & Fetal Res Unit, Div Reprod & Endocrinol, Sch Med,St Thomas Hosp, London SE1 7EH, England
[2] Groote Schuur Hosp, Dept Obstet & Gynaecol, ZA-7925 Cape Town, South Africa
关键词
BLOOD-PRESSURE DEVICE; HYPERTENSION-SOCIETY PROTOCOL; ADULT-POPULATION; VALIDATION;
D O I
10.3109/10641950902968650
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Objective. To assess the accuracy of the Dinamap ProCare 400 in pregnancy and preeclampsia, according to the British Hypertension Society protocol. Methods. Ethical approval was obtained. Forty-five women were recruited at two large teaching hospitals. Nine sequential same arm measurements were taken from each woman alternating between the Dinamap and mercury sphygmomanometry. Results. The device achieved an A/A grade (pregnancy) and an A/B grade (preeclampsia). Mean difference (SD) were 1.1(5.8)/0.1(5.7) mmHg (pregnancy) and -0.5(5.4)/0.9(9.5) mmHg (preeclampsia) for systolic/diastolic. Conclusion. The Dinamap ProCare 400 is recommended for use in pregnancy and has good accuracy in preeclampsia, although caution is still advised.</.
引用
收藏
页码:198 / 205
页数:8
相关论文
共 14 条
[1]  
*ASS ADV MED INSTR, 2007, SP1019922002 ANSIAAM
[2]   Validation of the Microlife BP3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol [J].
Cuckson, AC ;
Reinders, A ;
Shabeeh, H ;
Shennan, AH .
BLOOD PRESSURE MONITORING, 2002, 7 (06) :319-324
[3]   Clinical assessment of the DINAMAP ProCare monitor in an adult population according to the British Hypertension Society Protocol [J].
de Greeff, Annemarie ;
Reggiori, Fayrouge ;
Shennan, Andrew H. .
BLOOD PRESSURE MONITORING, 2007, 12 (01) :51-55
[4]   Validation and use of the Finometer™ for blood pressure measurement in normal, hypertensive and pre-eclamptic pregnancy [J].
Elvan-Taspinar, A ;
Uiterkamp, LAMS ;
Sikkema, JM ;
Bots, ML ;
Koomans, HA ;
Bruinse, HW ;
Franx, A .
JOURNAL OF HYPERTENSION, 2003, 21 (11) :2053-2060
[5]  
Golara M, 2002, BJOG-INT J OBSTET GY, V109, P1143
[6]   Accuracy of oscillometric blood pressure monitoring in pregnancy and pre-eclampsia [J].
Gupta, M ;
Shennan, AH ;
Halligan, A ;
Taylor, DJ ;
deSwiet, M .
BRITISH JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 1997, 104 (03) :350-355
[7]   Comparison of Portapres® with standard sphygmomanometry in pregnancy [J].
Hehenkamp, WJK ;
Rang, S ;
van Goudoever, J ;
Bos, WJW ;
Wolf, H ;
van der Post, JAM .
HYPERTENSION IN PREGNANCY, 2002, 21 (01) :65-76
[8]   Comparison of auscultatory and oscillometric automated blood pressure monitors in the setting of preeclampsia [J].
Natarajan, P ;
Shennan, AH ;
Penny, J ;
Halligan, AW ;
de Swiet, M ;
Anthony, J .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1999, 181 (05) :1203-1210
[9]   Determination of accuracy in neonates for non-invasive blood pressure device using an improved algorithm [J].
Nelson, RM ;
Stebor, AD ;
Groh, CM ;
Timoney, PM ;
Theobald, KS ;
Friedman, BA .
BLOOD PRESSURE MONITORING, 2002, 7 (02) :123-129
[10]  
O'Brien E., 1993, J HYPERTENSION S2, V11, pS43, DOI DOI 10.1097/00004872-199306000-00013