Left Ventricular Assist Device Inflow Cannula Angle and Thrombosis Risk

被引:75
|
作者
Chivukula, Venkat Keshav [2 ]
Beckman, Jennifer A. [1 ]
Prisco, Anthony R. [4 ]
Dardas, Todd [1 ]
Lin, Shin [1 ]
Smith, Jason W. [3 ]
Mokadam, Nahush A. [3 ]
Aliseda, Alberto [2 ]
Mahr, Claudius [1 ]
机构
[1] Univ Washington, Div Cardiol, 1959 NE Pacific St, Seattle, WA 98195 USA
[2] Univ Washington, Dept Mech Engn, Seattle, WA 98195 USA
[3] Univ Washington, Div Cardiothorac Surg, Seattle, WA 98195 USA
[4] Univ Minnesota, Dept Med, Box 736 UMHC, Minneapolis, MN 55455 USA
关键词
cannula; heart assist devices; heart ventricles; stroke; thrombosis; CONTINUOUS-FLOW; PUMP; MANAGEMENT; TRANSPLANT; THERAPY; SYSTEM; BRIDGE;
D O I
10.1161/CIRCHEARTFAILURE.117.004325
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND: As heart failure prevalence continues to increase in the setting of a static donor supply, left ventricular assist device (LVAD) therapy for end-stage heart failure continues to grow. Anecdotal evidence suggests that malalignment of the LVAD inflow cannula may increase thrombosis risk, but this effect has not been explored mechanistically or quantified statistically. Our objective is to elucidate the impact of surgical angulation of the inflow cannula on thrombogenicity. METHODS AND RESULTS: Unsteady computational fluid dynamics is used in conjunction with computational modeling and virtual surgery to model flow through the left ventricle for 5 different inflow cannula angulations. We use a holistic approach to evaluate thrombogenicity: platelet-based (Lagrangian) metrics to evaluate the platelet mechanical environment, combined with flow-based (Eulerian) metrics to investigate intraventricular hemodynamics. The thrombogenic potential of each LVAD inflow cannula angulation is quantitatively evaluated based on platelet shear stress history and residence time. Intraventricular hemodynamics are strongly influenced by LVAD inflow cannula angulation. Platelet behavior indicates elevated thrombogenic potential for certain inflow cannula angles, potentially leading to platelet activation. Our analysis demonstrates that the optimal range of inflow angulation is within 0 +/- 7 degrees of the left ventricular apical axis. CONCLUSIONS: Angulation of the inflow cannula >7 degrees from the apical axis (axis connecting mitral valve and ventricular apex) leads to markedly unfavorable hemodynamics as determined by computational fluid dynamics. Computational hemodynamic simulations incorporating Lagrangian and Eulerian metrics are a powerful tool for studying optimization of LVAD implantation strategies, with the long-term potential of improving outcomes.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] Neointima-inducing inflow cannula with titanium mesh for left ventricular assist device
    Yukiko Yamada
    Tomohiro Nishinaka
    Toshihide Mizuno
    Yoshiyuki Taenaka
    Eisuke Tatsumi
    Kenji Yamazaki
    Journal of Artificial Organs, 2011, 14 : 269 - 275
  • [22] Left ventricular assist device and pump thrombosis: the importance of the position of the inflow canula
    Kortekaas, K. Kirsten
    De Graaf, M. A.
    Palmen, M.
    Mertens, B. J. A.
    Tops, L. F.
    Beeres, S. L. M. A.
    EUROPEAN JOURNAL OF HEART FAILURE, 2021, 23 : 149 - 150
  • [23] Performance of Extracorporeally Adjustable Ventricular Assist Device Inflow Cannula
    Shiose, Akira
    Kim, Hyun-Il
    Takaseya, Tohru
    Kobayashi, Mariko
    Fujiki, Masako
    Chen, Ji-Feng
    Klatte, Ryan
    Gao, Shengqiang
    Massiello, Alex L.
    Saeed, Diyar
    Fukamachi, Kiyotaka
    ANNALS OF THORACIC SURGERY, 2010, 90 (05): : 1682 - 1687
  • [24] Development of the DexAide right ventricular assist device inflow cannula
    Ootaki, Yoshio
    Saeed, Diyar
    Ootaki, Chiyo
    Akiyama, Masatoshi
    Fumoto, Hideyuki
    Horai, Tetsuya
    Dessoffy, Raymond
    Catanese, Jacquelyn
    Massiello, Alex
    Horvath, David
    Chen, Ji-Feng
    Benefit, Stephen
    Golding, Leonard A. R.
    Fukamachi, Kiyotaka
    ASAIO JOURNAL, 2008, 54 (01) : 31 - 36
  • [25] Left ventricular assist device inflow obstruction
    Vinod Kumar Manikala
    Quynh Nguyen
    Qiang Fu
    Angela Neufeld
    Roderick MacArthur
    Jennifer Conway
    Holger Buchholz
    Indian Journal of Thoracic and Cardiovascular Surgery, 2023, 39 : 165 - 169
  • [26] Left ventricular assist device inflow obstruction
    Manikala, Vinod Kumar
    Nguyen, Quynh
    Fu, Qiang
    Neufeld, Angela
    MacArthur, Roderick
    Conway, Jennifer
    Buchholz, Holger
    INDIAN JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2023, 39 (02) : 165 - 169
  • [27] Transcatheter Aortic Valve Replacement Performed via Left Ventricular Assist Device Inflow Cannula
    Vavalle, John P.
    Kiefer, Todd L.
    Milano, Carmelo A.
    Harrison, J. Kevin
    Hughes, G. Chad
    CIRCULATION-HEART FAILURE, 2014, 7 (03) : 544 - 546
  • [28] Percutaneous Transcatheter Balloon Dilatation and Stenting to the Inflow Cannula Stenosis of a Left Ventricular Assist Device
    Carr, Stephen M.
    Lubbe, Dieter F.
    Huber, Philip R.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2017, 89 (07) : 1219 - 1223
  • [29] Explantation of a CircuLite left ventricular assist device without removal of the inflow cannula: how to do it?
    Sainte, Sarah
    Gewillig, Marc
    Droogne, Walter
    Van Cleemput, Johan
    Vanhaecke, Johan
    Meyns, Bart
    Rega, Filip
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2014, 18 (03) : 393 - 395
  • [30] PACING-INDUCED DYSSYNCHRONY TO RESOLVE LEFT VENTRICULAR ASSIST DEVICE INFLOW CANNULA OBSTRUCTION
    Guarache, Ricardo Cardona
    Kalahasty, Gautham
    Kontos, Michael
    Kasirajan, Vigneshwar
    Shah, Keyur
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2016, 67 (13) : 1059 - 1059