Evaluation of a high-volume portable bioaerosol sampler in laboratory and field environments

被引:48
作者
An, HR [1 ]
Mainelis, G [1 ]
Yao, M [1 ]
机构
[1] Rutgers State Univ, Dept Environm Sci, New Brunswick, NJ 08901 USA
关键词
bioaerosols; RCS High Flow sampler; BioSampler; collection efficiency;
D O I
10.1111/j.1600-0668.2004.00257.x
中图分类号
TU [建筑科学];
学科分类号
0813 ;
摘要
This study investigated the physical and biological performances of a portable centrifugal sampler for viable bioaerosols, RCS High Flow. The performance of the test sampler in the laboratory and field environments was compared with that of a reference sampler, BioSampler. The laboratory experiments with non-biological particles of KCl, oleic acid, and polystyrene latex showed that the test sampler's collection efficiency is about 22% for 0.5-mum particles, 48% for 1.0-mum particles, and approximately 100% for particles of 2.5 mum and larger. These tests indicated that the sampler's cut-off size (d(50)) was 1.1 mum. The test sampler's physical performance when collecting the spores and vegetative cells of Bacillus subtilis var. niger (BG) was similar to that when collecting non-biological particles of the same size. In the laboratory tests, the RCS High Flow sampler was found to enumerate approximately 40% of BG spores and cells relative to the reference sampler, BioSampler. A similar ratio was found during testing in an indoor environment. This ratio decreased to below 10% when testing was performed in an outdoor environment. We hypothesize that the test sampler's underperformance compared with the BioSampler could be caused by the damage to sensitive microorganisms during the collection process, test sampler's sensitivity to wind direction and speed as well as break-up of particle aggregates during the impingement process in BioSampler, which resulted in more colony-forming units (CFUs) being counted by the reference sampler than by the test sampler. Overall, when the RCS High Plus is used to sample culturable airborne microorganisms, the results obtained may have to be adjusted to avoid potential underestimation of microorganism concentration in the air.
引用
收藏
页码:385 / 393
页数:9
相关论文
共 36 条
  • [1] Development and evaluation of a new personal sampler for culturable airborne microorganisms
    Agranovski, IE
    Agranovski, V
    Reponen, T
    Willeke, K
    Grinshpun, SA
    [J]. ATMOSPHERIC ENVIRONMENT, 2002, 36 (05) : 889 - 898
  • [2] [Anonymous], 1997, MED ASPECTS CHEM BIO
  • [3] SAMPLING AND ANALYSIS OF BIOLOGICAL AEROSOLS
    BURGE, HA
    SOLOMON, WR
    [J]. ATMOSPHERIC ENVIRONMENT, 1987, 21 (02) : 451 - 456
  • [4] BURGE HA, 1989, GUIDELINES ASSESSMEN
  • [5] FRIEDLANDER AM, 1997, MED ASPECTS CHEM BIO
  • [6] Influence of Bacillus subtilis cell walls and EDTA on calcite dissolution rates and crystal surface features
    Friis, AK
    Davis, TA
    Figueira, MM
    Paquette, J
    Mucci, A
    [J]. ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2003, 37 (11) : 2376 - 2382
  • [7] Fungal fragments as indoor air biocontaminants
    Górny, RL
    Reponen, T
    Willeke, K
    Schmechel, D
    Robine, E
    Boissier, M
    Grinshpun, SA
    [J]. APPLIED AND ENVIRONMENTAL MICROBIOLOGY, 2002, 68 (07) : 3522 - 3531
  • [8] INLET CHARACTERISTICS OF BIOAEROSOL SAMPLERS
    GRINSHPUN, SA
    CHANG, CW
    NEVALAINEN, A
    WILLEKE, K
    [J]. JOURNAL OF AEROSOL SCIENCE, 1994, 25 (08) : 1503 - 1522
  • [9] EVALUATION OF MICROBIOLOGICAL AEROSOL SAMPLERS - A REVIEW
    HENNINGSON, EW
    AHLBERG, MS
    [J]. JOURNAL OF AEROSOL SCIENCE, 1994, 25 (08) : 1459 - 1492
  • [10] Hill SC, 1999, FIELD ANAL CHEM TECH, V3, P221, DOI 10.1002/(SICI)1520-6521(1999)3:4/5<221::AID-FACT2>3.0.CO