Biometry with a new swept-source optical coherence tomography biometer: Repeatability and agreement with an optical low-coherence reflectometry device

被引:78
|
作者
Kurian, Mathew [1 ]
Negalur, Nikhil [1 ]
Das, Sudeep [1 ]
Puttaiah, Narendra K. [1 ]
Haria, Dhawal [1 ]
Tejal, S. J. [1 ]
Thakkar, Maithil M. [1 ]
机构
[1] Narayana Nethralaya, Bangalore, Karnataka, India
关键词
INTRAOCULAR-LENS POWER; INTERFEROMETRY; ACCURACY;
D O I
10.1016/j.jcrs.2016.01.038
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To assess the repeatability of a new swept-source optical coherence tomography (SS-OCT) biometer (IOLMaster 700) and evaluate the agreement with an optical low-coherence reflectometry (OLCR) biometer (Lenstar 900) in cataractous eyes. SETTING: Tertiary eyecare facility. DESIGN: Prospective evaluation of a diagnostic test. METHODS: Eyes of patients had 3 consecutive scans acquired by the same operator. The repeatability of measurements was assessed using the within-subject standard deviation and coefficient of variation (CoV). Agreement between the mean measurements of each machine was evaluated. RESULTS: The study comprised 100 eyes of 100 patients. The median lens thickness and axial length (AL) measurements were significantly greater (P < .001) with SS-OCT (4.29 mm and 23.5 mm, respectively, versus 4.25 mm and 23.4 mm, respectively); anterior chamber depth (ACD) was greater with OLCR (2.72 mm versus 2.70 mm, respectively) (P < .001). The within-subject standard deviation and CoV of the AL (0.01 and 0.05, respectively) and the ACD (0.04 and 1.22, respectively) measurements showed lower variability with SS-OCT (0.05 and 0.21, respectively, for AL; 0.06 and 1.99, respectively, for ACD) than with OLCR. Good agreement was seen for anterior segment measurements between the 2 devices. The 95% limits of agreement of most parameters fell within the test retest variability of the parameters. The SS-OCT biometer also showed better penetration in dense posterior subcapsular cataracts, measuring AL successfully in 96% of cases. CONCLUSIONS: The new SS-OCT biometer showed good repeatability and agreement with the OLCR biometer. The SS-OCT biometer measured the AL with fewer dropouts compared with the OLCR even in dense cataracts. (C) 2016 ASCRS and ESCRS
引用
收藏
页码:577 / 581
页数:5
相关论文
共 50 条
  • [21] Clinical comparison of a new swept-source optical coherence tomography-based optical biometer and a time-domain optical coherence tomography-based optical biometer
    Srivannaboon, Sabong
    Chirapapaisan, Chareenun
    Chonpimai, Pratuangsri
    Loket, Siriwan
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (10) : 2224 - 2232
  • [22] Repeatability and reproducibility of optical biometry implemented in a new optical coherence tomographer and comparison with a optical low-coherence reflectometer
    Kanclerz, Piotr
    Hoffer, Kenneth J.
    Rozema, Jos J.
    Przewlocka, Katarzyna
    Savini, Giacomo
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (11) : 1619 - 1624
  • [23] Reliability of a New Swept-Source Optical Coherence Tomography Biometer in Healthy Children, Adults, and Cataract Patients
    Huang, Jinhai
    Zhao, Yuyu
    Savini, Giacomo
    Yu, Guanshun
    Yu, Jinjin
    Chen, Zhongxing
    Tu, Ruixue
    Zhao, Yune
    JOURNAL OF OPHTHALMOLOGY, 2020, 2020
  • [24] Comparison of 2 swept-source optical coherence tomography-based biometry devices
    Fisus, Andreea D.
    Hirnschall, Nino D.
    Findl, Oliver
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2021, 47 (01) : 87 - 92
  • [25] Meta-analysis of optical low-coherence reflectometry versus partial coherence interferometry biometry
    Huang, Jinhai
    McAlinden, Colm
    Huang, Yingying
    Wen, Daizong
    Savini, Giacomo
    Tu, Ruixue
    Wang, Qinmei
    SCIENTIFIC REPORTS, 2017, 7 : 1 - 10
  • [26] Comparison of two novel swept-source optical coherence tomography devices to a partial coherence interferometry-based biometer
    Chan, Tommy C. Y.
    Yu, Marco C. Y.
    Chiu, Vivian
    Lai, Gilda
    Leung, Christopher K. S.
    Chan, Poemen P. M.
    SCIENTIFIC REPORTS, 2021, 11 (01)
  • [27] Changes in the anterior segment after cycloplegia with a biometer using swept-source optical coherence tomography
    Higashiyama, Tomoaki
    Iwasa, Maki
    Ohji, Masahito
    PLOS ONE, 2017, 12 (08):
  • [28] Accuracy of a New Swept-Source Optical Coherence Tomography Biometer for IOL Power Calculation and Comparison to IOL Master
    Savini, Giacomo
    Hoffer, Kenneth J.
    Shammas, H. John
    Aramberri, Jaime
    Huang, Jinhai
    Barboni, Piero
    JOURNAL OF REFRACTIVE SURGERY, 2017, 33 (10) : 690 - 695
  • [29] Biometry measurements using a new large-coherence-length swept-source optical coherence tomographer
    Shammas, H. John
    Ortiz, Sergio
    Shammas, Maya C.
    Kim, Seong Hwan
    Chong, Changho
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (01) : 50 - 61
  • [30] Agreement between the Swept-Source Optical Coherence Tomography and the Image-Guided System for Biometry Assessment in Cataract Surgery
    Moreno-Paramo, Edel
    Garcia-Arroyo, Santiago
    Ortiz-Ramirez, Grecia Y.
    Garza-Leon, Manuel
    Gonzalez-Salinas, Roberto
    SEMINARS IN OPHTHALMOLOGY, 2022, 37 (03) : 324 - 329