Fast and simple procedure for liquid-liquid extraction of 136 analytes from different drug classes for development of a liquid chromatographic-tandem mass spectrometric quantification method in human blood plasma

被引:62
作者
Remane, Daniela [1 ]
Meyer, Markus R. [1 ]
Peters, Frank T. [1 ,2 ]
Wissenbach, Dirk K. [1 ]
Maurer, Hans H. [1 ]
机构
[1] Univ Saarland, Dept Expt & Clin Toxicol, Inst Expt & Clin Pharmacol & Toxicol, D-66421 Homburg, Saar, Germany
[2] Univ Jena, Inst Forens Med, D-07740 Jena, Germany
关键词
Liquid-liquid extraction; Recovery; Matrix effects; Internal standard; Multi-analyte LC-MS/MS method; Plasma; PRESSURE CHEMICAL-IONIZATION; LIBRARY-ASSISTED IDENTIFICATION; SOLID-PHASE EXTRACTION; VALIDATED QUANTIFICATION; ELECTROSPRAY-IONIZATION; WHOLE-BLOOD; FORENSIC TOXICOLOGY; ACTIVE METABOLITES; HUMAN SERUM; LC-MS;
D O I
10.1007/s00216-010-3820-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In clinical and forensic toxicology, different extraction procedures as well as analytical methods are used to monitor different drug classes of interest in biosamples. Multi-analyte procedures are preferable because they make the analytical strategy much simpler and cheaper and allow monitoring of analytes of different drug classes in one single body sample. For development of such a multi-analyte liquid chromatography-tandem mass spectrometry approach, a rapid and simple method for the extraction of 136 analytes from the following drug classes has been established: antidepressants, neuroleptics, benzodiazepines, beta-blockers, oral antidiabetics, and analytes relevant in the context of brain death diagnosis. Recovery, matrix effects, and process efficiency were tested at two concentrations using six different lots of blank plasma. The recovery results obtained using absolute peak areas were compared with those calculated using area ratios analyte/internal standard. The recoveries ranged from 8% to 84% for antidepressants, from 10% to 79% for neuroleptics, from 60% to 81% for benzodiazepines, from 1% to 71% for beta-blockers, from 10% to 73% for antidiabetics, and from 60% to 86% for analytes relevant in the context of brain death diagnosis. With the exception of 52 analytes at low concentration and 37 at high concentration, all compounds showed recoveries with acceptable variability with less than 15% and 20% coefficients of variation. Recovery results obtained by comparing peak area ratios were nearly the same, but 35 analytes at low concentration and 17 at high concentration lay above the acceptance criteria. Matrix effects with more than 25% were observed for 18 analytes. The results were acceptable for 119 analytes at high concentrations.
引用
收藏
页码:2303 / 2314
页数:12
相关论文
共 44 条
[1]   The AGNP-TDM expert group consensus guidelines:: Therapeutic Drug Monitoring in psychiatry [J].
Baumann, P ;
Hiemke, C ;
Ulrich, S ;
Eckermann, G ;
Gaertner, I ;
Gerlach, M ;
Kuss, HJ ;
Laux, G ;
Müller-Oerlinghausen, B ;
Rao, ML ;
Riederer, P ;
Zernig, G .
PHARMACOPSYCHIATRY, 2004, 37 (06) :243-265
[2]   Quantification of eight new antidepressants and five of their active metabolites in whole blood by high-performance liquid chromatography-tandem mass spectrometry [J].
Castaing, Nadege ;
Titier, Karine ;
Receveur-Daurel, Mathilde ;
Le-Deodic, Maite ;
Le-bars, Delphine ;
Moore, Nicholas ;
Molimard, Mathieu .
JOURNAL OF ANALYTICAL TOXICOLOGY, 2007, 31 (06) :334-341
[3]   High-throughput on-line solid-phase extraction-liquid chromatography-tandem mass spectrometry method for the simultaneous analysis of 14 antidepressants and their metabolites in plasma [J].
de Castro, Ana ;
Femandez, Maria Del Mar Ramirez ;
Laloup, Marleen ;
Samyn, Nele ;
De Boeck, Gert ;
Wood, Michelle ;
Maes, Viviane ;
Lopez-Rivadulla, Manuel .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1160 (1-2) :3-12
[4]   Clinical Utility of an LC-MS/MS Seizure Panel for Common Drugs Involved in Drug-Induced Seizures [J].
Drees, Julia C. ;
Stone, Judy A. ;
Olson, Kent R. ;
Meier, Kathryn H. ;
Gelb, Alan M. ;
Wu, Alan H. B. .
CLINICAL CHEMISTRY, 2009, 55 (01) :126-133
[5]   Quantification of benzodiazepines in whole blood and serum [J].
Dussy, Franz E. ;
Hamberg, Cornelia ;
Briellmann, Thomas A. .
INTERNATIONAL JOURNAL OF LEGAL MEDICINE, 2006, 120 (06) :323-330
[6]   Simultaneous screening and quantitation of 18 antihistamine drugs in blood by liquid chromatography ionspray tandem mass spectrometry [J].
Gergov, M ;
Robson, JN ;
Ojanperä, I ;
Heinonen, OP ;
Vuori, E .
FORENSIC SCIENCE INTERNATIONAL, 2001, 121 (1-2) :108-115
[7]   Fast gas chromatography-negative-ion chemical ionization mass spectrometry with microscale volume sample preparation for the determination of benzodiazepines and α-hydroxy metabolites, zaleplon and zopiclone in whole blood [J].
Gunnar, Teemu ;
Ariniemi, Kari ;
Lillsunde, Pirjo .
JOURNAL OF MASS SPECTROMETRY, 2006, 41 (06) :741-754
[8]   Therapeutic drug monitoring of 13 antidepressant and five neuroleptic drugs in serum with liquid chromatography-electrospray ionization mass spectrometry [J].
Gutteck, U ;
Rentsch, KM .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2003, 41 (12) :1571-1579
[9]  
International Conference on Harmonization (ICH), 2005, Q2 (R1): Text on Validation of Analytical Procedures
[10]  
SANCO. Guidance Document Onanalytical Quality Control and Validation Procedures for Pesticide Residues Analysis in Food and Feed