Validation of the SToP formula for calculating intraocular lens power in eyes with previous myopic excimer laser surgery

被引:13
作者
Savini, Giacomo [1 ]
Hoffer, Kenneth [4 ,5 ]
Barboni, Piero [2 ,3 ]
Balducci, Nicole [1 ]
Schiano-Lomoriello, Domenico [1 ]
机构
[1] IRCCS Fdn Bietti, Via Livenza 3, Rome, Italy
[2] Studio Oculist Azeglio, Bologna, Italy
[3] Ist Sci San Raffaele, Milan, Italy
[4] Univ Calif Los Angeles, Stein Eye Inst, Los Angeles, CA USA
[5] St Marys Eye Ctr, Santa Monica, CA USA
关键词
REQUIRING CATARACT-SURGERY; CORNEAL POWER; REFRACTIVE SURGERY; ADJUSTMENT; ACCURACY;
D O I
10.1016/j.jcrs.2019.06.011
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To validate the accuracy of the SToP formula for calculating the intraocular lens (IOL) power from measurements obtained with a rotating Scheimpflug camera (Pentacam) in eyes with previous myopic excimer laser surgery. Setting: IRCCS - Fondazione Bietti, Rome, Italy. Design: Retrospective case series. Methods: Corneal curvature measurements were taken in eyes that had cataract surgery after myopic excimer laser surgery. Values from both versions of the SToP calculator (based, respectively, on the Holladay 1 and the SRKTT formulas) were compared with historical (Masket and Savini) and no-history methods (Barrett True-K, BESSt formula, equivalent K-readings, Triple-S, and Shammas). Results: Fifty eyes of 50 patients with historical data were analyzed. The SToP formula based on the SRK/T achieved the lowest median absolute error (MedAE) (0.31 diopters [D]) and a higher percentage of eyes (62%) with a prediction error in refraction within +/- 0.5 D. Excellent outcomes were also achieved with the Masket method, the BESSt formula, the Savini method, and the Triple-S method, which led to similar MedAEs (from 0.31 to 0.37 D) and an even higher percentage of eyes with a prediction error within +/- 0.5 D (from 64% to 76%). The SToP formula based on the Holladay 1 was slightly less accurate. Conclusions: Several methods based on Scheimpflug camera measurements were used to successfully calculate the IOL power in eyes with previous myopic excimer laser surgery. The SToP formula based on the SRK/T was one of the best choices. (C) 2019 Published by Elsevier Inc. on behalf of ASCRS and ESCRS
引用
收藏
页码:1562 / 1567
页数:6
相关论文
共 34 条
[1]   Accuracy of the Barrett True-K formula for intraocular lens power prediction after laser in situ keratomileusis or photorefractive keratectomy for myopia [J].
Abulafia, Adi ;
Hill, Warren E. ;
Koch, Douglas D. ;
Wang, Li ;
Barrett, Graham D. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (03) :363-369
[2]   Intraocular lens power calculation after corneal refractive surgery: Double-K method [J].
Aramberri, J .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (11) :2063-2068
[3]  
Barrett GD, Barrett Universal II Formula
[4]  
Barrett GD, BARRETT TRUE K FORMU
[5]   Estimation of true corneal power after keratorefractive surgery in eyes requiring cataract surgery: BESSt formula [J].
Borasio, Edmondo ;
Stevens, Julian ;
Smith, Guy T. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (12) :2004-2014
[6]   A comparison of two methods to calculate axial length [J].
Cooke, David L. ;
Cooke, Timothy L. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (03) :284-292
[7]   Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis [J].
Haigis, W ;
Lege, B ;
Miller, N ;
Schneider, B .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2000, 238 (09) :765-773
[8]  
Hoffer KJ, 2007, J CATARACT REFR SURG, V33, P2, DOI 10.1016/j.jcrs.2006.08.054
[9]   Protocols for Studies of Intraocular Lens Formula Accuracy [J].
Hoffer, Kenneth J. ;
Aramberri, Jaime ;
Haigis, Wolfgang ;
Olsen, Thomas ;
Savini, Giacomo ;
Shammas, H. John ;
Bentow, Stanley .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2015, 160 (03) :403-405
[10]   Clinical results using the Holladay 2 intraocular lens power formula [J].
Hoffer, KJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (08) :1233-1237