Use of Arduino in the Development of a New and Fast Automated Online Preconcentration System Based on Double-Knotted Reactor for the Mn Determination in Tea Samples by Flame Atomic Absorption Spectrometry

被引:3
作者
Barreto, Jeferson A. [1 ]
Lemos, Valfredo A. [1 ]
de Oliveira, Djalma M. [1 ]
Cerqueira, Uillian Mozart F. M. [1 ]
Meira, Lucilia A. [1 ]
Bezerra, Marcos A. [1 ]
机构
[1] UESB, LQA, Campus Jequie, BR-45206191 Jequie, BA, Brazil
关键词
Arduino; double-knotted reactor; automated online system; multicomutation; manganese; SOLID-PHASE EXTRACTION; CLOUD POINT EXTRACTION; FLOW-ANALYSIS; MULTICOMMUTATION; MANGANESE; OPTIMIZATION; SPECIATION; CADMIUM; COPPER; TOOL;
D O I
10.21577/0103-5053.20190118
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
In the present paper, an automated online preconcentration system based on the use of two knotted reactors (KR) was developed for Mn determination in tea samples by flame atomic absorption spectrometry (F AAS). The system automation was performed by using four solenoid valves to switch sample streams and reagents. Valves were driven electronically by an integrated Darlington transistor circuit and controlled by an open-source Arduino hardware platform written in C language. The preconcentration was based on the precipitation process of metal ions complexed with 2-(5-bromo-2-pyridylazo)-5-(diethylamino)phenol (5-Br-PADAP) on the inner walls of the reactor, followed by elution with a solution of hydrochloric acid. The optimized system allowed the Mn determination with limit of detection (LOD) of 2.1 mu g L-1, precision (relative standard deviation (RSD), 50.0 mu g L-1) of 2.6% and enrichment factors of 8. The system presented a sampling frequency of 60 h(-1). The method accuracy was evaluated by analysis of the certified reference material (CRM) apple leaves (NIST 1515) and by addition/ recovery tests obtaining recoveries between 95.2 to 119%. This method was applied in the Mn determination in commercial tea samples purchased from local trade. Manganese levels were found in a concentration range of 2.95 to 123 mu g g(-1).
引用
收藏
页码:15 / 24
页数:10
相关论文
共 50 条
[1]  
[Anonymous], 2010, STAT SOFTW 10 0
[2]   Manganese and its Role in Parkinson's Disease: From Transport to Neuropathology [J].
Aschner, Michael ;
Erikson, Keith M. ;
Hernandez, Elena Herrero ;
Tjalkens, Ronald .
NEUROMOLECULAR MEDICINE, 2009, 11 (04) :252-266
[3]  
Azevedo F.A., 2003, METAIS GERENCIAMENTO
[4]  
Barros Neto B, 2010, COMO FAZER EXPERIMEN
[5]   Response surface methodology (RSM) as a tool for optimization in analytical chemistry [J].
Bezerra, Marcos Almeida ;
Santelli, Ricardo Erthal ;
Oliveira, Eliane Padua ;
Villar, Leonardo Silveira ;
Escaleira, Luciane Amlia .
TALANTA, 2008, 76 (05) :965-977
[6]  
CERDA A, 2009, INTRO FLOW ANAL
[7]   Knotted reactors and their role in flow-injection on-line preconcentration systems coupled to atomic spectrometry-based detectors [J].
Cerutti, S ;
Martinez, LD ;
Wuilloud, RG .
APPLIED SPECTROSCOPY REVIEWS, 2005, 40 (01) :71-101
[8]   Dual extraction based on solid phase extraction and solidified floating organic drop microextraction for speciation of arsenic and its distribution in tea leaves and tea infusion by electrothermal vaporization ICP-MS [J].
Chen, Shizhong ;
Li, Jianfen ;
Lu, Dengbo ;
Zhang, Yan .
FOOD CHEMISTRY, 2016, 211 :741-747
[9]   Analytical strategies for coupling separation and flow-injection techniques [J].
Clavijo, Sabrina ;
Avivar, Jessica ;
Suarez, Ruth ;
Cerda, Victor .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2015, 67 :26-33
[10]   Metal concentrations in traditional and herbal teas and their potential risks to human health [J].
de Oliveira, Letuzia M. ;
Das, Suchismita ;
da Silva, Evandro B. ;
Gao, Peng ;
Gress, Julia ;
Liu, Yungen ;
Ma, Lena Q. .
SCIENCE OF THE TOTAL ENVIRONMENT, 2018, 633 :649-657