Sorting a large set of heavily used LiF:Mg,Ti thermoluminescent detectors into repeatable subsets of similar response

被引:8
作者
Kearfott, Kimberlee J. [1 ]
Newton, Jill P. [1 ]
Rafique, Muhammad [1 ,2 ]
机构
[1] Univ Michigan, Dept Nucl Engn & Radiol Sci, Ann Arbor, MI 48109 USA
[2] Univ Azad Jammu & Kashmir Muzaffarbad, Dept Phys, Azad Kashmir 13100, Pakistan
关键词
Thermoluminescent dosimeters; Cs-137; source; Sensitivity; Statistics; Calibration; PEAK FADING PROPERTIES; DOSIMETERS; PERFORMANCE; CALIBRATION; READINGS; DESIGN; SYSTEM; DRIFT; CT;
D O I
10.1016/j.apradiso.2014.10.019
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A set of 920 heavily used LiF:Mg,Ti thermoluminescent dosimeters (TLDs) was placed into a polymethyl methacrylate (PMMA) plate attached to a 40 x 40 x 15 cm(3) PMMA phantom and irradiated to 4.52 mGy using a Cs-137 source. This was repeated three times to determine the mean and standard deviation of each TLD's sensitivity. Reader drift was tracked over time with 10 control dosimeters. Two test sets of 100 TLDs were divided into subsets with sensitivities within +/- 1% of their subset means. All dosimeters were re-irradiated four times to test the TLDs' response repeatability and determine the sensitivity uniformity within the subsets. Coefficients of variation revealed that, within a given subset, the dosimeters responded within +/- 2.5% of their subset mean in all calibrations. The coefficient of variation in any of the 200 TLDs' calibrations was below 6% across the four calibrations. The work validates the approach of performing three calibrations to separate heavily used and aged TLDs with overall sensitivity variations of +/- 25% into subsets that reproducibly respond within +/- 2.5%. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:180 / 187
页数:8
相关论文
共 27 条
  • [1] American National Standards Institute, 2009, N13112009 ANSIHPS
  • [2] Performance of Vintage Direct Reading Pocket Ionization Chambers
    Bergen, Robert J.
    Harvey, John A.
    Kearfott, Kimberlee J.
    [J]. HEALTH PHYSICS, 2010, 98 (05): : S56 - S62
  • [3] Chauvenet W., 1891, A Manual of Spherical and Practical Astronomy, VII
  • [4] Calibration drift in a laboratory high purity germanium detector spectrometry system
    Dewey, S. C.
    Kearfott, K. J.
    [J]. HEALTH PHYSICS, 2008, 94 (02): : S27 - S33
  • [5] Design of an inexpensive, flexible, shielded cave for environmental radioactivity measurements
    Dewey, SC
    Kearfott, KJ
    Henley, CP
    Bernal, SA
    [J]. HEALTH PHYSICS, 2005, 88 (05): : S110 - S114
  • [6] Non-radiation induced signals in TL dosimetry
    German, U
    Weinstein, M
    [J]. RADIATION PROTECTION DOSIMETRY, 2002, 101 (1-4) : 81 - 84
  • [7] Quantification of Various Factors Influencing the Precision of Thermoluminescent Detector Calibrations for New and Used Chip Sets
    Harvey, J. A.
    Thomas, E. M.
    Kearfott, K. J.
    [J]. HEALTH PHYSICS, 2011, 100 (05): : S79 - S91
  • [8] Effects of high ambient temperature on glow-peak fading properties of Lif:Mg,Ti thermoluminescent dosemeters
    Harvey, John A.
    Kearfott, Kimberlee J.
    [J]. RADIATION PROTECTION DOSIMETRY, 2012, 149 (02) : 109 - 115
  • [9] THE EFFECTS OF HIGH AMBIENT RADON ON THERMOLUMINESCENCE DOSIMETRY READINGS
    Harvey, John A.
    Kearfott, Kimberlee J.
    [J]. RADIATION PROTECTION DOSIMETRY, 2011, 147 (04) : 491 - 497
  • [10] A computerized glow curve analysis (GCA) method for WinREMS thermoluminescent dosimeter data using MATLAB
    Harvey, John A.
    Rodrigues, Miesher L.
    Kearfott, Kimberlee J.
    [J]. APPLIED RADIATION AND ISOTOPES, 2011, 69 (09) : 1282 - 1286