FINDING THE RIGHT POSITION: A THREE-YEAR, SINGLE-CENTER EXPERIENCE WITH THE "SELF-LOCATING" CATHETER

被引:21
作者
Bergamin, Brigitte [1 ]
Senn, Oliver [2 ]
Corsenca, Alf [1 ]
Dutkowski, Philipp [3 ]
Weber, Markus [3 ]
Wuethrich, Rudolf P. [1 ]
Segerer, Stephan [1 ]
Thut, Martina Pechula [1 ]
机构
[1] Univ Zurich Hosp, Div Nephrol, CH-8091 Zurich, Switzerland
[2] Univ Zurich, Inst Gen Practice & Hlth Serv Res, Zurich, Switzerland
[3] Univ Zurich Hosp, Div Visceral & Transplant Surg, CH-8091 Zurich, Switzerland
来源
PERITONEAL DIALYSIS INTERNATIONAL | 2010年 / 30卷 / 05期
关键词
Catheter; self-locating; complications; displacement; PERITONEAL-DIALYSIS; TENCKHOFF CATHETER; MALFUNCTION RATE; ACCESS; COMPLICATIONS; FOLLOW;
D O I
10.3747/pdi.2009.00106
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Background: Peritoneal dialysis (PD) catheter displacement is a major cause of dysfunction. Various catheter forms have been designed to improve outcome. One of them is based on the classical Tenckhoff catheter, but includes a small tungsten cylinder at the distal end to optimize location and to prevent dislocations. Methods: In this retrospective study, we analyzed the functionality and complication profile of the self-locating catheter implanted at our center from November 2005 to September 2008 in our PD program. Results: Data from 27 self-locating catheters implanted in 25 patients (12 women) were obtained. Patients were followed for up to 30 months (median observation time: 13 months), resulting in a cumulative study period of 357 PD months. A total of 22 complications were identified, including 1 episode of peritonitis per 52 patient-months, 1 tunnel infection, and 1 exit-site infection. The "functional catheter failure incidence rate" was approximately 0.01 catheters per month on PD (or approximately 1 catheter loss per 100 PD months). The catheter survival was 93% at 1 year. Conclusions: The "self-locating" PD catheter demonstrates low rates of catheter migration and dislocations, and good catheter survival.
引用
收藏
页码:519 / 523
页数:5
相关论文
共 15 条
[1]  
Cavagna R, 1999, PERITON DIALYSIS INT, V19, P540
[2]  
Dell'Aquila R, 2007, PERITON DIALYSIS INT, V27, pS119
[3]   The self-locating catheter: Review and cost analysis [J].
Di Paolo, N ;
Sansoni, E ;
Cappelletti, F ;
Cavatorta, F ;
Galli, S ;
Nicolai, GA ;
Gaggiotti, E .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2006, 29 (01) :113-122
[4]  
Di Paolo N, 2004, PERITON DIALYSIS INT, V24, P359
[5]  
DiPaolo N, 1996, PERITON DIALYSIS INT, V16, P623
[6]  
Flanigan M, 2005, PERITON DIALYSIS INT, V25, P132
[7]   PERITONEAL ACCESS AND RELATED COMPLICATIONS IN CONTINUOUS AMBULATORY PERITONEAL-DIALYSIS [J].
GLOOR, HJ ;
NICHOLS, WK ;
SORKIN, MI ;
PROWANT, BF ;
KENNEDY, JM ;
BAKER, B ;
NOLPH, KD .
AMERICAN JOURNAL OF MEDICINE, 1983, 74 (04) :593-598
[8]  
Krediet RT, 2007, PERITON DIALYSIS INT, V27, pS35
[9]   Our nine-year experience with the self-locating catheter: Comparison of malfunction rate with other Tenckhoff catheter variants [J].
Lanuza, M ;
Minguela, JI ;
Rodado, R ;
Muriel, J ;
Ruiz-De-Gauna, R .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2006, 29 (01) :138-141
[10]  
Minguela I, 2001, PERITON DIALYSIS INT, V21, pS209