Low-cost EDA device for screening diabetic neuropathy

被引:2
|
作者
Daniel Bolanos, Luis [1 ]
Maria Vicente-Samper, Jose [2 ]
Zambrana Vinaroz, David [2 ]
Andres Vivas, Oscar [1 ]
Esteve Sala, Vicente [3 ]
Sabater-Navarro, Jose [2 ]
机构
[1] Univ Cauca, Dept Elect Instrumentat & Control, Popayan, Colombia
[2] Miguel Hernandez Univ Elche, nBio Res Grp, Elche, Spain
[3] Univ Alicante, Dept Lenguaje & Sistemas Informat, Alicante, Spain
来源
2019 IEEE 32ND INTERNATIONAL SYMPOSIUM ON COMPUTER-BASED MEDICAL SYSTEMS (CBMS) | 2019年
关键词
Electrodermal activity (EDA); Galvanic skin response (GSR); diabetic neuropathy; Bluetooth Low Energy (BLE); RESPONSES;
D O I
10.1109/CBMS.2019.00061
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Purpose: Poor electrodermal activity (EDA) may be related to diabetic neuropathy. The aims of this study were to investigate the characteristics of EDA, to design a portable electronic device capable of acquiring it and to formulate a hypothesis regarding the relationship between EDA and diabetic foot. Materials and methods: This study investigated the relationship between EDA signal and diabetic neuropathy using a self-designed device, and tested it with a procedure defined for diabetic foot screening. After that, mean skin conductance level (M-SCL), mean skin conductance response (M-SCR) to stimuli, and difference in mean SCL between feet (DBF) were obtained. Participants were assessed based on whether they had diabetes or not and on their clinical history and care habits. Afterwards, 3 groups were selected, namely reference, people with diabetes out of risk, and people with diabetes at risk of diabetic foot. Results: 22 patients were included in the final test. Several types of stimuli were applied through a test, and responses were recorded and analyzed. Diabetic patients reflected lower MSCL and M-SCR values. Statistical difference was found for those being at risk regarding M-SCR (P<0.01). Conclusion: A test protocol was defined, based on prior tests and measures, to acquire proper and reliable data. Lower SCL and SCR were found according to the diabetes progress. In particular, significant difference was found for M-SCR. No significant difference in DBF values for those at risk was found. These results have the potential to be used as part of a diagnostic tool, which may include other parameters and questionnaires, for diabetic foot screening.
引用
收藏
页码:253 / 258
页数:6
相关论文
共 50 条
  • [11] A new method of screening for diabetic neuropathy using laser Doppler and photoplethysmography
    Sung Woo Kim
    Soo Chan Kim
    Ki Chang Nam
    Eun Seok Kang
    Jae Jung Im
    Deok Won Kim
    Medical & Biological Engineering & Computing, 2008, 46 : 61 - 67
  • [12] Screening of Oxidative Stress Biomarkers and Inflammatory Cytokines in the Pathogenesis of Diabetic Neuropathy
    Devarapalli, P. Patricia
    Sudhakar, G.
    RESEARCH JOURNAL OF BIOTECHNOLOGY, 2024, 19 (07): : 108 - 119
  • [13] A new method of screening for diabetic neuropathy using laser Doppler and photoplethysmography
    Kim, Sung Woo
    Kim, Soo Chan
    Nam, Ki Chang
    Kang, Eun Seok
    Im, Jae Jung
    Kim, Deok Won
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2008, 46 (01) : 61 - 67
  • [14] Diagnostic Accuracy of Screening Tests for Diabetic Peripheral Neuropathy: An Umbrella Review
    Mogilevskaya, Maria
    Gaviria-Carrillo, Mariana
    Feliciano-Alfonso, John Edwin
    Barragan, Ana M.
    Calderon-Ospina, Carlos A.
    Nava-Mesa, Mauricio O.
    JOURNAL OF DIABETES RESEARCH, 2024, 2024
  • [15] Turkish version of the Michigan Neuropathy Screening Instrument in the assessment of diabetic peripheral neuropathy: a validity and reliability study
    Kaymaz, Serdar
    Alkan, Hakan
    Karasu, Ugur
    Cobankara, Veli
    DIABETOLOGY INTERNATIONAL, 2020, 11 (03) : 283 - 292
  • [16] Turkish version of the Michigan Neuropathy Screening Instrument in the assessment of diabetic peripheral neuropathy: a validity and reliability study
    Serdar Kaymaz
    Hakan Alkan
    Ugur Karasu
    Veli Çobankara
    Diabetology International, 2020, 11 : 283 - 292
  • [17] A low-cost Portable Device to Deliver Smoke, Volatile or Vaporized Substances to Drosophila melanogaster, Useful for Research and/or Educational Assays
    Santalla, Manuela
    Gomez, Ivana M.
    Valverde, Carlos A.
    Ferrero, Paola
    BIO-PROTOCOL, 2021, 11 (23):
  • [18] Development of Low-Cost in-the-Ear EEG Prototype
    Athavipach, Chanavit
    Pan-Ngum, Setha
    Israsena, Pasin
    2018 15TH INTERNATIONAL JOINT CONFERENCE ON COMPUTER SCIENCE AND SOFTWARE ENGINEERING (JCSSE), 2018, : 12 - 17
  • [19] Low-Cost Chlorophyll Fluorescence Imaging for Stress Detection
    Legendre, Reeve
    Basinger, Nicholas T.
    van Iersel, Marc W.
    SENSORS, 2021, 21 (06) : 1 - 17
  • [20] Wearable Movement Exploration Device with Machine Learning Algorithm for Screening and Tracking Diabetic Neuropathy-A Cross-Sectional, Diagnostic, Comparative Study
    Radunovic, Goran
    Velickovic, Zoran
    Pavlov-Dolijanovic, Slavica
    Janjic, Sasa
    Stojic, Biljana
    Velkova, Irena Jeftovic
    Suljagic, Nikola
    Soldatovic, Ivan
    BIOSENSORS-BASEL, 2024, 14 (04):