In situ and ex vivo evaluation of a wireless magnetoelastic biliary stent monitoring system

被引:13
作者
Green, Scott Ryan [1 ]
Kwon, Richard S. [2 ]
Elta, Grace H. [2 ]
Gianchandani, Yogesh B. [1 ,3 ]
机构
[1] Univ Michigan, Dept Mech Engn, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Dept Internal Med Gastroenterol, Ann Arbor, MI 48109 USA
[3] Univ Michigan, Dept Elect Engn & Comp Sci, Ann Arbor, MI 48109 USA
基金
美国国家科学基金会;
关键词
Implant; Resonant sensors; Biliary sludge; Mass sensors; Viscosity sensors;
D O I
10.1007/s10544-010-9404-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper presents the in situ and ex vivo evaluation of a system that wirelessly monitors the accumulation of intimal tissue and sludge in a biliary stent. The sensing element, located within the stent, is a magnetoelastic resonator that is queried by a wireless radio frequency signal. The in situ testing uses a commercially-available self-expanding biliary stent enhanced with a 1 mm x 25 mm magnetoelastic ribbon sensor (formed from Metglas (TM) 2605SA1). The stent has a conformal magnetic layer (consisting of strontium ferrite particles suspended in polydimethylsiloxane) that biases the sensor. The external interrogation module is able to acquire a signal from the sensor from a distance of at least 5 cm while the sensor is implanted in a porcine carcass and loaded with biological fluids. The ex vivo testing uses bile harvested from the porcine carcass. The response of a 1 mm x 25 mm magnetoelastic ribbon sensor is first calibrated with fluids of known density and viscosity, and the calibrated sensor is used to estimate that the viscosity of the harvested bile is 2.7-3.7 cP. The test results presented in this paper illustrate the fundamental usability of the system when the sensor is implanted, loaded by biological fluids, and interrogated in a surgical setup.
引用
收藏
页码:477 / 484
页数:8
相关论文
共 12 条
[1]  
*ASM INT, 1989, ASM HDB, V16
[2]  
Colas A., 2004, SILICONE BIOMATERIAL, V2nd, P80, DOI DOI 10.1038/LEU.2017.253
[3]   Boundary effect of a viscous fluid on a longitudinally vibrating bar: Theory and application [J].
Darvennes, CM ;
Pardue, SJ .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2001, 110 (01) :216-224
[4]   Plastic Biliary Stent Occlusion: Factors Involved and Possible Preventive Approaches [J].
Donelli, Gianfranco ;
Guaglianone, Emilio ;
Di Rosa, Roberta ;
Fiocca, Fausto ;
Basoli, Antonio .
CLINICAL MEDICINE & RESEARCH, 2007, 5 (01) :53-60
[5]  
Engdahl G., 2000, Handbook of Giant Magnetostrictive Materials, P1, DOI [10.1016/B978-0-12-238640-4.X5014-1, DOI 10.1016/B978-0-12-238640-4.X5014-1]
[6]  
GREEN S, 2010, J MICROMECH IN PRESS
[7]   Wireless Magnetoelastic Monitoring of Biliary Stents [J].
Green, Scott R. ;
Glanchandani, Yogesh B. .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2009, 18 (01) :64-78
[8]  
Jüngst D, 2001, WORLD J GASTROENTERO, V7, P203
[9]  
Kozarek RA, 2000, WORLD J GASTROENTERO, V6, P643
[10]   Magnetic and mechanical properties of micromachined strontium ferrite polyimide composites [J].
Lagorce, LK ;
Allen, MG .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 1997, 6 (04) :307-312