Biomechanical evaluation of implant-supported prosthesis with various tilting implant angles and bone types in atrophic maxilla: A finite element study

被引:26
|
作者
Gumrukcu, Zeynep [1 ]
Korkmaz, Yavuz Tolga [1 ]
Korkmaz, Fatih Mehmet [2 ]
机构
[1] Fac Dent, Dept Oral & Maxillofacial Surg, Trabzon, Turkey
[2] Fac Dent, Dept Prosthodont, Trabzon, Turkey
关键词
Sinus augmentation; Tilted implant; Stress analysis; Bone type; Von Mises stress; SHORT DENTAL IMPLANTS; RETAINED OVERDENTURES; STRESS-DISTRIBUTION; CORTICAL BONE; LENGTH; STABILITY; QUALITY; THICKNESS; DIAMETER; SURVIVAL;
D O I
10.1016/j.compbiomed.2017.04.015
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background and Objective: The purpose of this study is to evaluate and compare bone stress that occurs as a result of using vertical implants with simultaneous sinus augmentation with bone stress generated from oblique implants without sinus augmentation in atrophic maxilla. Methods: Six, three-dimensional (3D) finite element (FE) models of atrophic maxilla were generated with SolidWorks software. The maxilla models were varied for two different bone types. Models 2a, 2b and 2c represent maxilla models with D2 bone type. Models 3a, 3b and 3c represent maxilla models with D3 bone type. Five implants were embedded in each model with different configurations for vertical implant insertion with sinus augmentation: Model 2a/Model 3a, 30 degrees tilted insertion; Model 2b/Model 3b and 45 degrees tilted insertion; Model 2c/Model 3c. A 150 N load was applied obliquely on the hybrid prosthesis. The maximum von Mises stress values were comparatively evaluated using color scales. Results: The von Mises stress values predicted by the FE models were higher for all D3 bone models in both cortical and cancellous bone. For the vertical implant models, lower stress values were found in cortical bone. Tilting of the distal implants by 30 degrees increased the stress in the cortical layer compared to vertical implant models. Tilting of the distal implant by 45 degrees decreased the stress in the cortical bone compared to the 30 degrees models, but higher stress values were detected in the 45 degrees models compared to the vertical implant models. Conclusions: Augmentation should be the first treatment option in atrophic maxilla in terms of biomechanics. Tilted posterior implants can create higher stress values than vertical posterior implants. During tilting implant planning, the use of a 45 degrees tilted implant results in better biomechanical performance in peri-implant bone than 30 degrees tilted implant due to the decrease in cantilever length.
引用
收藏
页码:47 / 54
页数:8
相关论文
共 50 条
  • [1] Stress evaluation of different implant lengths on atrophic edentulous mandibles with fixed full-arch implant-supported prosthesis: a finite element analysis
    Benevides, Fernanda
    Cimoes, Renata
    Vajgel, Andre
    Wilmersdorf, Ramiro Brito
    Farias Vajgel, Bruna de Carvalho
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2021, 24 (04) : 358 - 374
  • [2] Finite element analysis of implant-supported prosthesis with pontic and cantilever in the posterior maxilla
    de Souza Batista, Victor Eduardo
    Verri, Fellippo Ramos
    de Faria Almeida, Daniel Augusto
    Santiago Junior, Joel Ferreira
    Lemos, Cleidiel Aparecido Araujo
    Pellizzer, Eduardo Piza
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2017, 20 (06) : 663 - 670
  • [3] Influence of implant number, length, and tilting degree on stress distribution in atrophic maxilla: a finite element study
    Zeynep Gümrükçü
    Yavuz Tolga Korkmaz
    Medical & Biological Engineering & Computing, 2018, 56 : 979 - 989
  • [4] Influence of implant number, length, and tilting degree on stress distribution in atrophic maxilla: a finite element study
    Gumrukcu, Zeynep
    Korkmaz, Yavuz Tolga
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2018, 56 (06) : 979 - 989
  • [5] Biomechanical Three-Dimensional Finite Element Analysis of Single Implant-Supported Prostheses in the Anterior Maxilla, with Different Surgical Techniques and Implant Types
    Verri, Fellippo Ramos
    Santiago, Joel Ferreira, Jr.
    Faria Almeida, Daniel Augusto
    de Souza Batista, Victor Eduardo
    Araujo Lemos, Cleidiel Aparecido
    Mello, Caroline Cantieri
    Pellizzer, Eduardo Piza
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2017, 32 (04) : E191 - E198
  • [6] Implant-Supported Prostheses in the Edentulous Mandible: Biomechanical Analysis of Different Implant Configurations via Finite Element Analysis
    Anitua, Eduardo
    de Ibarra, Naiara Larrazabal Saez
    Rotaeche, Luis Saracho
    DENTISTRY JOURNAL, 2023, 11 (01)
  • [7] Effect of Implant Diameter and Cortical Bone Thickness on Biomechanical Performance of Short Dental Implant-Supported Distal Cantilever: A Finite Element Study
    Muangsisied, Suthep
    Chantarapanich, Nattapon
    Veerasakul, Melvin Stanley
    Inglam, Samroeng
    ENGINEERING JOURNAL-THAILAND, 2021, 25 (02): : 175 - 182
  • [8] Finite Element Analysis of the Influence of the Offset Placement of an Implant-Supported Prosthesis on Bone Stress Distribution
    Anitua, E.
    Orive, G.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2009, 89B (02) : 275 - 281
  • [9] The effect of implant splinting on the load distribution in bone bed around implant-supported fixed prosthesis with different framework materials: A finite element study
    Hasan, I.
    Bourauel, C.
    Keilig, L.
    Stark, H.
    Lueckerath, W.
    ANNALS OF ANATOMY-ANATOMISCHER ANZEIGER, 2015, 199 : 43 - 51
  • [10] Evaluation of the Use of PEEK Material in Implant-Supported Fixed Restorations by Finite Element Analysis
    Tekin, S.
    Deger, Y.
    Demirci, F.
    NIGERIAN JOURNAL OF CLINICAL PRACTICE, 2019, 22 (09) : 1252 - 1258