Evaluation of a commercially-available block for spatially fractionated radiation therapy

被引:41
|
作者
Buckey, Courtney [1 ,2 ]
Stathakis, Sotirios [1 ,2 ]
Cashon, Ken [3 ]
Gutierrez, Alonso [1 ,2 ]
Esquivel, Carlos [1 ,2 ]
Shi, Chengyu [1 ,2 ]
Papanikolaou, Nikos [1 ,2 ]
机构
[1] Univ Texas Hlth Sci Ctr San Antonio, Dept Radiol, Canc Therapy & Res Ctr, San Antonio, TX 78229 USA
[2] Univ Texas Hlth Sci Ctr San Antonio, Dept Radiat Oncol, Canc Therapy & Res Ctr, San Antonio, TX 78229 USA
[3] Dot Decimal Inc, Sanford, FL 32771 USA
来源
关键词
spatially fractionated radiotherapy; GRID therapy; compensators; CANCER;
D O I
10.1120/jacmp.v11i3.3163
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this paper, we present the dosimetric characteristics of a commercially-produced universal GRID block for spatially fractioned radiation therapy. The dosimetric properties of the GRID block were evaluated. Ionization chamber and film measurements using both Kodak EDR2 and Gafchromic EBT film were performed in a solid water phantom to determine the relative output of the GRID block as well as its spatial dosimetric characteristics. The surface dose under the block and at the openings was measured using ultra thin TLDs. After introducing the GRID block into the treatment planning system, a treatment plan was created using the GRID block and also by creating a GRID pattern using the multi-leaf collimator. The percent depth doses measured with film showed that there is a shift of the dmax towards shallower depths for both energies (6 MV and 18 MV) under investigation. It was observed that the skin dose at the GRID openings was higher than the corresponding open field by a factor as high as 50% for both photon energies. The profiles showed the transmission under the block was in the order of 15-20% for 6 MV and 30% for 18 MV. The MUs calculated for a real patient using the block were about 80% less than the corresponding MUs for the same plan using the multileaf collimator to define the GRID. Based on this investigation, this brass GRID compensator is a viable alternative to other solid compensators or MLC-based fields currently in use. Its ease of creation and use give it decided advantages. Its ability to be created once and used for multiple patients (by varying the collimation of the linear accelerator jaws) makes it attractive from a cost perspective. We believe this compensator can be put to clinical use, and will allow more centers to offer GRID therapy to their patients.
引用
收藏
页码:2 / 11
页数:10
相关论文
共 50 条
  • [1] An optimal grid block design for spatially fractionated radiation therapy
    Gholami, S.
    Nedaie, H. A.
    Longo, F.
    Ay, M. R.
    Meigooni, A. S.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S920 - S920
  • [2] Use of a commercially-available bio-feedback device in respiratory correlated radiation therapy (RCRT)
    Waldron, T
    Prado, K
    Starkschall, G
    MEDICAL PHYSICS, 2004, 31 (06) : 1926 - 1926
  • [3] Evaluation of Five Commercially-Available Algorithms for Deformable Image Registration
    Shah, A.
    Pukala, J.
    Johnson, P.
    Staton, R.
    Meeks, S.
    MEDICAL PHYSICS, 2014, 41 (06) : 399 - 399
  • [4] DOSIMETRIC EVALUATION OF A GRID COMPENSATOR FOR SPATIALLY FRACTIONATED RADIATION THERAPY
    Buckey, C.
    Stathakis, S.
    Papanikolaou, N.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S234 - S234
  • [5] Evaluation of Quantitative Error of a Commercially-Available Positron Emission Mammography Scanner
    Springer, A.
    Mawlawi, O.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [6] Evaluation of three commercially-available chikungunya virus immunoglobulin G immunoassays
    De Salazar, Pablo M.
    Marie Valadere, Anne
    Goodman, Christin H.
    Johnson, Barbara W.
    REVISTA PANAMERICANA DE SALUD PUBLICA-PAN AMERICAN JOURNAL OF PUBLIC HEALTH, 2017, 41
  • [7] Evaluation of quantitative error of a commercially-available positron emission mammography scanner
    Springer, Adam
    Mawlawi, Osama
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [8] Evaluation of GRID and Spatially Fractionated Radiation Therapy: Dosimetry and Preclinical Trial
    Johnson, T.
    Bassil, A.
    Kent, C.
    Williams, N.
    Palmer, G.
    Mowery, Y.
    Oldham, M.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [9] An overview of the diversity represented in commercially-available databases
    Mary P. Bradley
    Molecular Diversity, 2000, 5 : 175 - 183
  • [10] COMMERCIALLY-AVAILABLE GAMES FOR SPEECH COMMUNICATION COURSES
    WEATHERLY, M
    SPEECH TEACHER, 1974, 23 (04): : 312 - 319