Revision total hip arthroplasty performed after fracture of a ceramic femoral head - A multicenter survivorship study

被引:117
作者
Allain, J [1 ]
Roudot-Thoraval, F [1 ]
Delecrin, J [1 ]
Anract, P [1 ]
Migaud, H [1 ]
Goutallier, D [1 ]
机构
[1] Hop Henri Mondor, Serv Chirurg Orthoped, F-94010 Creteil, France
关键词
D O I
10.2106/00004623-200305000-00009
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The alumina ceramic femoral head was introduced for total hip arthroplasty approximately thirty years ago. One of its main drawbacks was the risk of implant fracture. The aim of this study was to examine the results of revision total hip replacement performed specifically to treat a fracture of a ceramic femoral head and to identify technical factors that affected the outcomes. Methods: One hundred and five surgical revisions to treat a fracture of a ceramic femoral head, performed at thirty-five institutions, were studied. The patients were examined clinically by the operating surgeon at the time of the last follow-up. The surgeon provided the latest follow-up radiographs, which were compared with the immediate postoperative radiographs. The success of the revisions was assessed with Kaplan-Meier survivorship analysis, with the need for repeat revision as the end point. We evaluated the complication rate and radiographic changes indicative of implant loosening. The average duration of follow-up between the index revision and the last clinical and radiographic review was 3.5 years (range, six months to twenty years). Results: Following the revisions, radiographic evidence of cup loosening was seen in twenty-two hips (21%) and radiographic evidence of femoral loosening was seen in twenty-two (21%). One or several repeat revisions were necessary in thirty-three patients (31%) because of infection (four patients), implant loosening (twenty), osteolysis (eight), or fracture of the revision femoral head component (one). The survival rate at five years was 63% (95% confidence interval, 51% to 75%). The survival rate was significantly worse when the cup had not been changed, when the new femoral head was made of stainless steel, when a total synovectomy had not been done, and when the patient was less than fifty years old. Conclusions: Fracture of a ceramic femoral head component is a rare but potentially serious event. A suitable surgical approach, including total synovectomy, cup exchange, and insertion of a cobalt-chromium or new ceramic femoral ball minimizes the chance of early loosening of the implants and the need for one or more repeat revisions. Level of Evidence: Therapeutic study, Level I (case series [no, or historical, control group]). See Instructions to Authors for a complete description of levels of evidence.
引用
收藏
页码:825 / 830
页数:6
相关论文
共 25 条
  • [1] Poor eight-year survival of cemented zirconia-polyethylene total hip replacements
    Allain, J
    Le Mouel, S
    Goutallier, D
    Voisin, MC
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (05): : 835 - 842
  • [2] Failure of a stainless-steel femoral head of a revision total hip arthroplasty performed after a fracture of a ceramic femoral head - A case report
    Allain, J
    Goutallier, D
    Voisin, MC
    Lemouel, S
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1998, 80A (09) : 1355 - 1360
  • [3] LONG-TERM RESULTS OF UNCEMENTED ALUMINA ACETABULAR IMPLANTS
    BOEHLER, M
    KNAHR, K
    PLENK, H
    WALTER, A
    SALZER, M
    SCHREIBER, V
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (01): : 53 - 59
  • [4] THE USE OF DENSE ALUMINA ALUMINA CERAMIC COMBINATION IN TOTAL HIP-REPLACEMENT
    BOUTIN, P
    CHRISTEL, P
    DORLOT, JM
    MEUNIER, A
    DEROQUANCOURT, A
    BLANQUAERT, D
    HERMAN, S
    SEDEL, L
    WITVOET, J
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1988, 22 (12): : 1203 - 1232
  • [5] BOUTIN P, 1971, PRESSE MED, V79, P639
  • [6] Boutin P.M., 1981, ORTHOPEDIC CERAMIC I, V1, P11
  • [7] Callaway GH, 1995, J ARTHROPLASTY, V10, P855
  • [8] MECHANICAL-PROPERTIES AND SHORT-TERM INVIVO EVALUATION OF YTTRIUM-OXIDE-PARTIALLY-STABILIZED ZIRCONIA
    CHRISTEL, P
    MEUNIER, A
    HELLER, M
    TORRE, JP
    PEILLE, CN
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1989, 23 (01): : 45 - 61
  • [9] CHRISTEL PS, 1989, B HOSP JOINT DIS ORT, V49, P170
  • [10] Fritsch EW, 1996, CLIN ORTHOP RELAT R, P129