Direct measurement of oxygen reduction reactions at neurostimulation electrodes

被引:29
作者
Ehlich, Jiri [1 ]
Migliaccio, Ludovico [1 ]
Sahalianov, Ihor [1 ]
Nikic, Marta [1 ,2 ]
Brodsky, Jan [1 ]
Gablech, Imrich [1 ]
Xuan Thang Vu [3 ]
Ingebrandt, Sven [3 ]
Glowacki, Eric Daniel [1 ]
机构
[1] Brno Univ Technol, Cent European Inst Technol CEITEC, Bioelect Mat & Devices Lab, Purkynova 123, Brno 61200, Czech Republic
[2] Tech Univ Munich, Inst Neuroelect, Munich, Germany
[3] Rhein Westfal TH Aachen, Inst Mat Elect Engn 1, D-52074 Aachen, Germany
基金
欧洲研究理事会;
关键词
bioelectronics; neurostimulation; faradaic reactions; hypoxia; reactive oxygen species; HYDROGEN-PEROXIDE; CHARGE INJECTION; DIFFUSION-COEFFICIENTS; NEURAL STIMULATION; PLATINUM; ELECTROCHEMISTRY;
D O I
10.1088/1741-2552/ac77c0
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. Electric stimulation delivered by implantable electrodes is a key component of neural engineering. While factors affecting long-term stability, safety, and biocompatibility are a topic of continuous investigation, a widely-accepted principle is that charge injection should be reversible, with no net electrochemical products forming. We want to evaluate oxygen reduction reactions (ORR) occurring at different electrode materials when using established materials and stimulation protocols. Approach. As stimulation electrodes, we have tested platinum, gold, tungsten, nichrome, iridium oxide, titanium, titanium nitride, and poly(3,4-ethylenedioxythiophene):poly(styrene sulfonate). We use cyclic voltammetry and voltage-step amperometry in oxygenated versus inert conditions to establish at which potentials ORR occurs, and the magnitudes of diffusion-limited ORR currents. We also benchmark the areal capacitance of each electrode material. We use amperometric probes (Clark-type electrodes) to quantify the O-2 and H2O2 concentrations in the vicinity of the electrode surface. O-2 and H2O2 concentrations are measured while applying DC current, or various biphasic charge-balanced pulses of amplitude in the range 10-30 mu C cm(-2)/phase. To corroborate experimental measurements, we employ finite element modelling to recreate 3D gradients of O-2 and H2O2. Main results. All electrode materials support ORR and can create hypoxic conditions near the electrode surface. We find that electrode materials differ significantly in their onset potentials for ORR, and in the extent to which they produce H2O2 as a by-product. A key result is that typical charge-balanced biphasic pulse protocols do lead to irreversible ORR. Some electrodes induce severely hypoxic conditions, others additionally produce an accumulation of hydrogen peroxide into the mM range. Significance. Our findings highlight faradaic ORR as a critical consideration for neural interface devices and show that the established biphasic/charge-balanced approach does not prevent irreversible changes in O-2 concentrations. Hypoxia and H2O2 can result in different (electro)physiological consequences.
引用
收藏
页数:15
相关论文
共 33 条
[1]   Faradaic Pixels for Precise Hydrogen Peroxide Delivery to Control M-Type Voltage-Gated Potassium Channels [J].
Abdullaeva, Oliya S. ;
Sahalianov, Ihor ;
Silvera Ejneby, Malin ;
Jakesova, Marie ;
Zozoulenko, Igor ;
Liin, Sara I. ;
Glowacki, Eric Daniel .
ADVANCED SCIENCE, 2022, 9 (03)
[2]   Tutorial: guidelines for standardized performance tests for electrodes intended for neural interfaces and bioelectronics [J].
Boehler, Christian ;
Carli, Stefano ;
Fadiga, Luciano ;
Stieglitz, Thomas ;
Asplund, Maria .
NATURE PROTOCOLS, 2020, 15 (11) :3557-3578
[3]   Safety and efficacy of spinal cord stimulation for the treatment of chronic pain: a 20-year literature review [J].
Cameron, T .
JOURNAL OF NEUROSURGERY, 2004, 100 (03) :254-267
[4]   Neural stimulation and recording electrodes [J].
Cogan, Stuart F. .
ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 2008, 10 :275-309
[5]   Contribution of Oxygen Reduction to Charge Injection on Platinum and Sputtered Iridium Oxide Neural Stimulation Electrodes [J].
Cogan, Stuart F. ;
Ehrlich, Julia ;
Plante, Timothy D. ;
Gingerich, Marcus D. ;
Shire, Douglas B. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2010, 57 (09) :2313-2321
[6]   PEDOT:PSS-Coated Stimulation Electrodes Attenuate Irreversible Electrochemical Events and Reduce Cell Electropermeabilization [J].
Dijk, Gerwin ;
Ruigrok, Hermanus J. ;
O'Connor, Rodney P. .
ADVANCED MATERIALS INTERFACES, 2021, 8 (19)
[7]   Tailoring PEDOT properties for applications in bioelectronics [J].
Donahue, Mary J. ;
Sanchez-Sanchez, Ana ;
Inal, Sahika ;
Qu, Jing ;
Owens, Roisin M. ;
Mecerreyes, David ;
Malliaras, George G. ;
Martin, David C. .
MATERIALS SCIENCE & ENGINEERING R-REPORTS, 2020, 140
[8]   A novel organotypic long-term culture of the rat hippocampus on substrate-integrated multielectrode arrays [J].
Egert, U ;
Schlosshauer, B ;
Fennrich, S ;
Nisch, W ;
Fejtl, M ;
Knott, T ;
Muller, T ;
Hammerle, H .
BRAIN RESEARCH PROTOCOLS, 1998, 2 (04) :229-242
[9]   Oxidative modification of M-type K+ channels as a mechanism of cytoprotective neuronal silencing [J].
Gamper, Nikita ;
Zaika, Oleg ;
Li, Yang ;
Martin, Pamela ;
Hernandez, Ciria C. ;
Perez, Michael R. ;
Wang, Andrew Y. C. ;
Jaffe, David B. ;
Shapiro, Mark S. .
EMBO JOURNAL, 2006, 25 (20) :4996-5004
[10]   Neuroprosthetic applications of electrical stimulation [J].
Grill, WM ;
Kirsch, RF .