DYNAMIC BEHAVIOR OF BONE-IMPLANT SUBSTITUTE SYSTEMS. NUMERICAL AND EXPERIMENTAL STUDY

被引:0
作者
Nowak, Bartosz [1 ]
Kaczmarek, Mariusz [1 ]
Kubik, Jozef [1 ]
机构
[1] Kazimierz Wielki Univ Bydgoszcz, Inst Environm Mech & Appl Comp Sci, Bydgoszcz, Poland
关键词
dynamics; biomechanics; bone; implant; TOTAL HIP-ARTHROPLASTY;
D O I
暂无
中图分类号
O3 [力学];
学科分类号
08 ; 0801 ;
摘要
In this paper, a study of substitute systems of an implanted bone was presented from the point of view of elaboration of a method for diagnosing the state of integrity of the bone-implant attachment based on vibrational techniques. Dynamic parameters of the systems were eshablished as sets of eigenfrequencies/resonance frequencies using numerical procedures and experimental modal analysis. The shifts of values of the parameters were discussed in relation to the loosening of contact between the bone and implant. The numerical and experimental results were compared and their consequences for application of vibrational techniques to real anatomical systems discussed.
引用
收藏
页码:617 / 628
页数:12
相关论文
共 11 条
[1]  
[Anonymous], 1996, FINITE ELEMENT PROCE
[2]   Accurate diagnosis of hip prosthesis loosening using a vibrational technique [J].
Georgiou, AP ;
Cunningham, JL .
CLINICAL BIOMECHANICS, 2001, 16 (04) :315-323
[3]   Twenty-five-year results after charnley total hip arthroplasty in patients less than fifty years old - A concise follow-up of a previous report [J].
Keener, JD ;
Callaghan, JJ ;
Goetz, DD ;
Pederson, DR ;
Sullivan, PM ;
Johnston, RC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (06) :1066-1072
[4]  
Kleiber M., 1995, KOMPUTEROWE METODY M
[5]  
LIEBERMAN JR, 1993, J BONE JOINT SURG BR, V75, P475
[6]  
ONIELL DA, 1984, J BONE JOINT SURG, V66, P540
[7]   PROSTHETIC HIP REPLACEMENTS - PLAIN FILMS AND ARTHROGRAPHY FOR COMPONENT LOOSENING [J].
PHILLIPS, WC ;
KATTAPURAM, SV .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1982, 138 (04) :677-682
[8]   How much can a vibrational diagnostic tool reveal in total hip arthroplasty loosening? [J].
Qi, G ;
Mouchon, WP ;
Tan, TE .
CLINICAL BIOMECHANICS, 2003, 18 (05) :444-458
[9]  
Rosenstein A D, 1989, Proc Inst Mech Eng H, V203, P77, DOI 10.1243/PIME_PROC_1989_203_014_01
[10]  
Zienkiewicz OC, 2001, FINITE ELEMENT METHO, V1