Requirements for synthetic diamond devices for radiotherapy dosimetry applications

被引:50
作者
Guerrero, MJ [1 ]
Tromson, D [1 ]
Rebisz, M [1 ]
Mer, C [1 ]
Bazin, B [1 ]
Bergonzo, P [1 ]
机构
[1] CEA Saclay, LTD, SSTC, DETECS,LIST,CEA Rech Technol, F-91191 Gif Sur Yvette, France
关键词
polycrystalline CVD diamond; radiotherapy; dosimetry; ionisation chamber; thermoluminescence;
D O I
10.1016/j.diamond.2004.07.026
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
CVD diamond is a remarkable material for the fabrication of radiation detectors. Radiation hardness, chemical resistance and hightemperature operation capabilities of diamond motivate its use for fabrication of devices operating in hostile environments such as that encountered in nuclear industry and high energy physics. Its potentialities for such applications have been well documented and recent studies have led to the developments of a few applications that are addressing specific industrial needs. One particular interest of diamond stands in the fact that its atomic number is close to that of human tissues. This implies that the response of a diamond device to radiation is close to that received by the human body. Its thus enables the straightforward measurement of the dose for radiotherapy applications. However, this requires high reproducibility and linearity. It is widely observed that radiation exposure is modifying the initial performances of diamond detectors and priming devices is therefore required to obtain the required linearity. However, the nature of defects in the material strongly influences the type of priming required. This paper will address this problem from the study of trapping levels and their influence on the device response. We present here the current status of the development of polycrystalline diamond for this type of application, and propose new techniques of improving the material characteristics toward the optimisation of ionisation chamber performances as well as that of thermoluminescent dosimeters for the particular field of radiotherapy applications. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:2046 / 2051
页数:6
相关论文
共 16 条
[1]   An investigation of the operating characteristics of two PTW diamond detectors in photon and electron beams [J].
De Angelis, C ;
Onori, S ;
Pacilio, M ;
Cirrone, GAP ;
Cuttone, G ;
Raffaele, L ;
Bucciolini, M ;
Mazzocchi, S .
MEDICAL PHYSICS, 2002, 29 (02) :248-254
[2]   PTW-diamond detector: Dose rate and particle type dependence [J].
Fidanzio, A ;
Azario, L ;
Miceli, R ;
Russo, A ;
Piermattei, A .
MEDICAL PHYSICS, 2000, 27 (11) :2589-2593
[3]  
FIDANZIO A, 2004, ACCEL SPECTROM DETEC, V524, P115
[4]   Properties of nickel nitrogen complexes in diamond: stability and electronic structure [J].
Johnston, K ;
Mainwood, A .
DIAMOND AND RELATED MATERIALS, 2003, 12 (3-7) :516-520
[5]   Modelling transition metals in diamond [J].
Johnston, K ;
Mainwood, A .
DIAMOND AND RELATED MATERIALS, 2002, 11 (3-6) :631-634
[6]   Experimental and theoretical studies of cobalt defects in diamond [J].
Johnston, K ;
Mainswood, A ;
Collins, AT ;
Davies, G ;
Twitchen, D ;
Newton, M ;
Baker, JM .
DIAMOND AND RELATED MATERIALS, 2000, 9 (3-6) :424-427
[7]   DIAMOND DOSIMETER FOR X-RAY AND GAMMA-RADIATION [J].
KOZLOV, SF ;
KONOROVA, EA ;
KUZNETSOV, YA ;
SALIKOV, YA ;
REDKO, VI ;
GRINBERG, VR ;
MEILMAN, ML .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1977, 24 (01) :235-237
[8]   Application of CVD diamonds as dosimeters of soft X-ray emission from plasma sources [J].
Krása, J ;
Juha, L ;
Vorlicek, V ;
Cejnarová, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 524 (1-3) :332-339
[9]   The value of the PinPoint ion chamber for characterization of small field segments used in intensity-modulated radiotherapy [J].
Martens, C ;
De Wagter, C ;
De Neve, W .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (09) :2519-2530
[10]  
McKeever S. W. S., 1985, CAMBRIDGE SOLID STAT