Determination of 238Pu in plutonium bearing fuels by thermal ionization mass spectrometry

被引:15
作者
Aggarwal, S. K. [1 ]
Alamelu, D. [1 ]
Khodade, P. S. [1 ]
Shah, P. M. [1 ]
机构
[1] Bhabha Atom Res Ctr, Div Fuel Chem, Bombay 400085, Maharashtra, India
关键词
ISOTOPE CORRELATIONS; ALPHA-SPECTROMETRY; TIMS;
D O I
10.1007/s10967-007-0947-x
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Determination of Pu-238 in plutonium bearing fuels is required as a part of the chemical quality assurance of nuclear fuels. In addition, the determination of Pu-238 is required in nuclear technology for many other applications, e.g., for developing isotope correlations and while using Pu-238 as a spike (tracer) in isotope dilution alpha-spectrometry (IDAS). This determination usually involves the use of alpha-spectrometry on purified Pu sample. In view of the random errors associated with the counting statistics and the systematic errors due to (1) in-growth of Am-241 in purified Pu sample and (2) tail contribution correction methodology in alpha-spectrometry, the precision and accuracy obtainable by alpha-spectrometry are limited. Thermal ionization mass spectrometry (TIMS) is generally used for the determination of different Pu isotopes other than Pu-238. This is due to the ubiquitous isobaric interference from U-238 at Pu-238 in TIMS. Recently, we have carried out studies on the formation of atomic and oxide ions of U and Pu by TIMS and developed a novel approach using interfering element correction methodology to account for the isobaric interference of U-238 at Pu-238 in TIMS. This methodology is based on the addition of U-235 (enrichment > 90 atom%) to Pu sample followed by the determination of U-238/U-235 atom ratio using UO+ ion and determination of Pu isotope ratios using Pu+ ion, from the same filament loading. The TIMS methodology was used for the determination of Pu-238 in different Pu samples in U based nuclear fuels from PHWRs with Pu-238 content about 0.2 atom%. The Pu-238 determination was also carried out using alpha-spectrometry. This paper reports the results obtained by the two methods and presents the merits and shortcomings of the two approaches.
引用
收藏
页码:775 / 778
页数:4
相关论文
共 6 条
[1]   Novel approach for the determination of 238Pu by thermal ionization mass spectrometry (TIMS) using interfering element correction methodology [J].
Aggarwal, SK ;
Alamelu, D .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2005, 241 (01) :83-88
[2]  
Aggarwal SK, 1998, RADIOCHIM ACTA, V81, P129
[3]  
AGGARWAL SK, 1980, RADIOCHIM ACTA, V27, P1
[4]   Investigations on atomic and oxide ion formation of plutonium and uranium in thermal ionization mass spectrometry (TIMS) for determination of 238Pu [J].
Alamelu, D ;
Khodade, PS ;
Shah, PM ;
Aggarwal, SK .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2004, 239 (01) :51-56
[5]   Isotope correlations for determining the isotopic composition of plutonium [J].
Alamelu, D ;
Aggarwal, SK .
RADIOCHIMICA ACTA, 2001, 89 (03) :131-134
[6]   ISOTOPE-DILUTION ALPHA-SPECTROMETRY FOR THE DETERMINATION OF PLUTONIUM CONCENTRATION IN IRRADIATED FUEL DISSOLVER SOLUTION - IDAS AND R-IDAS [J].
RAMANIAH, MV ;
JAIN, HC ;
AGGARWAL, SK ;
CHITAMBAR, SA ;
KAVIMANDAN, VD ;
ALMAULA, AI ;
SHAH, PM ;
PARAB, AR ;
SANT, VL .
NUCLEAR TECHNOLOGY, 1980, 49 (01) :121-128