Does metal transfer differ on retrieved Biolox® Delta composites femoral heads? Surface investigation on three Biolox® generations from a biotribological point of view

被引:22
作者
Affatato, Saverio [1 ]
Ruggiero, Alessandro [2 ]
Merola, Massimiliano [2 ]
Logozzo, Silvia [3 ]
机构
[1] Rizzoli Orthopaed Inst, Med Technol Lab, Via Barbiano 1-10, I-40136 Bologna, Italy
[2] Univ Salerno, Dept Ind Engn, Salerno, Italy
[3] V GER Srl, Dept Res & Dev, Via Mori 6, Bologna, Italy
关键词
Biotribology; Biolox (R) Delta; Metal transfer; Digital surface analysis; Roughness; IN-VITRO; TRIBOLOGICAL BEHAVIOR; ROUGHNESS MEASUREMENT; PHASE-TRANSFORMATION; ACETABULAR CUP; CERAMIC HIP; STRIPE WEAR; ALUMINA; JOINT; MICROSEPARATION;
D O I
10.1016/j.compositesb.2017.01.026
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Background: Currently, little is known about the effects that metal transfer affecting hip implants may have on roughness. The main aim of this study is to propose a metrological technique in order to better characterize this phenomenon on the surface of composite ceramic femoral heads, discussing its influence on the biotribological behaviour of the coupling. In addition, through the acquired data, we investigate whether the metal transfer extension and roughness alterations on a femoral head is correlated to the material, the follow-up of the implant and whether metal transfer is associated with changes in the roughness of alumina surface, considering different follow-ups. Methods: Three generations of femoral heads manufactured in ceramic (Biolox (R), Biolox (R) Forte, and Biolox (R) Delta) were retrieved from patients after different implantation periods (mean follow up = 6.9 years) and they were analysed using a roughness tester and an optical non-contact 3D scanner, in order to characterize the metal transfer on the bearing surface. Results: Surface area where the metal transfer was observed ranged between 29.6 mm(2) and 573.6 mm(2). The mean values of Sa (average roughness) were different between the Affected Zone (0.3 +/- 0.1 mu m) and Unaffected Zone (0.03 +/- 0.1 mu m). In particular, regions of metal transfer were rougher than the non-metal transfer regions of the articular surface and this is demonstrated by both the areal roughness parameters considered: Sa (arithmetic mean height) and St (maximum height of the surface) parameters (p = 0.002, Wilcoxon paired-test). St parameter acquired on the Biolox (R) Delta femoral heads showed the lowest mean value (11.92 mu m). Conclusion: We found no correlation between follow-up vs. roughness and total surface area of metal transfer. Furthermore, we found that the amount of areal coverage of metal transfer is correlated with the surface roughness. Finally, MT effect on composite ceramic implants caused lower height of peaks and valleys when compared to common ceramic. (C) 2017 Elsevier Ltd. All rights reserved.
引用
收藏
页码:164 / 173
页数:10
相关论文
共 53 条
[11]   Microseparation and stripe wear in alumina-on-alumina hip implants [J].
Affatato, Saverio ;
Traina, Francesco ;
Toni, Aldo .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2011, 34 (06) :506-512
[12]  
[Anonymous], 1996, Geometrical product specification (GPS) - Surface texture: profile method
[13]  
surfaces having stratified functional properties -part2: Height characterization using the linear material ratio curve
[14]   On the Improvement of AA2024 Wear Properties through the Deposition of a Cold-Sprayed Titanium Coating [J].
Astarita, Antonello ;
Rubino, Felice ;
Carlone, Pierpaolo ;
Ruggiero, Alessandro ;
Leone, Claudio ;
Genna, Silvio ;
Merola, Massimiliano ;
Squillace, Antonino .
METALS, 2016, 6 (08)
[15]   The significance of metal staining on alumina femoral heads in total hip arthroplasty [J].
Bal, B. Sonny ;
Rahaman, Mohamed N. ;
Aleto, Thomas ;
Miller, F. Scott ;
Traina, Francesco ;
Toni, Aldo .
JOURNAL OF ARTHROPLASTY, 2007, 22 (01) :14-19
[16]   Clinical failure analysis of contemporary ceramic-on-ceramic total hip replacements [J].
Brandt, Jan-M ;
Gascoyne, Trevor C. ;
Guenther, Leah E. ;
Allen, Andrew ;
Hedden, David R. ;
Turgeon, Thomas R. ;
Bohm, Eric R. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2013, 227 (08) :833-846
[17]   High strength and toughness alumina matrix composites by transformation toughening and 'in situ' platelet reinforcement (ZPTA) - The new generation of bioceramics [J].
Burger, W ;
Richter, HG .
BIOCERAMICS, 2000, 192-1 :545-548
[18]  
Chan FW, 1999, CLIN ORTHOP RELAT R, P10
[19]   Transfer of metallic debris from the meta surface of an acetabular cup to artificial femoral heads by scraping: Comparison between alumina and cobalt-chrome heads [J].
Chang, Chong Bum ;
Yoo, Jeong Joon ;
Song, Won Seok ;
Kim, Deug Joong ;
Koo, Kyung-Hoi ;
Kim, Hee Joong .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2008, 85B (01) :204-209
[20]  
Chen D, 2010, MINERVA ORTOP TRAUMA, V61, P43