Cardiac tamponade as complication of active-fixation atrial lead perforations: Proposed mechanism and management algorithm

被引:32
作者
Geyfman, Vitaly
Storm, Randle H.
Lico, Serrie C.
Oren, Jess W.
机构
[1] Geisinger Med Ctr, Dept Cardiol, Danville, PA 17822 USA
[2] Geisinger Med Ctr, Dept Cardiothorac Surg, Danville, PA 17822 USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2007年 / 30卷 / 04期
关键词
cardiac tamponade; pericardium; pacemaker lead perforation;
D O I
10.1111/j.1540-8159.2007.00699.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Cardiac tamponade is a rare complication after implantation of dual chamber pacemaker or defibrillator systems. Its pathophysiology and optimal management are not currently well established. Methods: Three cases of cardiac tamponade following successful implantation of transvenous dual chamber pacemakers with active-fixation atrial leads were identified. Results: All three patients with post-implant cardiac tamponade were suspected to have the same etiology of bleeding into the pericardial space. This was due to protrusion of the helix of the active-fixation atrial pacing lead through the atrial wall with subsequent abrasion of visceral pericardial layer and bleeding from the atrium through the perforation. In two patients, the perforation sites were visualized and repaired during open thoracotomy in the operating room. The third patient underwent lead repositioning under fluoroscopic guidance in the electrophysiology laboratory. Conclusion: Based on the reviewed cases, we describe the pathophysiology of, and recommend a safe conservative algorithm for, the management of cardiac tamponade after successful transvenous lead implantation. Percutaneous pericardiocentesis with placement of the pericardial drain followed by lead repositioning under fluoroscopic guidance with surgical backup appears to be safe and effective.
引用
收藏
页码:498 / 501
页数:4
相关论文
共 16 条
[1]   Oozing from the pericardium as an etiology of cardiac tamponade associated with screw-in atrial leads [J].
Aizawa, K ;
Kaneko, Y ;
Yamagishi, T ;
Utsugi, T ;
Suzuki, T ;
Ishikawa, S ;
Otaki, A ;
Morishita, Y ;
Hasegawa, A ;
Kurabayashi, M ;
Nagai, R .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (03) :381-383
[2]   Delayed complications following pacemaker implantation [J].
Ellenbogen, KA ;
Wood, MA ;
Shepard, RK .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (08) :1155-1158
[3]  
GIKSON M, 1994, PACE, V17, P1399
[4]   ACUTE PERICARDITIS RESULTING FROM AN ENDOCARDIAL ACTIVE FIXATION SCREW-IN ATRIAL LEAD [J].
GREENE, TO ;
PORTNOW, AS ;
HUANG, SKS .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (01) :21-25
[5]   Right pneumothorax resulting from an endocardial screw-in atrial lead [J].
Ho, WJ ;
Kuo, CT ;
Lin, KH .
CHEST, 1999, 116 (04) :1133-1134
[6]   Aortic perforation by active-fixation atrial pacing lead [J].
Kashani, A ;
Mehdirad, A ;
Fredman, C ;
Barold, SS .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (03) :417-418
[7]   Delayed lead perforation: A disturbing trend [J].
Khan, MN ;
Joseph, G ;
Khaykin, Y ;
Ziada, KM ;
Wilkoff, BL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (03) :251-253
[8]   Bipolar active fixation atrial leads: Comparison of two new lead models [J].
Kindermann, M ;
Schwaab, B ;
Frohlig, G ;
Lawall, P ;
Schieffer, H .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (11) :2285-2290
[9]   RELIABILITY OF ATRIAL SCREW-IN LEADS [J].
MARKEWITZ, A ;
WENKE, K ;
WEINHOLD, C .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11) :1777-1783
[10]   Right pneumothorax resulting from an endocardial screw-in atrial lead in an implantable cardioverter defibrillator system [J].
Oginosawa, Y ;
Abe, H ;
Nakashima, Y .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (08) :1278-1279