Proton transfer reaction mass spectrometry at high drift tube pressure
被引:39
作者:
Hanson, DR
[1
]
Greenberg, J
[1
]
Henry, BE
[1
]
Kosciuch, E
[1
]
机构:
[1] Natl Ctr Atmospher Res, Div Atmospher Chem, Boulder, CO 80307 USA
关键词:
proton transfer reaction mass spectrometry;
drift tube pressure;
D O I:
10.1016/S1387-3806(02)00924-7
中图分类号:
O64 [物理化学(理论化学)、化学物理学];
O56 [分子物理学、原子物理学];
学科分类号:
070203 ;
070304 ;
081704 ;
1406 ;
摘要:
Anion drift tube operated at a total pressure of 10 Torr (similar to13 mbar) coupled to amass spectrometer is described. Proton transfer from water molecules was used to detect organic compounds and calibrations of the instrument yield normalized sensitivities that are in accord with what is expected for operation at this pressure. For example, the sensitivity for acetone was found to be 240 Hz ppbv(-1) for 1 MHz of reagent ion signal while that for isoprene was 120 Hz ppbv(-1). Issues particular to using this technique at pressures of 10 Torr, considerably higher than the conventional pressure, are addressed. Detection limits for a 1 s integration (S/N = 2) depend on the cleanliness history of the system but can reach as low as 20 pptv. (Int J Mass Spectrom 223-224 (2003) 507-518) (C) 2002 Elsevier Science B.V. All rights reserved.