Portable gas chromatograph with tunable retention and sensor array detection for determination of complex vapor mixtures

被引:102
作者
Lu, CJ
Whiting, J
Sacks, RD
Zellers, ET [1 ]
机构
[1] Univ Michigan, Dept Environm Hlth Sci, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Dept Chem, Ctr Wireless Integrated MicroSyst, Ann Arbor, MI 48109 USA
关键词
D O I
10.1021/ac026092n
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A prototype portable gas chromatograph that combines a multiadsorbent preconcentrator/focuser, a tandem-column separation stage with individual column temperature control and junction point pressure modulation, and a detector consisting of an integrated array of polymer-coated surface acoustic wave microsensors is described. Using scheduled first-column stop-flow intervals and independent temperature programming of the two columns, it is possible to adjust the retention of eluting analyte vapors to maximize vapor recognition with the microsensor array and minimize the time of analysis. A retention window approach is combined with Monte Carlo simulations to guide retention tuning requirements and facilitate pattern recognition analyses. The determination of a 30-vapor mixture of common indoor air contaminants in < 10 min is demonstrated using ambient air as the carrier gas. Detection limits of < 10 ppb are achieved for the majority of compounds from a 1-L air sample on the basis of the most sensitive sensor in the array. Performance is assessed in the context of near-real-time indoor air quality monitoring applications.
引用
收藏
页码:1400 / 1409
页数:10
相关论文
共 46 条
[31]  
REPKA D, 1990, HRC-J HIGH RES CHROM, V13, P333
[32]   Advancing the science of column selectivity [J].
Sacks, R ;
Coutant, C ;
Grall, A .
ANALYTICAL CHEMISTRY, 2000, 72 (15) :524A-533A
[33]   SELECTIVITY AND SELECTIVITY TUNING IN CAPILLARY GAS-CHROMATOGRAPHY [J].
SANDRA, P ;
DAVID, F ;
PROOT, M ;
DIRICKS, G ;
VERSTAPPE, M ;
VERZELE, M .
JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY & CHROMATOGRAPHY COMMUNICATIONS, 1985, 8 (11) :782-798
[34]   Sign of eye irritation in female hospital workers and the indoor environment [J].
Smedbold, HT ;
Ahlen, C ;
Norbäck, D ;
Hilt, B .
INDOOR AIR, 2001, 11 (04) :223-231
[35]   Pressure-tunable GC columns with electronic pressure control [J].
Smith, H ;
Sacks, R .
ANALYTICAL CHEMISTRY, 1997, 69 (24) :5159-5164
[36]  
*US EPA, 1997, EPA625R96010B
[37]   STANDARDIZATION OF 2ND-ORDER INSTRUMENTS [J].
WANG, YD ;
KOWALSKI, BR .
ANALYTICAL CHEMISTRY, 1993, 65 (09) :1174-1180
[38]   Selectivity enhancement for high-speed GC analysis of volatile organic compounds with portable instruments designed for vacuum-outlet and atmospheric-pressure inlet operation using air as the carrier gas [J].
Whiting, J ;
Sacks, R .
ANALYTICAL CHEMISTRY, 2002, 74 (01) :246-252
[39]   A portable, high-speed, vacuum-outlet GC vapor analyzer employing air as carrier gas and surface acoustic wave detection [J].
Whiting, JJ ;
Lu, CJ ;
Zellers, ET ;
Sacks, RD .
ANALYTICAL CHEMISTRY, 2001, 73 (19) :4668-4675
[40]  
Williams D, 1999, FIELD ANAL CHEM TECH, V3, P45, DOI 10.1002/(SICI)1520-6521(1999)3:1<45::AID-FACT6>3.3.CO