PERFORMANCE EVALUATION OF PROBING SYSTEMS IN DATA CAPTURE FOR KINEMATIC PARAMETER IDENTIFICATION AND VERIFICATION OF ARTICULATED ARM COORDINATE MEASURING MACHINES

被引:0
作者
Santolaria, Jorge [1 ]
Jose Aguilar, Juan [1 ]
Brau, Agustin [1 ]
Javier Brosed, Francisco [1 ]
机构
[1] CPS Univ Zaragoza, Design & Mfg Engn Dept, Zaragoza, Spain
来源
XIX IMEKO WORLD CONGRESS: FUNDAMENTAL AND APPLIED METROLOGY, PROCEEDINGS | 2009年
关键词
Articulated arm coordinate measuring machines; Performance evaluation; Probing system comparison; CALIBRATION; IMPROVEMENT; ACCURACY;
D O I
暂无
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This paper presents a comparison of different techniques to capture nominal data for its use in later verification and kinematic parameter identification procedures for articulated arm coordinate measuring machines (AACMM). By using four different probing systems ( passive spherical probe, active spherical probe, self-centering passive probe and self-centering active probe) the accuracy and repeatability of captured points has been evaluated. The nominal points are materialized by a ball-bar gauge distributed in several positions of the measurement volume. By comparing these systems it is possible to characterize the influence of the force over the final results for each of the gauge and probing system configurations. The results with each of the systems studied show the advantages and original accuracy obtained by active probes, and thus their suitability in verification ( active probes) and kinematic parameter identification (self-centering active probes) procedures.
引用
收藏
页码:1846 / 1851
页数:6
相关论文
共 13 条
  • [1] A systematic technique to estimate positioning errors for robot accuracy improvement using laser interferometry based sensing
    Alici, G
    Shirinzadeh, B
    [J]. MECHANISM AND MACHINE THEORY, 2005, 40 (08) : 879 - 906
  • [2] [Anonymous], 1955, J APPL MECH
  • [3] EVERETT LJ, 1987, IEEE C ROBOTICS AUTO, V1, P183
  • [4] Nongeometric error identification and compensation for robotic system by inverse calibration
    Gong, CH
    Yuan, JX
    Ni, J
    [J]. INTERNATIONAL JOURNAL OF MACHINE TOOLS & MANUFACTURE, 2000, 40 (14) : 2119 - 2137
  • [5] IMPROVING THE ABSOLUTE POSITIONING ACCURACY OF ROBOT MANIPULATORS
    HAYATI, S
    MIRMIRANI, M
    [J]. JOURNAL OF ROBOTIC SYSTEMS, 1985, 2 (04): : 397 - 413
  • [6] Hsu T.-W., 1985, Proceedings of the Computation in Engineering Conference, P263
  • [7] PUTTOCK MJ, 1969, 25 CSIRO NAT STDS LA
  • [8] Kinematic parameter estimation technique for calibration and repeatability improvement of articulated arm coordinate measuring machines
    Santolaria, Jorge
    Aguilar, Juan-Jose
    Yaguee, Jose-Antonio
    Pastor, Jorge
    [J]. PRECISION ENGINEERING-JOURNAL OF THE INTERNATIONAL SOCIETIES FOR PRECISION ENGINEERING AND NANOTECHNOLOGY, 2008, 32 (04): : 251 - 268
  • [9] SHETH PN, 1971, ASME, P102
  • [10] STONE HW, 1986, IEEE INT C ROB AUT, V1, P41